![Masimo Radical Signal Extraction Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/masimo/radical-signal-extraction/radical-signal-extraction_operators-manual_1735244019.webp)
Radical Signal Extraction Pulse Oximeter Operator’s Manual
4-3
4
o p e r a t i o n
S u c c e s s f u l S p O
2
m o n i t o r i n g
The following general points will aid in ensuring oximetry monitoring success.
■
Place the sensor on a site that has sufficient perfusion and provides proper alignment
of the LED’s and photodetector.
■
Place the sensor on a site that has unrestricted blood flow.
■
Do not constrict the monitoring site when securing a sensor with tape.
■
Do not select a site near potential electrical interference (electrosurgical unit, for
example).
■
Read the sensor Directions for Use for proper sensor application.
MASIMO SENSORS
Before use, carefully read the pertinent sensor Directions for Use.
Use only Masimo oximetry sensors for SpO
2
measurements.
Tissue damage can be caused by incorrect application or use of a sensor, for example by
wrapping the sensor too tightly. Inspect the sensor site as directed in the sensor Directions for
Use to ensure skin integrity and correct positioning and adhesion of the sensor.
CAUTIONS
:
■
DO NOT USE DAMAGED SENSORS. DO NOT USE A SENSOR WITH EXPOSED
OPTICAL OR ELECTRICAL COMPONENTS. DO NOT IMMERSE THE SENSOR
IN WATER, SOLVENTS, OR CLEANING SOLUTIONS (THE SENSORS AND
CONNECTORS ARE NOT WATERPROOF). DO NOT STERILIZE BY
IRRADIATION, STEAM, AUTOCLAVE OR ETHYLENE OXIDE.
SEE THE
CLEANING INSTRUCTIONS IN THE DIRECTIONS FOR USE FOR REUSABLE
MASIMO SENSORS.
■
DO NOT USE DAMAGED PATIENT CABLES. DO NOT IMMERSE THE PATIENT
CABLES IN WATER, SOLVENTS, OR CLEANING SOLUTIONS (THE PATIENT
CABLE CONNECTORS ARE NOT WATERPROOF). DO NOT STERILIZE BY
IRRADIATION, STEAM, AUTOCLAVE OR ETHYLENE OXIDE.
■
DO NOT ATTEMPT TO REPROCESS, RECONDITION OR RECYCLE MASIMO
SENSORS OR PATIENT CABLES AS THESE PROCESSES MAY DAMAGE THE
ELECTRICAL COMPONENTS, POTENTIALLY LEADING TO PATIENT HARM.
10.
Verify that the patient alarms are functional by setting the high and low SpO
2
and
pulse rate alarm limits beyond the patient readings.
■
An alarm tone sounds.
■
The violated alarm limit and reading flash on the display.
■
The red alarm indicator flashes on the Docking Station (standalone operation).
11.
Verify the sensor alarms are functional by removing the sensor from the sensor site.
■
“SENSOR OFF” appears in the message area of the graphic display.
■
The alarm tone sounds.
■
The alarm indicator flashes.
■
Disconnect the sensor from the patient cable or oximeter.
■
Confirm that “NO SENSOR” appears in the message area of the graphic display.
12.
Verify alarm silence operation.
■
Create an alarm condition by lowering the SpO
2
or pulse rate high alarm
limits beyond the patient readings.
■
Press the Alarm Silence button.
■
The alarm tone ceases for the displayed amount of time.
13.
To begin patient monitoring:
■
Adjust the alarm limits.
■
Adjust the alarm volumes.
■
Adjust the pulse beep volume.
14.
Verify the sensor is on correctly and that the measured data is appropriate, see
Section 4,
Successful SpO
2
Monitoring.
15.
Monitor the patient.
16.
After monitoring is complete, remove the sensor from the patient and store or
dispose of the sensor according to governing rules. See the Directions for Use of
the sensor.
17.
Press and hold the Power/Standby Button for 2 seconds to turn the oximeter off.
o p e r a t i o n
4-2
Radical Signal Extraction Pulse Oximeter Operator’s Manual
4