sensor in that back pressure at the sensing point may change. For 

these reasons, the following is recommended:
1. Calibrate the MAXO

2

®

 analyzer at the same pressure as the 

sample gas.

2. If sample gases flow through tubing, use the same apparatus and 

flow rates when calibrating as when measuring.

3. The MAXO

2

®

 analyzer oxygen sensor has been validated at 

pressures up to 2 atmospheres absolute. Calibration or operation 

above this pressure is beyond the intended use.

Humidity Effect

Humidity has no effect on the performance of the MAXO

2

®

 analyzer 

other than diluting the gas, as long as there is no condensation. 

Depending on the humidity, the gas may be diluted by as much as 

4%, which proportionally reduces the oxygen concentration from the 

dry concentration. Environments where condensation may occur 

are to be avoided since condensate may obstruct passage of gas 

to the sensing surface, resulting in erroneous readings and slower 

response time. For this reason avoid usage in environments greater 

than 95% relative humidity.

 3.0

 

OPERATING INSTRUCTIONS

To check the oxygen concentration of a sample gas: (after the unit has 

been calibrated)
1. Connect the sample hose to the barbed fitting on the bottom of the unit.
2. Attach the other end of the sample hose to the sample gas source and 

initiate flow of the sample to the unit at a rate of 1-5 liters per minute (2 

liters per minute is recommended).

3. Using the ON/OFF key 

, make sure the unit is in the power on 

mode.

4. Allow the oxygen reading to stabilize. This will normally take about 30 

seconds or more.

 4.0

 

SENSOR REMOVAL AND REPLACEMENT

The OM-25A is shipped with a new MAX-250 oxygen sensor installed. 

Although the sensor has a very long expected life, eventually the sensor 

will require replacement. Removing or installing a sensor, when necessary, 

is a very simple procedure. To remove and install a new

sensor:
1. With the front of the analyzer facing downward and the bottom facing 

away from you, press down with both thumbs on the center of the battery 

compartment cover and slide the cover off.

Содержание MaxO2 OM-25A

Страница 1: ...MaxO2 O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E O M 2 5 A R213M15 Rev P...

Страница 2: ......

Страница 3: ...l of other components WARRANTY The MAXO2 Analyzer is designed for medical oxygen delivery equipment and systems Under normal operating conditions Maxtec warrants the MAXO2 Analyzer to be free from def...

Страница 4: ...become thoroughly familiar with the information contained in this Operation Manual Strict adherance to the operating instructions is necessary for safe effective product performance This product will...

Страница 5: ...3 2 2 2 To Calibrate the MAXO2 Analyzer 3 2 2 3 Automatic Calibration to Room Air 5 2 2 4 Factors Influencing Calibration 5 3 0 OPERATION INSTRUCTIONS 6 4 0 SENSOR REMOVAL AND REPLACEMENT 6 5 0 PROBL...

Страница 6: ...s Low Battery Use by prescription only Meets ETL standards Do not throw away Follow local guidelines for disposal RXonly EC REP Authorized Representative in the European Community LOT SN Lot code Batc...

Страница 7: ...f analog and microprocessor circuitry Low battery indication 1 2 Components Description please refer to page IV LCD Readout 1 3 1 2 Digit Display The 3 1 2 digit liquid crystal display LCD provides di...

Страница 8: ...ent there is no electrochemical reaction and therefore negligible current is produced In this sense the sensor is self zeroing CAUTION The Maxtec MAX 250 oxygen sensor is a sealed device containing a...

Страница 9: ...phic elevation at which the product is being used changes by more than 500 feet In addition calibration is recommended if the user is unsure when the last calibration procedure was performed or if the...

Страница 10: ...etween key actuations the system will store the latest calibration value and will revert to normal operation If this occurs inadvertently simply repeat the calibration procedure 8 When the calibration...

Страница 11: ...ion 2 2 4 Factors Influencing Calibration The primary factors influencing the MAXO2 analyzer are temperature pressure and humidity Effects of Temperature The MAXO2 analyzer will hold calibration and r...

Страница 12: ...ous readings and slower response time For this reason avoid usage in environments greater than 95 relative humidity 3 0 OPERATING INSTRUCTIONS To check the oxygen concentration of a sample gas after t...

Страница 13: ...bottom of the case by turning the barb clockwise with a 7 16 wrench The barb fitting needs to be tightened firmly against the bottom of the sensor 8 Slide the battery compartment cover back onto the...

Страница 14: ...at 23 C Warm up Time none required Operating Temperature 15 C 40 C 59 F 104 F Storage Temperature 15 C 50 C 5 F 122 F Atmospheric Pressure 800 1013 mBars Humidity 0 95 non condensing Power Requiremen...

Страница 15: ...essure the greater the difference in sensor signal oxygen reading on the analyzer By the same token if an analyzer is calibrated on a pressurized gas sample containing a known concentration of oxygen...

Страница 16: ...le gas stream For example if the MAXO2 analyzer is calibrated in dry gas and then the gas is humidified the analyzer will correctly display a reading which is slightly lower than previously displayed...

Отзывы: