Maxtec MaxO2 Plus Скачать руководство пользователя страница 5

WWW.MAXTEC.COM  •  (800) 748-5355 

5

 ENGLISH

3.2 Temperature Effects

The MaxO2+ will hold calibration and read correctly within ±3% when at thermal equilibrium 

within the operating temperature range. The device must be thermally stable when calibrated 

and allowed to thermally stabilize after experiencing temperature changes before readings 

are accurate. For these reasons, the following is recommended:

 

For best results, perform the calibration procedure at a temperature close to the 

temperature where analysis will occur.

 

Allow adequate time for the sensor to equilibrate to a new ambient temperature.

 CAUTION:

 “CAL Err St” may result from a sensor that has not reached thermal equilibrium.

When used in a breathing circuit, place the sensor upstream of the heater.

3.3 Pressure Effects

Readings from the MaxO2+ are proportional to the partial pressure of oxygen. The partial pres-

sure is equal to the concentration times the absolute pressure.
Thus, the readings are proportional to the concentration if the pressure is held constant. 

Therefore, the following are recommended:

 

Calibrate the MaxO2+ at the same pressure as the sample gas.

 

If sample gases flow through tubing, use the same apparatus and flow rates when 

calibrating as when measuring.

3.4 Humidity Effects

Humidity (non-condensing) has no effect on the performance of the MaxO2+ other than dilut-

ing the gas, as long as there is no condensation. Depending on the humidity, the gas may be 

diluted by as much as 4%, which proportionally reduces the oxygen concentration. The device 

responds to the actual oxygen concentration rather than the dry concentration. Environments 

where condensation may occur are to be avoided since moisture may obstruct passage of 

gas to the sensing surface, resulting in erroneous readings and slower response time. For this 

reason, the following is recommended:

 

Avoid usage in environments greater than 95% relative humidity.

 

When used in a breathing circuit, place the sensor upstream of the humidifier.

HELPFUL HINT:

 Dry sensor by lightly shaking moisture out, or flow a dry gas at two liters per 

minute across the sensor membrane.

4.0 CALIBRATION ERRORS AND 

ERROR CODES

The MaxO2+ analyzers have a self test feature built into the software to detect faulty calibra-

tions, oxygen sensor failures, and low operating voltage. These are listed below, and include 

possible actions to take, if an error code occurs.

E02:

 No sensor attached

 

MaxO2+A: Open unit and disconnect and reconnect sensor. Unit should perform an auto 

calibration and should read 20.9%. If not, contact Maxtec Customer Service for possible 

sensor replacement.

 

MaxO2+AE: Disconnect and reconnect external sensor. Unit should perform an auto 

calibration, and should read 20.9%. If not, contact Maxtec Customer Service for possible 

sensor replacement or cable replacement.

E03:

 No valid calibration data available

 

Make sure unit has reached thermal equilibrium. Press and hold the Calibration Button 

for three seconds to manually force a new calibration.

E04:

 Battery below minimum operating voltage

 

Replace batteries.

E06:

 Non-compatible oxygen sensor, see note below.

 

Replace batteries.

CAL ERR ST:

 O2 Sensor reading not stable

 

Wait for displayed oxygen reading to stabilize, when calibrating the device at 100% 

oxygen.

 

Wait for unit to reach thermal equilibrium, (Please note that this can take up to one 

half hour, if the device is stored in temperatures outside the specified operating 

temperature range).

CAL ERR LO:

 Sensor voltage too low

 

Press and hold the Calibration Button for three seconds to manually force a new 

calibration. If unit repeats this error more than three times, contact Maxtec Customer 

Service for possible sensor replacement.

CAL ERR HI:

 Sensor voltage too high

 

Press and hold the Calibration Button for three seconds to manually force a new 

calibration. If unit repeats this error more than three times, contact Maxtec Customer 

Service for possible sensor replacement.

CAL ERR BAT:

 Battery voltage too low to recalibrate

 

Replace batteries.

NOTE:

 If you receive a E01, E05, or an E07 error code, correct by ensuring the calibration gas is 

either room air or 100% oxygen. Also ensure the calibration gas flow, pressure and concentra-

tion is constant. Allow sufficient time for the sensor to stabilize in the calibration gas and with 

room temperature, then attempt to calibrate again.
If these steps do not correct the error, contact Maxtec for technical support.

NOTE:

 Use only a Maxtec approved Max-550E sensor called out in Section 9.0 of the Spare Parts 

List. The Max550E sensor is equipped with an authentication chip to ensure the monitor is used 

with an approved sensor.

NOTE:

 Correcting E02, E06, or E07 errors:

1.  Disconnect the sensor and reconnect, making sure the male plug is fully inserted into 

the receptacle before tightening the threaded locking shroud. The analyzer should now 

perform a new calibration with the error cleared.

2.  If the error still persists, remove the batteries, wait 30 seconds, then reinstall to 

perform a factory reset and diagnostic on the analyzer. The analyzer should again 

perform a new calibration with the error cleared.

3.  Contact Maxtec Customer Service Department if the error code cannot be cleared.

5.0 CHANGING THE BATTERIES

Batteries should be changed by service personnel.

 

Use only brand name batteries.

 

Replace with two AA batteries and insert per orientation marked on the device.

Should the batteries require changing, the device will indicate this in one of two ways:

 

The battery icon on the bottom of the display will begin to flash. This icon will continue 

to flash until the batteries are changed. The unit will continue to function normally for 

approx. 200 hours.

 

If the device detects a very low battery level, an error code of “E04” will be present on 

the display, and the unit will not function until the batteries are changed.

To change the batteries, begin by removing the 

three screws from the back of the device. 

A #1 Phillips screwdriver is required to 

remove these screws.
Once the screws are removed, 

gently separate the two halves 

of the device.
The batteries can now be replaced from 

the back half of the case. Be sure to orient the new 

batteries as indicated in the embossed polarity on 

the back case.

NOTE:

 If the batteries are installed incorrectly the 

batteries will not make contact and the device will 

not operate.

FIGURE 2

Содержание MaxO2 Plus

Страница 1: ...MaxO2 Instructions for Use E N G L I S H R217M40 REV U...

Страница 2: ...ient s head or neck which may result in strangulation Before use all individuals who will be using the MaxO2 must become thoroughly familiar with the information contained in this Operation Manual Str...

Страница 3: ...achieves 90 of final value in approximately 15 seconds at room temperature Large easy to read 3 1 2 digit LCD display for readings in the 0 100 range Simple operation and easy one key calibration Sel...

Страница 4: ...the sensor open to static ambient air For optimum accuracyMaxtecrecommendsthattheSensorbeplacedinaclosedloopcircuitwheregasflow ismovingacrossthesensorinacontrolledmanner Calibratewiththesametypeofci...

Страница 5: ...oxygen sensor see note below Replace batteries CAL ERR ST O2 Sensor reading not stable Wait for displayed oxygen reading to stabilize when calibrating the device at 100 oxygen Wait for unit to reach...

Страница 6: ...nterclockwiseand pullthesensorfromtheconnection Replacethenewsensorbyinsertingtheelectricalplugfrom the coiled cord into the receptacle on the oxygen sensor Rotate the thumbscrew clockwise until snug...

Страница 7: ...reasonable assurance of such This information may not be applicable to other medical electrical equipment older equipment may be particularly susceptible to interference Note Medical electrical equip...

Страница 8: ...mon mode 2 kV Differential mode 1 kV 3 A m power frequency magnetic field 50 60 Hz IEC 61000 4 8 30 A m 50 Hz or 60 Hz Voltage dips and short interruptions on AC mains input lines IEC 61000 4 11 Dip 9...

Страница 9: ...9 This page intentionally left blank...

Страница 10: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com...

Отзывы: