7
d. Interlock
High voltage output can be cut off with external switch off. For safety reason, once the
output is cut off, the unit shall not resume the output even turning the external switch
back on.
To reset this status, turn the Power switch on after shorting the interlock terminal.
e. External voltage control
By turning Local/Remote change switch on the front panel to REMOTE, output voltage
can be set with external voltage 0V to +10V.
f. Output current monitor
±10V output for the maximum rated output current. Output impedance is 1k ohm. The
polarity of monitor output is identical to that of high voltage output.
3.Operation Manual
Interlock action by switch
switch
Interlock action
Interlock action by open
collector
NOTE1
In place of switch, an open collector will
suffice for it.
E
C
24
25
SHORT
O P E N
…… Output ON
…… Output OFF
COM
⑮
⑭
CH②
Vcon-in
②
①
COM
CH①
Vcon-in
COM
⑤
V
COM
⑱
V
CH①
Imoni
CH②
Imoni
④
⑰