Maxtec MaxBlend 2 Скачать руководство пользователя страница 3

WWW.MAXTEC.COM  •  (800) 748-5355 

3

 ENGLISH

 

If a condition is detected that could possibly prevent the monitor from continuing to oper-

ate safely, it will sound an alarm. If, at any time, EOx (i.e. EO2, EO4, etc.) appears on the LCD 

refer to section 4.0, or contact a Maxtec Certified Service Technician.

 

All service should be referred to a Maxtec Certified Service Technician.

 

Maxtec recommends that the MaxBlend 2 be serviced by Maxtec at a minimum of every 

three years, or if a leak or other performance problem is suspected.

 

If the MaxBlend 2 is dropped, follow the procedures outlined in section 3.0 for a perfor-

mance check before reusing the device.

 

When the unit is going to be stored (not in use for more than 30 days), always remove the 

batteries to protect the unit from potential damage due to battery leakage.

 

Always replace batteries with high-quality, name-brand  “AA” alkaline batteries.

 

The low oxygen alarm can be set below 18%,  and as low as 15%. See section 2.3 on setting 

alarms. This is in accordance with IEC 80601-2-55.

 

In the event of exposure to an ELECTROMAGNETIC DISTURBANCE the analyzer may display 

an E06 or E02 error message. If this occurs, refer to Section 4.0 for instructions to resolve 

the problem.

To prevent risk of burns, fire, or injury to person(s):

 

Mixed gases continuously bleed into the atmosphere at the oxygen concentration rate 

set at the control knob. Bleeding oxygen into any closed area could increase the risk of 

fire or explosion.

 DO NOT 

operate this device in the presence of any flame or source of ignition, or when 

using devices such electrosurgical equipment or defibrillators.

 DO NOT 

operate the MaxBlend 2 in the presence of flammable anesthetics or in an atmo-

sphere of explosive gases. Operating the MaxBlend 2 in flammable or explosive atmo-

spheres may result in fire or explosion.

 

Galvanic O2 sensor electrolyte gel is acidic and may cause skin or eye irritation and/or 

burns. Take care when handling or replacing exhausted or damaged disposable O2 sen-

sors. Be sure to dispose of expired sensors in accordance with hospital and/or govern-

mental regulations (O2 Sensor SDS available upon request from Maxtec).

 DO NOT 

smoke in an area where oxygen is being used.

CAUTIONS 

Indicates a potentially hazardous situation which, if not avoided, could result in minor or mod-

erate injury and property damage.

 DO NOT 

store the MaxBlend 2 in hot areas for prolonged periods of time. Temperatures 

above 80°F (27°C) can shorten battery life.

 

To minimize the potential for electrostatic shock,   

DO NOT

 use antistatic or electrically 

conductive hoses with the MaxBlend 2.

When cleaning the MaxBlend 2: 

 DO NOT 

use harsh abrasives.

 DO NOT 

immerse the MaxBlend 2 in liquid sterilizing agents or liquids of any kind.

 DO NOT 

spray cleaning solution directly onto the sensor port, bleed muffler, or buzzer 

opening.

 DO NOT 

allow cleaning solution to pool on the front panel, sensor port, or bleed muffler.

 DO NOT 

sterilize the MaxBlend 2. Standard sterilization techniques may damage the 

blender.

 DO NOT 

clean or dry the MaxBlend 2 with a high pressure air gun. Applying high pressure 

air to the MaxBlend 2 may damage components and render the system inoperable.

 DO NOT 

over-clean the MaxBlend 2. Repeated use of a cleaning agent can cause residue 

build-up on critical components. Excessive residue buildup can affect the MaxBlend 2’s 

performance.

 

Dropping or severely jarring the sensor after calibration may shift the calibration point 

enough to require recalibration.

 

Always operate the MaxBlend 2 with clean, dry medical-grade gases. Contaminants 

or moisture can cause defective operation. Oxygen should have a minimum dewpoint 

of -80°F (-62°C) or moisture content less than 7.9 PPM (0.0059mg/L). Oxygen “purity” 

should be at least 99.0% and the air used should be medical-grade. Water vapor content 

must not exceed a dew point of 5°F (-15°C) below the lowest ambient temperature to 

which the delivery system is exposed.

 

Particulate content must not exceed that which would be found immediately down-

stream of a 15 micron absolute filter. Refer to CGA commodity specifications G-4.3 and G7.1 

for more information. Water vapor content of medical air or O2 supply to the blender must 

not exceed 5.63 x 103 milligrams H2O per cubic meter of non-condensable gas.

