![MaxQ A12X24-PR Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/maxq/a12x24-pr/a12x24-pr_user-manual_1745473005.webp)
www.pack
MaxQ
.com
5
MaxPlus Alpha Cooler®
User Manual | SKU # A12X24-PR, A12X24-PR-C
V2.0 | March 7
th
, 2022
1.4 Integrated monitoring system
1.4.1 Wireless sensor module -
Placed inside the cooler along with the blood products for measuring payload temperature and communicating with
Main controller module.
1.4.2 Main controller module -
Removable controller with a display that can be placed in the lid of the Alpha Cooler.
1.4.3 System Specifications
1
Memory
8GB (SD card, pre-installed)
2
Program Interval
2 minutes (minimum) to 1 hour
3
Temperature Sensor
Integrated circuit temperature sensor
4
Resolution
0.1°C
5
Accuracy
±0.5°C
6
Sensor Response Time
>2s
7
Operating Temperature
-20°C to +70°C
8
Alarm
Yes (Audible)
9
Battery
2800mAh Lipo battery
10 Calibration
3 point NIST Traceable Calibration Certificate
12 Mobile Application
Connect to
“MaxTemp”
Mobile Application (FREE Download
from Apple or Google Store)
13 Desktop Application
Windows application “
MaxTemp
” (FREE download from
Microsoft app store)
13 Application Compatibility
Mobile phones running operating systems greater than Android
5.0 or iOS 11
Microsoft windows 10 operating system
14 Data / Report export
Built-in PDF report generator in the mobile applications