WWW.MAXTEC.COM • (800) 748-5355
7
ENGLISH
Drift of Measurement .......< +/-1% of full scale at constant temperature, pressure and humidity
A Model Dimensions.................................. 3.0”(W) x 4.0”(H) x 1.5”(D) [76mm x 102mm x 38mm]
A Weight ......................................................................................................................... 0.4 lbs. (170g)
AE Model Dimensions.............................3.0”(W) x 36.0”(H) x 1.5”(D) [76mm x 914mm x38mm] ]
Height includes external cable length (retracted)
AE Weight .......................................................................................................................0.6 lbs. (285g)
8.2 Sensor Specifications
Type .....................................................................................................Galvanic fuel sensor (0-100%)
Life .......................................................................................................2-years in typical applications
9.0 MAXO2+ SPARE PARTS AND
ACCESSORIES
9.1 Included With Your Unit
PART NUMBER ITEM
A MODEL
AE MODEL
R217M40
User’s Guide and Operating Instructions
x
x
RP76P06
Lanyard
x
x
R110P10-001
Flow Diverter
x
x
RP16P02
Blue Tee Adapter
x
x
R217P23
Dovetail Bracket
x
9.2 Standard Replacement Parts and
Accessories
PART NUMBER ITEM
A MODEL
AE MODEL
R125P02-011
Max-250+ Oxygen Sensor
x
R125P03-002
Max-250E Oxygen Sensor
x
R115P85
Max-250ESF Oxygen Sensor
x
R217P08
Gasket
x
x
RP06P25
#4-40 Pan Head Stainless Steel Screw
x
x
R217P16-001
Front Assembly (Includes Board & LCD)
x
x
R217P11-002
Back Assembly
x
x
R217P24
Coiled Cable Assembly
x
R217P09-001
Overlay
x
x
9.3 Optional Accessories
9.3.1 Optional Adapters
PART NUMBER
ITEM
RP16P02
Blue Tee Adapter
R103P90
Perfusion Tee Adapter
RP16P05
Pediatric Tee Adapter
R207P17
Threaded Adapter with Tygon Tubing
9.3.2 Mounting Options (requires dovetail
R217P23)
PART NUMBER
ITEM
R206P75
Pole Mount
R205P86
Wall Mount
R100P10
Rail Mount
R206P76
Horizontal Pole Mount
NOTE:
Repair of this equipment must be performed by a qualified service technician experi-
enced in repair of portable hand held medical equipment.
Equipment in need of repair shall be sent to:
Maxtec
Service Department
2305 South 1070 West
Salt Lake City, Ut 84119
(Include RMA number issued by customer service)
10.0 ELECTROMAGNETIC
COMPATIBILITY
The information contained in this section (such as separation distances) is in general spe-
cifically written with regard to the MaxO2+ A/AE. The numbers provided will not guarantee
faultless operation but should provide reasonable assurance of such. This information may
not be applicable to other medical electrical equipment; older equipment may be particularly
susceptible to interference.
Note: Medical electrical equipment requires special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC infor-
mation provided in this document and the remainder of the instructions for use this device.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Cables and accessories not specified within the instructions for use are not authorized. Using
other cables and/or accessories may adversely impact safety, performance and electromag-
netic compatibility (increased emission and decreased immunity).
Care should be taken if the equipment is used adjacent to or stacked with other equipment;
if adjacent or stacked use is inevitable, the equipment should be observed to verify normal
operation in the configuration in which it will be used.
ELECTROMAGNETIC EMISSIONS
This equipment is intended for use in the electromagnetic environment specified below.
The user of this equipment should assure that it is used in such an environment.
EMISSIONS
COMPLIANCE
ACCORDING TO
ELECTROMAGNETIC ENVIRONMENT
RF Emissions
(CISPR 11)
Group 1
The MaxO2+ uses RF energy only for its
internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
CISPR Emissions
Classification
Class A
The MaxO2+ is suitable for use in all
establishments other than domestic and
those directly connected to the public low-
voltage power supply network that supplies
buildings used for domestic purposes.
NOTE:
The EMISSIONS characteristics of this
equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used
in a residential environment (for which CISPR
11 class B is normally required) this equipment
might not offer adequate protection to radio-
frequency communication services. The user
might need to take mitigation measures, such
as relocating or re-orienting the equipment.
Harmonic Emissions
(IEC 61000-3-2)
Class A
Voltage Fluctuations Complies