 DO NOT 

disassemble the MaxBlend 2. All service should be performed by a Maxtec 

Certified Service Technician.

 

Be sure the MaxBlend 2 is securely mounted. This device is usually mounted to a hospital 

rail system or an infusion stand. Dropping the device may cause injury or damage to the 

device.

 

The outlets of this device are capable of delivering pressure as high as the inlet pressure. 

Ensure that the devices that carry the gas from the blender to the patient prevent exces-

sive pressure to the patient.

 DO NOT 

use the oxygen sensor if it appears to be damaged or is leaking. The oxygen 

sensors contain a weak acidic solution encapsulated in a plastic housing. Under normal 

operating conditions the solution (electrolyte) is never exposed.

NOTES

Indicates supplemental information to assist in the use of the device.

 

The MaxBlend 2 is tested for compliance with ISO 11195, and meets requirements regarding 

reverse gas flow as delivered.

 

Applicable parts used in the MaxBlend 2 have been cleaned and degreased for oxygen 

service. Any lubricants used are designed specifically for the application.

 

As long as the absolute pressure of the gas mixture being monitored is constant, the 

MaxBlend 2 will accurately read oxygen concentrations. However, if the absolute pres-

sure varies the reading will fluctuate proportionately, as the sensor actually measures the 

partial pressure of oxygen in the mixture. The sensor readings will also change propor-

tionately with barometric pressure changes; because of this, frequent calibration of the 

sensor is recommended.

 

Users are advised to use pressure regulators which display the outlet pressure.

 

All specifications assume the following standard environmental conditions, unless speci-

fied otherwise: ambient and sample gas temperatures of 77°F (25°C), barometric pressure 

of 30inHg (760mmHg), sea level altitude, ambient relative humidity of 50%, sample gas 

relative humidity of 0%.

 

It is important to note that the oxygen concentration selection scale is provided only as a 

guideline for selecting O2 concentrations. The clinician should use the MaxBlend 2’s dis-

play panel to adjust the O2 concentration to the desired setting.

 

The alarm limits can be set to levels that would render them useless for a particular 

patient’s clinical condition. Ensure that the delivered oxygen level and flow rate are set 

to values prescribed by the patient’s physician. Also ensure that the high and low alarm 

limits are set to levels such that they will sound if the oxygen level is outside of safe limits. 

Be sure to review and, if necessary, re-set the alarm limits when the patient’s clinical con-

dition changes, or when the patient’s physician prescribes a change in oxygen therapy.

 

This device does not contain automatic barometric pressure compensation.

 

Gas leaks that cause room air to mix with the gas sample may cause inaccurate oxygen 

readings. Ensure the O-rings on the sensor and flow diverter are in place and intact prior 

to use.

Содержание MaxBlend 2

Страница 1: ...60 50 40 30 21 70 80 90 100 O2 Instructions for Use E N G L I S H MaxBlend 2 L O W F L O W H I G H F L O W R229M01 REV H ...

Страница 2: ...uency energy If not installed and used in accordance with the instructions in this manual electromagnetic interference may result The equipment has been tested and found to comply with the limits set forth in IEC 60601 1 2 for medical products These limits provide reasonable protection against electromagnetic interference when operated in the intended use environments described in this manual MRI ...

Страница 3: ...xBlend 2 with clean dry medical grade gases Contaminants or moisture can cause defective operation Oxygen should have a minimum dewpoint of 80 F 62 C or moisture content less than 7 9 PPM 0 0059mg L Oxygen purity should be at least 99 0 and the air used should be medical grade Water vapor content must not exceed a dew point of 5 F 15 C below the lowest ambient temperature to which the delivery sys...

Страница 4: ...al Power Supply Operation 9 2 8 Calibration Procedures 9 2 8 1 Calibration to 100 Oxygen 9 2 8 2 Calibration to Room Air 9 2 9 Flowmeter Operation 9 3 0 PERFORMANCE CHECK 9 3 1 Blender Safety Check 10 4 0 TROUBLESHOOTING 10 5 0 CLEANING AND DISINFECTING 11 6 0 SERVICE AND MAINTENANCE 11 6 1 Maintenance 11 6 2 Replacing O2 Sensor 11 6 3 Monitor Alarm Testing 11 7 0 ABBREVIATION GUIDE 12 8 0 SPECIFI...

Страница 5: ...c to oxygen the current generated is proportional to the amount of oxygen present in the sample gas When no oxygen is present there is no electrochemical reaction andtherefore negligiblecurrentisproduced Inthissense thesensorisself zeroing CAUTION The MAX 550E oxygen sensor is a sealed device containing a mild acid electro lyte and lead Pb These materials are hazardous waste constituents and shoul...

Страница 6: ... is used to help set the High Low Alarm win dow quickly See section 2 3 3 for instructions on using the Smart Alarm setting y DOWN LOW ALARM KEY The down key is used in setting the low FiO2 alarm limit The device must be in the unlocked state for the key to operate See section 2 3 1 for instructions on setting the low FiO2 alarm limit u BACKLIGHTKEY Thebacklightkeywillmanuallyactivatethebacklightf...

Страница 7: ...ES 2 1 Setup and Installation 2 1 1 Battery Installation AllMaxBlend2unitsarepoweredbyfour AA alkalinebatteries 4x1 5Volts andareshipped without the batteries installed Only name brand batteries should be used Batteries should be replaced by trained service personnel To install the batteries Open the battery drawer by squeezing inward on both tabs as shown in the figure below If you have difficult...

Страница 8: ...indicate two dashes This special condition turns off or deactivates the high alarm 5 When the high alarm value is set press the Unlock again to accept the high alarm setting and return to normal operation NOTE The default high alarm setting is 50 O2 Removing the batteries will reset the high alarm limit to 50 2 3 3 Smart Alarm Mode 1 Press the Unlock Key to unlock the keypad Note the LOW SMART ALA...

Страница 9: ...s for the reading to stabilize 4 Press the Unlock key to unlock the keypad Note the LOW SMART ALARM CAL and HIGH icons will begin to flash indicating the SET OPERATING MODE 5 Press the CAL Calibration key on the keypad The word CAL will appear on the display for approximately 5 seconds and then finish with 100 0 6 The unit is now calibrated and in the normal operating mode 2 8 2 Calibration to Roo...

Страница 10: ...ed alarm cap damaged or defective Contact Maxtec for repair PROBLEM Selected oxygen concentration accurate only when gas pressures are equal POTENTIAL CAUSES AND SOLUTIONS MaxBlend 2 balance module not functioning properly Contact Maxtec for repair PROBLEM Blank display POTENTIAL CAUSES AND SOLUTIONS Batteries not installed Install batteries Refer to section 2 1 1 Battery Installation Battery comp...

Страница 11: ... the device and its accessories Be sure to closely inspect and remove contamination from seams and recesses on the device that may trap contaminants Wipe with clean paper towel to remove debris and bioburden 3 After all visible contamination is removed use a second germicidal wipe to thoroughly wet the surfaces of the device and accessories Allow to remain wet for 4 minutes Use additional wipes if...

Страница 12: ... High Oxygen Alarm Range 16 100 1 higher than low alarm according to IEC 60601 1 8 Audible Alarms in Medical Equipment 8 3 O2 Sensor Specifications Total accuracy 3 actual oxygen level over full operating temperature range Oxygen Measurement Accuracy 1 oxygen Linearity 1 at constant temperature and pressure Error Over Operating Temp Range 3 oxygen maximum Response Time to 90 of Final Reading 77 F ...

Страница 13: ...everse gas flow from either the air or the oxygen supply systems Thetwogasesthenpassthroughatwo stagebalanceregulator Thepurposeofthisregulator is to equalize the operating pressures of the air and oxygen gas sources Once these pres sures have been balanced the gases are proportioned according to the oxygen concentration selected on the oxygen concentration selection knob The oxygen concentration ...

Страница 14: ...High Flow Blender 0 5 10 15 20 25 0 5 10 15 20 25 30 35 Pressure drop PSIG Flow Rate LPM 75 psig 50 psig 30 psig For Low Flow Blender 13 0 ELECTROMAGNETIC COMPATABILITY The information contained in this section such as separation distances is in general specifi cally written with regard to the MaxBlend 2 The numbers provided will not guarantee fault less operation but should provide reasonable ass...

Страница 15: ...er manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propogation is affected by absorption and reflection from structures objects and people This equipment is intended for use in the electromagnetic environment specified below The customer or the user of this equipment sho...

Страница 16: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com ...

Отзывы: