Page 3 P/N 07823145 Rev D
2271PB-07 Hood
Instructions For Use
2. Either continue grasping the HLF and lift it up, over, and off the
Helmet and discard appropriately, or ALTERNATIVELY,
you may temporarily leave it on the Hood and proceed to Doff it
with the Hood (next section).
1. Begin HLF doffing by curling its back bottom up over Helmet
back bottom.
Doff HLF
Don HFR FCC (Optional, depending upon fluid resistance protection desired.)
3. Ensure HFR FCC is down to or below all
around bottom edge of Hood Filter.
1. Initiate donning the HFR FCC (High Fluid Resis-
tant Filter Cover Cap) by placing it over the front
of the Helmet and down towards the back.
2. Continue bringing the HFR
FCC down over the Helmet
so that it covers the HLF.
1. Hold front of HFR FCC in
place while lifting its rear
up over Helmet.
Doff HFR FCC (If it was Donned)
2. Remove HFR FCC com-
pletely and decon for next
use.
Don HLF (NOT optional - Required per NIOSH Approval)
3. Ensure Hood Skirt is fully down on all
sides, secure neck ties to within about 1/2
inch of neck, and feed body ties through
front hood shroud slits and tie securely.
1. Initiate donning the HLF
(Heavy Loading Filter) by
tucking its back bottom up
under back bottom of Helmet
2. Complete by pulling HLF over and down, rear-to-
front, of the Helmet and secure front bottom on
Hood front Velcro strip.
16. Don appropriate body gown per
gown manufacturer’s protocol. Keep
Hood Shroud up while closing gown
so Hood is outside of gown.
15. Pull Hood Shroud
down all around.
14. Tighten Ratchet Knob (clockwise) as tight as
possible without causing discomfort to secure
Helmet on head for all anticipated activities.