Radical Signal Extraction Pulse Oximeter Operator’s Manual
4-9
4
o p e r a t i o n
S i m p l i f i e d u s e r i n t e r f a c e
By enabling the Simplified User Interface, users are exposed to only the most common
oximeter features, while the all the remaining settings remain available behind password
protection.
FIRST PAGE
SECOND PAGE
NEXT MENU
PAGE
Press the Next Menu Page button to access the first page of
selections.
Press and hold for 6 seconds to toggle between Traditional User
Interface and the Simplified User Interface. Enter password when
prompted. The Radical will retain this setting after a power cycle.
ALARM
MENU
Press the Alarm Menu Access button to enter the alarm settings menu.
INCREASE
LOUDNESS
Press the Increase Loudness button to increase the volume of the
pulse beep.
Seven levels of volume exist.
DECREASE
LOUDNESS
Press the Decrease Loudness button to decrease the volume of the
pulse beep. The lowest volume level will silence the pulse beep.
NEXT MENU
PAGE
Press the Next Menu Page button to access the second page of
selections.
Press and hold for 6 seconds to toggle between Traditional User
Interface and the Simplified User Interface. Enter password when
prompted. The Radical will retain this setting after a power cycle.
ALARM
MENU
Press the Alarm Menu Access button to enter the alarm settings menu.
SENSITIVITY
Press the Sensitivity button to toggle between the Normal, APOD
and Maximum Sensitivity modes. Use the Normal Sensitivity setting
for typical monitoring purposes. Use the APOD setting where there
is a high probability of the sensor becoming detached. Use the
Maximum Sensitivity setting for patients with low perfusion or when
the low perfusion message is displayed on the screen in APOD or
normal sensitivity mode. The default is APOD.
Note:
When using the Maximum Sensitivity setting, the SENSOR
OFF detection performance may be compromised
Note:
In "Custom" mode (see Section 4, Operation) the unit will
remain in Normal or APOD setting after a power cycle. Maximum
Sensitivity will automatically reset to Normal Sensitivity after a power
cycle.
In "Neo" or "Adult" mode (see Section 4, Operation) the unit will reset
the sensitivity to the hospital specified setting (Normal or APOD)
after a power cycle.
ROTATE
DISPLAY
Press the Rotate Display button to reconfigure the display screen in
a vertical or horizontal format. The display rotates counter-clockwise
in 90 degree increments.
T o u c h k e y c o n t r o l b u t t o n a n d i c o n s
The touch key control buttons are the four dark grey control buttons to the right of the
Handheld display. To select a touch key icon, press and release the dark gray control
button to the right of the icon.
On the Radical display, four icons are shown on the right side or bottom of the LCD
display.
T r a d i t i o n a l u s e r i n t e r f a c e
FIRST PAGE
SECOND PAGE
NEXT MENU
PAGE
Press the Next Menu Page button to access the second page of
selections.
Press and hold for 6 seconds to toggle between Traditional User
Interface and the Simplified User Interface. Enter password when
prompted. The Radical will retain this setting after a power cycle.
TREND
DISPLAY
Press the Trend Display button to show the trend data on the display.
INCREASE
LOUDNESS
Press the Increase Loudness button to increase the volume of the
pulse beep.
Seven levels of volume exist.
DECREASE
LOUDNESS
Press the Decrease Loudness button to decrease the volume of the
pulse beep. The lowest volume level will silence the pulse beep.
NEXT MENU
PAGE
Press the Next Menu Page button to access the second page of
selections.
Press and hold for 6 seconds to toggle between Traditional User
Interface and the Simplified User Interface. Enter password when
prompted. The Radical will retain this setting after a power cycle.
MENU
ACCESS
Press the Menu Access button to enter the main menu.
SENSITIVITY
Press the Sensitivity button to toggle between the Normal, APOD and
Maximum Sensitivity modes. Use the Normal Sensitivity setting for
typical monitoring purposes. Use the APOD setting where there is a
high probability of the sensor becoming detached. Use the Maximum
Sensitivity setting for patients with low perfusion or when the low
perfusion message is displayed on the screen in APOD or normal
sensitivity mode. The default is APOD.
Note:
When using the Maximum Sensitivity setting, the SENSOR OFF
detection performance may be compromised
Note:
In "Custom" mode (see Section 4, Operation) the unit will remain
in Normal or APOD setting after a power cycle. Maximum Sensitivity
will automatically reset to Normal Sensitivity after a power cycle.
In "Neo" or "Adult" mode (see Section 4, Operation) the unit will reset
the sensitivity to the hospital specified setting (Normal or APOD) after
a power cycle.
ROTATE
Press the Rotate Display button to reconfigure the display screen in a
vertical or horizontal format. The display rotates counter-clockwise in 90
degree increments.
o p e r a t i o n
4-8
Radical Signal Extraction Pulse Oximeter Operator’s Manual
4