MaxPlus Alpha Cooler®
User Manual | SKU # A12X24-PR, A12X24-PR-C
V2.0 | March 7
th
, 2022
6.2 Setting up payload temperature range
●
Step 5:
Under this step, the user can select the acceptable lower and upper payload temperature range for the product
that will be stored inside the cooler. The standard temperature range for the cooler is 1-6
°
C for refrigerated
RBC/Plasma. However, if the cooler is being used for transportation of refrigerated RBC/Plasma, then the user can
change the temperature range to 1-10
°
C
o
Setting the lower limit: Use the ‘
UP
’ or ‘
DOWN
’ button to increase or decrease the value
o
Press ‘
SELECT
’ to accept the change and move to the next screen.
o
Setting the upper limit: Use the ‘
UP
’ button to increase the value to acceptable upper temperature value.
o
Press ‘
SELECT
’ to accept the change and move to the Date/Time Screen
6.3 Setting up Date and Time on the main controller
●
Step 6:
Inside the Date and Time setting the user can change year, month, day, hour and minute sequentially (pictures
below).
o
Use the ‘
UP
’ or ‘
DOWN
’ button to increase or decrease the value
o
Press ‘
SELECT
’ to accept the change and move to the next screen.
After the user has adjusted the ‘minute’ and accepted the change, the system will automatically go to the next screen.