 3.0 FACTORS INFLUENCING ACCURATE READINGS

3.1 Effects of Temperature

The UltraMaxO

2

 compensates for temperature and will perform within specifications 

throughout the operating temperature range. However, taking measurements during 
rapid changes in gas temperature should be avoided.

3.2 Effects of Humidity

The UltraMaxO

2

 has a humidity sensor to detect and compensate for the humidity of 

gas entering the device. However, high levels (condensing) of humidity can affect the 
accuracy and reliability of the UltraMaxO

2

. To prevent possible damage: 

• 

Avoid usage in environments of greater than 95% relative humidity. 

• 

DO NOT

 use this device in a breathing circuit.

• 

DO NOT

 breathe or blow into the UltraMaxO

2

3.3 Effects of Other Gases

The UltraMaxO

2

 is designed to measure two different types of gas mixtures:

• 

Oxygen, nitrogen and argon from oxygen concentrators.

• 

Pure oxygen during calibration verification mode.

Any other concentrations or combinations of gases will cause the UltraMaxO

2

 to 

measure oxygen concentration incorrectly.

3.4 Effects of Low Flow

Oxygen concentrators function on the principle of removing nitrogen gas from air, 
leaving concentrated oxygen and argon at a specific oxygen to argon ratio.  This 
operating principle may be altered when concentrators are set to flow at the low end of 
their operational range.  At low flows they may output a low oxygen concentration, e.g. 
85% to 91%, for reasons other than high nitrogen, possibly due to an increase in argon 
content.  The UltraMaxO2 requires that the ratio of oxygen to argon remain constant in 
order to guarantee an accuracy of +/-1.5% oxygen.

• 

DO NOT use the UltraMaxO

to measure the oxygen concentration of a 

concentrator when flowing at rates lower than its optimal performance as specified 
by the concentrator manufacturer; generally 4 LPM or less on concentrators that 
have a maximum flow of 10 LPM, and 1 LPM or less on concentrators that have a 
maximum flow of 5 LPM.

 4.0 ERROR CODES

The UltraMaxO2 has self diagnostic features built into the software to detect 
faulty readings outside of normal operating ranges. The codes, descriptions and 
recommended actions are: 

• 

E01:

 Oxygen measurement out of range Hi (

102.0% calculated by algorithm).

Recommended Action: Verify that the UltraMaxO

2

 is being used in the correct 

mode (Concentrator or Calibration Verification mode). If error code repeats; 
perform a calibration verification per section 2.3 of this manual. If error code 
repeats again; contact Customer Service.

Содержание UltraMaxO2

Страница 1: ...UltraMaxO2 O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 2 1 P 1 1 R221M11 Rev C O X Y G E N A N A LY Z E R...

Страница 2: ......

Страница 3: ...ieu of all other warranties expressed or implied including warranty of merchantability and fitness for a particular purpose CLASSIFICATIONS Classification Class II electrical device Protection against...

Страница 4: ...ther than conventional oxygen concentrators Not for use with inhalation agents Operating the UltraMaxO2 in flammable or explosive environments may result in fire or explosion Not suitable for use in t...

Страница 5: ...in minor or moderate injury and property damage Replace the batteries with high quality AA Alkaline or Lithium batteries DO NOT use rechargeable batteries When not in use for periods greater than 30...

Страница 6: ...1 Oxygen Flow and Pressure Measurement 3 2 2 Switching Pressure Units of Measure 3 2 3 Calibration Verification Procedure 3 3 0 FACTORS INFLUENCING ACCURATE READINGS 4 3 1 Effects of Temperature 4 3...

Страница 7: ...following features and operational benefits Accurate oxygen measurements No in field calibration required Convenient ability to measure pressure in PSI or kPa Durable compact design Large easy to rea...

Страница 8: ...nated next to the flow measurement Not shown when in calibration verification mode KPA Indicates the pressure measurement is in units of kilopascals PSI Indicates the pressure measurement is in units...

Страница 9: ...pressure cover the gas sample outlet with thumb or finger while gas is flowing 7 Wait 5 seconds for the display to read pressure 2 2 Switching Pressure Units of Measure The UltraMaxO2 can measure pre...

Страница 10: ...gas from air leaving concentrated oxygen and argon at a specific oxygen to argon ratio This operating principle may be altered when concentrators are set to flow at the low end of their operational r...

Страница 11: ...closer to room temperature before use E08 Device self check found error Recommended Action Remove and replace the batteries If error code repeats return the UltraMaxO2 to the manufacturer for factory...

Страница 12: ...ent Range 0 10 LPM Flow Measurement Accuracy 0 2 LPM Flow Measurement Resolution 0 1 LPM PRESSURE PSI kPa Pressure Measurement Range 0 5 50 3 4 344 Pressure Measurement Accuracy 0 5 0 5 Pressure Measu...

Страница 13: ...Your Unit Part Number Item R211M11 Operating Manual and Instructions for Use RP46P11 003 Gas Sample Tubing 8 2 Optional Accessories Part Number Item R221P15 Soft Cover An electronic version of this Op...

Отзывы: