Maxtec MaxO2 OM-25A Скачать руководство пользователя страница 8

866.4.Maxtec    www.maxtecinc.com

2

 

7

    

 and 

 Keys-

 The up and down arrow keys are used in 

conjunction with the CALIBRATE key to calibrate the device. 

C

 Sample Inlet Connection

 1.3 MAX-250 Oxygen Sensor

MAX-250 oxygen sensors offer quick response, stability and extra life on 

the order of 1,500,000 percent hours.
The MAX-250 is a galvanic, partial pressure sensor that is specific 

to oxygen. It consists of two electrodes (a cathode and an anode), 

a teflon membrane and an electrolyte. Oxygen diffuses through the 

teflon membrane and immediately reacts electrochemically at a gold 

cathode.  Concurrently, oxidation occurs electrochemically at a lead 

anode, generating an electrical current and providing a voltage output.  

Electrodes are immersed in a unique gelled weak acid electrolyte 

which is responsible for the sensors long life and motion insensitive 

characteristic.  Since the sensor is specific to oxygen, the current 

generated is proportional to the amount of oxygen present in the sample 

gas. When no oxygen is present, there is no electrochemical reaction 

and therefore, negligible current is produced. In this sense, the sensor is 

self-zeroing.

CAUTION: The Maxtec MAX-250 oxygen sensor is a sealed device 

containing a mild acid electrolyte, lead (Pb), and lead acetate. Lead 

and lead acetate are hazardous waste constituents and should be 

disposed of properly, or returned to Maxtec for proper disposal or 

recovery.
CAUTION: Do not use ethylene oxide for sterilization. Do not 

immerse the sensor in any cleaning solution, autoclave or expose 

the sensor to high temperatures.
CAUTION: Dropping or severely jarring the sensor after calibration 

may shift the calibration point enough to require recalibration.

 2.0 SET-UP PROCEDURE

2.1  Battery Installation

All MAXO

2

®

 analyzer units are powered by two, AA alkaline batteries (2 

x 1.5 Volts) and are shipped without the batteries installed. The battery 

compartment is accessible from the back side of the unit. To install the 

batteries:
1. With the thumb, press down on the center of the battery compartment 

cover and slide the cover off of the instrument case.

2. Install the two, AA, alkaline batteries (2 x 1.5 Volts) in the unit, observing 

the orientation shown on the label inside the compartment.

Содержание MaxO2 OM-25A

Страница 1: ...MaxO2 O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E O M 2 5 A R213M15 Rev P...

Страница 2: ......

Страница 3: ...l of other components WARRANTY The MAXO2 Analyzer is designed for medical oxygen delivery equipment and systems Under normal operating conditions Maxtec warrants the MAXO2 Analyzer to be free from def...

Страница 4: ...become thoroughly familiar with the information contained in this Operation Manual Strict adherance to the operating instructions is necessary for safe effective product performance This product will...

Страница 5: ...3 2 2 2 To Calibrate the MAXO2 Analyzer 3 2 2 3 Automatic Calibration to Room Air 5 2 2 4 Factors Influencing Calibration 5 3 0 OPERATION INSTRUCTIONS 6 4 0 SENSOR REMOVAL AND REPLACEMENT 6 5 0 PROBL...

Страница 6: ...s Low Battery Use by prescription only Meets ETL standards Do not throw away Follow local guidelines for disposal RXonly EC REP Authorized Representative in the European Community LOT SN Lot code Batc...

Страница 7: ...f analog and microprocessor circuitry Low battery indication 1 2 Components Description please refer to page IV LCD Readout 1 3 1 2 Digit Display The 3 1 2 digit liquid crystal display LCD provides di...

Страница 8: ...ent there is no electrochemical reaction and therefore negligible current is produced In this sense the sensor is self zeroing CAUTION The Maxtec MAX 250 oxygen sensor is a sealed device containing a...

Страница 9: ...phic elevation at which the product is being used changes by more than 500 feet In addition calibration is recommended if the user is unsure when the last calibration procedure was performed or if the...

Страница 10: ...etween key actuations the system will store the latest calibration value and will revert to normal operation If this occurs inadvertently simply repeat the calibration procedure 8 When the calibration...

Страница 11: ...ion 2 2 4 Factors Influencing Calibration The primary factors influencing the MAXO2 analyzer are temperature pressure and humidity Effects of Temperature The MAXO2 analyzer will hold calibration and r...

Страница 12: ...ous readings and slower response time For this reason avoid usage in environments greater than 95 relative humidity 3 0 OPERATING INSTRUCTIONS To check the oxygen concentration of a sample gas after t...

Страница 13: ...bottom of the case by turning the barb clockwise with a 7 16 wrench The barb fitting needs to be tightened firmly against the bottom of the sensor 8 Slide the battery compartment cover back onto the...

Страница 14: ...at 23 C Warm up Time none required Operating Temperature 15 C 40 C 59 F 104 F Storage Temperature 15 C 50 C 5 F 122 F Atmospheric Pressure 800 1013 mBars Humidity 0 95 non condensing Power Requiremen...

Страница 15: ...essure the greater the difference in sensor signal oxygen reading on the analyzer By the same token if an analyzer is calibrated on a pressurized gas sample containing a known concentration of oxygen...

Страница 16: ...le gas stream For example if the MAXO2 analyzer is calibrated in dry gas and then the gas is humidified the analyzer will correctly display a reading which is slightly lower than previously displayed...

Отзывы: