Maxtec MaxO2 Plus Скачать руководство пользователя страница 7

WWW.MAXTEC.COM  •  (800) 748-5355 

7

 ENGLISH

Drift of Measurement .......< +/-1% of full scale at constant temperature, pressure and humidity 

A Model Dimensions.................................. 3.0”(W) x 4.0”(H) x 1.5”(D) [76mm x 102mm x 38mm]

A Weight ......................................................................................................................... 0.4 lbs. (170g)

AE Model Dimensions.............................3.0”(W) x 36.0”(H) x 1.5”(D) [76mm x 914mm x38mm] ]

Height includes external cable length (retracted)

AE Weight .......................................................................................................................0.6 lbs. (285g)

8.2 Sensor Specifications

Type .....................................................................................................Galvanic fuel sensor (0-100%)

Life .......................................................................................................2-years in typical applications

9.0 MAXO2+ SPARE PARTS AND 

ACCESSORIES

9.1 Included With Your Unit

PART NUMBER ITEM

A MODEL

AE MODEL

R217M40

User’s Guide and Operating Instructions

x

x

RP76P06

Lanyard

x

x

R110P10-001

Flow Diverter

x

x

RP16P02

Blue Tee Adapter

x

x

R217P23

Dovetail Bracket

x

9.2 Standard Replacement Parts and 

Accessories

PART NUMBER ITEM

A MODEL

AE MODEL

R125P02-011

Max-250+ Oxygen Sensor

x

R125P03-002

Max-250E Oxygen Sensor

x

R115P85

Max-250ESF Oxygen Sensor

x

R217P08

Gasket

x

x

RP06P25

#4-40 Pan Head Stainless Steel Screw

x

x

R217P16-001

Front Assembly (Includes Board & LCD)

x

x

R217P11-002

Back Assembly

x

x

R217P24

Coiled Cable Assembly

x

R217P09-001

Overlay

x

x

9.3 Optional Accessories
9.3.1 Optional Adapters

PART NUMBER 

ITEM

RP16P02 

Blue Tee Adapter

R103P90 

Perfusion Tee Adapter

RP16P05 

Pediatric Tee Adapter

R207P17 

Threaded Adapter with Tygon Tubing

9.3.2 Mounting Options (requires dovetail 

R217P23)

PART NUMBER 

ITEM

R206P75

Pole Mount

R205P86

Wall Mount

R100P10

Rail Mount

R206P76

Horizontal Pole Mount

NOTE:

 Repair of this equipment must be performed by a qualified service technician experi-

enced in repair of portable hand held medical equipment.
Equipment in need of repair shall be sent to:
Maxtec 

Service Department 

2305 South 1070 West 

Salt Lake City, Ut  84119
(Include RMA number issued by customer service)

10.0 ELECTROMAGNETIC 

COMPATIBILITY

The information contained in this section (such as separation distances) is in general spe-

cifically written with regard to the MaxO2+ A/AE. The numbers provided will not guarantee 

faultless operation but should provide reasonable assurance of such. This information may 

not be applicable to other medical electrical equipment; older equipment may be particularly 

susceptible to interference.
Note: Medical electrical equipment requires special precautions regarding electromagnetic 

compatibility (EMC) and needs to be installed and put into service according to the EMC infor-

mation provided in this document and the remainder of the instructions for use this device.
Portable and mobile RF communications equipment can affect medical electrical equipment.
Cables and accessories not specified within the instructions for use are not authorized. Using 

other cables and/or accessories may adversely impact safety, performance and electromag-

netic compatibility (increased emission and decreased immunity).
Care should be taken if the equipment is used adjacent to or stacked with other equipment; 

if adjacent or stacked use is inevitable, the equipment should be observed to verify normal 

operation in the configuration in which it will be used.

ELECTROMAGNETIC EMISSIONS

This equipment is intended for use in the electromagnetic environment specified below.
The user of this equipment should assure that it is used in such an environment.

EMISSIONS

COMPLIANCE 
ACCORDING TO

ELECTROMAGNETIC ENVIRONMENT

RF Emissions 
(CISPR 11)

Group 1

The MaxO2+ uses RF energy only for its 
internal function. Therefore, its RF emissions 
are very low and are not likely to cause any 
interference in nearby electronic equipment.

CISPR Emissions 
Classification

Class A

The MaxO2+ is suitable for use in all 
establishments other than domestic and 
those directly connected to the public low-
voltage power supply network that supplies 
buildings used for domestic purposes.

NOTE:

 The EMISSIONS characteristics of this 

equipment make it suitable for use in industrial 
areas and hospitals (CISPR 11 class A). If it is used 
in a residential environment (for which CISPR 
11 class B is normally required) this equipment 
might not offer adequate protection to radio-
frequency communication services. The user 
might need to take mitigation measures, such 
as relocating or re-orienting the equipment.

Harmonic Emissions  
(IEC 61000-3-2)

Class A

Voltage Fluctuations Complies

Содержание MaxO2 Plus

Страница 1: ...MaxO2 Instructions for Use E N G L I S H R217M40 REV U...

Страница 2: ...ient s head or neck which may result in strangulation Before use all individuals who will be using the MaxO2 must become thoroughly familiar with the information contained in this Operation Manual Str...

Страница 3: ...achieves 90 of final value in approximately 15 seconds at room temperature Large easy to read 3 1 2 digit LCD display for readings in the 0 100 range Simple operation and easy one key calibration Sel...

Страница 4: ...the sensor open to static ambient air For optimum accuracyMaxtecrecommendsthattheSensorbeplacedinaclosedloopcircuitwheregasflow ismovingacrossthesensorinacontrolledmanner Calibratewiththesametypeofci...

Страница 5: ...oxygen sensor see note below Replace batteries CAL ERR ST O2 Sensor reading not stable Wait for displayed oxygen reading to stabilize when calibrating the device at 100 oxygen Wait for unit to reach...

Страница 6: ...nterclockwiseand pullthesensorfromtheconnection Replacethenewsensorbyinsertingtheelectricalplugfrom the coiled cord into the receptacle on the oxygen sensor Rotate the thumbscrew clockwise until snug...

Страница 7: ...reasonable assurance of such This information may not be applicable to other medical electrical equipment older equipment may be particularly susceptible to interference Note Medical electrical equip...

Страница 8: ...mon mode 2 kV Differential mode 1 kV 3 A m power frequency magnetic field 50 60 Hz IEC 61000 4 8 30 A m 50 Hz or 60 Hz Voltage dips and short interruptions on AC mains input lines IEC 61000 4 11 Dip 9...

Страница 9: ...9 This page intentionally left blank...

Страница 10: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com...