Maxtec MaxBlend 2 Скачать руководство пользователя страница 8

ENGLISH 

8

 

WWW.MAXTEC.COM  •  (800) 748-5355

2.  Should oxygen level exceed either the HIGH or LOW alarm set points, the alarm 

indicator on the front panel will illuminate indicating either a high or low oxygen 
condition. The alarm will cease when oxygen levels are within limits, or the limits are 
adjusted.

NOTE:

 The MaxBlend 2 will monitor the oxygen concentration of the gas delivered from the 

blender regardless of which outlet port is in use. Therefore, the integrated flowmeter has ulti-
mately no effect on the oxygen sensing.

2.3 Alarm Setting Procedure
2.3.1 Low Alarm Setting 

To adjust the low alarm setting:

1.  Press the Unlock key 

 to unlock the keypad. Note the LOW, SMART ALARM, CAL, and 

HIGH icons will begin to flash indicating the SET OPERATING MODE.

2.  Press the DOWN (Low Alarm) key 

 on the keypad.

NOTE: 

The Low Alarm digits begin to flash indicating the Low Alarm manual setting.

3.  Use the UP 

 and DOWN 

 keys to set the low alarm to the desired value. 

Pressing the arrow keys changes the value in 1% increments. If the keys are held down 
for more than 1 second, the display will scroll at a rate of 1% per second.

NOTE: 

If 30 seconds elapse between key actuations, the system will store the latest low alarm 

setting and will revert to normal operation. If this occurs inadvertently, simply repeat the 
alarm setting procedure.
There is a special condition that allows the low oxygen alarm to be set below 18%. To access 
this condition, press the DOWN arrow key for three seconds while the low alarm reading dis-
plays 18%. The alarm setting can now be adjusted to 17, 16, or 15%. A bar will blink above the 
setting to provide further indication that the alarm has been set to this special <18% condition.
The low alarm value cannot be set lower than 15%, nor can it be set closer than 1% from the 
high alarm value. For example, if the high alarm is set at 25%, the system will not accept a low 
alarm setting greater than 24%.
When the low alarm value is set, press the Unlock key again to accept the low alarm setting 
and return to normal operation.

NOTE: 

The default low alarm setting is 18% O2. Removing the batteries, or shutting the unit OFF 

will reset the low alarm limit to 18%.

2.3.2 High Alarm Setting

To adjust the high alarm setting:

1.  Press the Unlock key to unlock the keypad. Note the LOW, SMART ALARM, CAL and HIGH 

icons will begin to flash indicating the SET OPERATING MODE.

2.  Press the UP (High Alarm) key on the keypad.

NOTE: 

The High Alarm digits begin to flash indicating the High Alarm manual setting.

3.  Use the UP and DOWN keys to set the high alarm to the desired value.  Pressing the 

arrow keys changes the value   in 1% increments. If the keys are held down for more 
than 1 second the display will scroll at a rate of 1% per second.

NOTE: 

If 30 seconds elapse between key actuations, the system will store the latest high alarm 

setting and will revert to normal operation. If this occurs inadvertently, simply repeat the 
alarm setting procedure.

4.  When the high alarm setting is set above 100%, the high alarm will indicate two dashes 

– –. This special condition turns off, or deactivates, the high alarm.

5.  When the high alarm value is set, press the Unlock again to accept the high alarm 

setting and return to normal operation.

NOTE: 

The default high alarm setting is 50% O2. Removing the batteries will reset the high 

alarm limit to 50%.

2.3.3 Smart Alarm Mode

1.  Press the Unlock Key 

 to unlock the keypad. Note the LOW, SMART ALARM, CAL, and 

HIGH icons will begin to flash indicating the SET OPERATING MODE.

2.  Press the Smart Alarm key 

 on the keypad. Note the LOW digits, Alarm Mode and 

HIGH digits begin a slow flash indicating SMART ALARM MODE. The high alarm will now 
be set equal to the current oxygen r3% (rounded to the nearest integer). The 
low alarm will now be set equal to the current oxygen reading -3% (rounded to the 
nearest integer, but never lower than 18%.

3.  Pressing of the UP key 

 will add 1% to the high alarm setting and subtract 1% 

from the low alarm setting. Pressing the DOWN key 

 will subtract 1% from the 

high alarm setting and add 1% to the low alarm setting. In other words, the UP key 
widens the alarm band and the DOWN key tightens the alarm band. This feature will 
not set the alarm levels above 100% or below 18%.

4.  Once the desired alarm settings are attained, press the Unlock key 

 again to save 

the settings and return to normal operation mode. If 30 seconds elapse without a 
key press by the user, the device will automatically save the latest alarm settings and 
return to normal operation mode.

2.4 Basic Operation

To check the oxygen concentration of the delivered gas:

1.  Ensure the sensor is connected to the flow diverter and inserted completely in the 

sensor port behind the flowmeter.

2.  Set the oxygen blender control knob to the desired oxygen setting.
3.  Using the ON/OFF key 

, make sure the unit is in the power on mode.

4.  Allow the oxygen reading to stabilize.
5.  Adjust the flowmeter to the desired flow rate. Read the flow rate at the center of the float ball.

2.5 Alarm Conditions and Priorities

In the event of either a low alarm or high alarm condition, the corresponding LED will begin 
to flash, accompanied by the audio buzzer. Pressing the Silent key 

 will deactivate the 

buzzer, but the LED and the alarm value digits on the display will continue to flash until the 
alarm condition has been rectified. If the alarm condition still exists 120 seconds after silencing 
the audio buzzer, the beeper will start to sound again.

ALARM

ALARM 

PRIORITY

LOW 

ALARM LED

HIGH 

ALARM LED

AUDIBLE 

ALARM

AUDIBLE 

ALARM REPEAT

Line Power 

Plugged In

Informational

Off

Off

2 Pulses

No Repeat

Line Power 

Unplugged

Informational

Single 

Yellow Pulse

Single 

Yellow Pulse

2 Pulses

No Repeat

External DC 

Power Supply 

Voltage Out 

of Range

Informational

Solid Yellow

Solid Yellow

2 Pulses

Every 15 Sec.

Battery Voltage 

too low for 

device to 

operate (E04)

Medium

Pulsing 

Yellow

Pulsing 

Yellow

3 Pulses

Every 25 Sec.

Oxygen/

pressure level 

above the high 

alarm setting

Medium

Off

Pulsing 

Yellow

3 Pulses

Every 25 Sec.

Oxygen/pressure 

level below the 

low alarm setting

Medium

Pulsing 

Yellow

Off

3 Pulses

Every 25 Sec.

Oxygen level 

below the low 

oxygen alarm 

setting and 

lower than 18%

High

Pulsing Red

Off

5+5 Pulses

Every 15 Sec.

Содержание MaxBlend 2

Страница 1: ...60 50 40 30 21 70 80 90 100 O2 Instructions for Use E N G L I S H MaxBlend 2 L O W F L O W H I G H F L O W R229M01 REV H ...

Страница 2: ...uency energy If not installed and used in accordance with the instructions in this manual electromagnetic interference may result The equipment has been tested and found to comply with the limits set forth in IEC 60601 1 2 for medical products These limits provide reasonable protection against electromagnetic interference when operated in the intended use environments described in this manual MRI ...

Страница 3: ...xBlend 2 with clean dry medical grade gases Contaminants or moisture can cause defective operation Oxygen should have a minimum dewpoint of 80 F 62 C or moisture content less than 7 9 PPM 0 0059mg L Oxygen purity should be at least 99 0 and the air used should be medical grade Water vapor content must not exceed a dew point of 5 F 15 C below the lowest ambient temperature to which the delivery sys...

Страница 4: ...al Power Supply Operation 9 2 8 Calibration Procedures 9 2 8 1 Calibration to 100 Oxygen 9 2 8 2 Calibration to Room Air 9 2 9 Flowmeter Operation 9 3 0 PERFORMANCE CHECK 9 3 1 Blender Safety Check 10 4 0 TROUBLESHOOTING 10 5 0 CLEANING AND DISINFECTING 11 6 0 SERVICE AND MAINTENANCE 11 6 1 Maintenance 11 6 2 Replacing O2 Sensor 11 6 3 Monitor Alarm Testing 11 7 0 ABBREVIATION GUIDE 12 8 0 SPECIFI...

Страница 5: ...c to oxygen the current generated is proportional to the amount of oxygen present in the sample gas When no oxygen is present there is no electrochemical reaction andtherefore negligiblecurrentisproduced Inthissense thesensorisself zeroing CAUTION The MAX 550E oxygen sensor is a sealed device containing a mild acid electro lyte and lead Pb These materials are hazardous waste constituents and shoul...

Страница 6: ... is used to help set the High Low Alarm win dow quickly See section 2 3 3 for instructions on using the Smart Alarm setting y DOWN LOW ALARM KEY The down key is used in setting the low FiO2 alarm limit The device must be in the unlocked state for the key to operate See section 2 3 1 for instructions on setting the low FiO2 alarm limit u BACKLIGHTKEY Thebacklightkeywillmanuallyactivatethebacklightf...

Страница 7: ...ES 2 1 Setup and Installation 2 1 1 Battery Installation AllMaxBlend2unitsarepoweredbyfour AA alkalinebatteries 4x1 5Volts andareshipped without the batteries installed Only name brand batteries should be used Batteries should be replaced by trained service personnel To install the batteries Open the battery drawer by squeezing inward on both tabs as shown in the figure below If you have difficult...

Страница 8: ...indicate two dashes This special condition turns off or deactivates the high alarm 5 When the high alarm value is set press the Unlock again to accept the high alarm setting and return to normal operation NOTE The default high alarm setting is 50 O2 Removing the batteries will reset the high alarm limit to 50 2 3 3 Smart Alarm Mode 1 Press the Unlock Key to unlock the keypad Note the LOW SMART ALA...

Страница 9: ...s for the reading to stabilize 4 Press the Unlock key to unlock the keypad Note the LOW SMART ALARM CAL and HIGH icons will begin to flash indicating the SET OPERATING MODE 5 Press the CAL Calibration key on the keypad The word CAL will appear on the display for approximately 5 seconds and then finish with 100 0 6 The unit is now calibrated and in the normal operating mode 2 8 2 Calibration to Roo...

Страница 10: ...ed alarm cap damaged or defective Contact Maxtec for repair PROBLEM Selected oxygen concentration accurate only when gas pressures are equal POTENTIAL CAUSES AND SOLUTIONS MaxBlend 2 balance module not functioning properly Contact Maxtec for repair PROBLEM Blank display POTENTIAL CAUSES AND SOLUTIONS Batteries not installed Install batteries Refer to section 2 1 1 Battery Installation Battery comp...

Страница 11: ... the device and its accessories Be sure to closely inspect and remove contamination from seams and recesses on the device that may trap contaminants Wipe with clean paper towel to remove debris and bioburden 3 After all visible contamination is removed use a second germicidal wipe to thoroughly wet the surfaces of the device and accessories Allow to remain wet for 4 minutes Use additional wipes if...

Страница 12: ... High Oxygen Alarm Range 16 100 1 higher than low alarm according to IEC 60601 1 8 Audible Alarms in Medical Equipment 8 3 O2 Sensor Specifications Total accuracy 3 actual oxygen level over full operating temperature range Oxygen Measurement Accuracy 1 oxygen Linearity 1 at constant temperature and pressure Error Over Operating Temp Range 3 oxygen maximum Response Time to 90 of Final Reading 77 F ...

Страница 13: ...everse gas flow from either the air or the oxygen supply systems Thetwogasesthenpassthroughatwo stagebalanceregulator Thepurposeofthisregulator is to equalize the operating pressures of the air and oxygen gas sources Once these pres sures have been balanced the gases are proportioned according to the oxygen concentration selected on the oxygen concentration selection knob The oxygen concentration ...

Страница 14: ...High Flow Blender 0 5 10 15 20 25 0 5 10 15 20 25 30 35 Pressure drop PSIG Flow Rate LPM 75 psig 50 psig 30 psig For Low Flow Blender 13 0 ELECTROMAGNETIC COMPATABILITY The information contained in this section such as separation distances is in general specifi cally written with regard to the MaxBlend 2 The numbers provided will not guarantee fault less operation but should provide reasonable ass...

Страница 15: ...er manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propogation is affected by absorption and reflection from structures objects and people This equipment is intended for use in the electromagnetic environment specified below The customer or the user of this equipment sho...

Страница 16: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com ...

Отзывы: