Accuracy of oxygen concentration will be affected if bleed is not activated at flow settings below 15
l/min for the High Flow Blender, and 3 l/min for the Low Flow Blender.
DO NOT
obstruct the alarm.
DO NOT
use Blender when alarm is sounding.
DO NOT
use oil in or around the Blender.
DO NOT
occlude or obstruct the bleed port on the auxiliary outlet of the Blender.
DO NOT
use near any type of flame or flammable/explosive substances, vapors or atmosphere.
◆
◆
Oxygen Concentration Dial does not rotate 360 degrees. Rotating the dial less than 21% or over
100% oxygen will damage the Blender.
CAUTION
◆
◆
Turn off gas supplies when Air-Oxygen Blender is not in use.
◆
◆
Store the Air-Oxygen Blender in a clean, dry area when not in use.
◆
◆
The Air-Oxygen Blender contains magnetic, ferrous material that may affect the results of an MRI.
◆
◆
Ensure all connections are tight and leak free.
◆
◆
Avoid excessive pressure surges greater than 100 psi (6.9 bar) when pressuring the Blender inlets.
DO NOT
steam autoclave.
DO NOT
immerse Air-Oxygen Blender into any liquid.
DO NOT
gas sterilize with (EtO) Ethylene Oxide.
DO NOT
use if dirt or contaminants are present on or around the Blender or connecting devices.
DO NOT
smoke in an area where oxygen is being administered.
DO NOT
clean with aromatic hydrocarbons.
◆
◆
Inlet pressure of device used in conjunction with Blender must match inlet pressure of Blender.
◆
◆
When using a bottled high pressure gas source, always use a pressure reducing regulator set within
30-75 psi (2.1-5.2 bar).