BENCH MODEL MULTIMETER
Open the tools cover on the top cover, and open the fuse cover in the tool case.
Remove the blown fuse, replace with fuse of the same size and rating. Make sure the new fuse is centered in the fuse
holder.
Replace the blown fuse with same ratings. The A input terminal is protected by a F10A/250V fast blow ceramic fuse,
Φ6×32mm. The mA input terminal is protected by a F300mA/250V fast blow fuse, Φ5×20mm..
Fuse rarely need replacement and blow almost always as a result of the operator’s error.
Never use the bench multimeter unless the fuse cover is in place and fastened fully.
Note:
To avoid electrical shock, disconnect the test leads and any input signals before replacing the fuses. Replace only with
same type of fuses. The
A
input terminal is protected by a F10A/250V fast blow ceramic fuse. The
mA
input terminal is
protected by a F300mA/250V fast blow fuse.
Replacing the Power Fuse
Use the following procedure to examine or replace the power fuses:
1. Turn the power switch to off, and remove the power cable from live power source.
2. Disconnect test leads from any live source, turn the rotary switch to off, and remove the test leads from the input terminals.
3. Replace the power fuse from the supply power inlet with fuse. Replace the blown fuse with same ratings.
4. The power fuse: 100mA / 250V, Fast, Φ5×20mm.
Note:
To avoid electrical shock, disconnect power cable from live power source , and remove the test leads and any input
signals before replacing the power fuses. Replace only with same type of fuses.