Rad-5/5v   

Chapter 4: Operation 

 

www.masimo.com 

30 

  Masimo   

 

addition, after a power off and on cycle, the sensitivity will change from the HI to the factory 

default or user configured default setting of APO or Nor. 

CAUTION:

 When using the Maximum Sensitivity setting, the performance of the SENSOR 

OFF detection may be compromised. If the device is in this setting and the sensor becomes 
dislodged from the patient, the potential for false readings may occur due to environmental 

‘noise’ such as light, vibration and excessive air movement. 

 

Alarm Limits 

WARNING:

 To ensure that alarm limits are appropriate for the patient being monitored, 

check the limits each time the pulse oximeter is used. 

On Rad-5, an audible alarm and a flashing alarm status indicator will occur when an alarm 

limit is met or exceeded for greater than five seconds. Directions for alarm suspension are 
indicated below. When a sensor is not connected to a patient, or when a sensor is not 

connected to its cable, the display will read SEn OFF or NO SEn. An audible alarm will 
accompany the display unless the oximeter has been set to Alarm Suspend Mode. To modify 

the alarm limits, see 

Menu Level 2 - Averaging, Sensitivity, FastSat, and SmartTone

 on 

page 34. 

Setting 

Range 

SpO

2

 High 

Limit 

The SpO

2

 high alarm limit can be set anywhere between 2% and 100%, with 

a 1% step size. In the “----” (off) setting, the SpO

2

 High Limit alarm is 

disabled. 

SpO

2

 Low 

Limit 

The SpO

2

 low alarm limit can be set anywhere between 1% and 100%, with 

a 1% step size. 

Note:

 The low alarm limit must always be set below the high alarm setting. 

When attempting to set the high alarm limit below the low alarm limit, the 
low alarm limit will automatically adjust to the next setting below the 

newly entered high alarm limit setting. 

Pulse Rate 
High Limit 

(BPM) 

The pulse rate high alarm limit can be set anywhere between 30 BPM and 
240 BPM, with a 5 BPM step size. 

Pulse Rate 

Low Limit 
(BPM) 

The pulse rate low alarm limit can be set anywhere between 25 BPM and 

235 BPM, with a 5 BPM step size. 

Note:

 The low alarm limit must always be set below the high alarm setting. 

When attempting to set the high alarm limit below the low alarm limit, the 
low alarm limit will automatically adjust to the next setting below the 

newly entered high alarm limit setting. 

 

 

Содержание Rad-5

Страница 1: ...Operator s Manual Rad 5 5v Pulse Oximeter...

Страница 2: ......

Страница 3: ...hysician See instructions for use for full prescribing information including indications contraindications warnings and precautions Masimo Corporation 52 Discovery Irvine CA 92618 USA Tel 949 297 7000...

Страница 4: ...www masimo com 2 Masimo 2020 Masimo Corporation...

Страница 5: ...and Cautions 12 Compliance Warnings and Cautions 13 Chapter 1 Technology Overview 15 Signal Extraction Technology SET 15 Chapter 2 Description 19 General System Description 19 Front View 19 Back View...

Страница 6: ...50 Guidance and Manufacturer s Declaration Electromagnetic Emissions 51 Guidance and Manufacturer s Declaration Electromagnetic Immunity 52 Recommended Separation Distances 54 Symbols 55 Citations 56...

Страница 7: ...s cautions and notes A warning is given when actions may result in a serious outcome for example injury serious adverse effect death to the patient or user WARNING This is an example of a warning stat...

Страница 8: ......

Страница 9: ...raging times Features that apply to the Rad 5 only will be indicated with Rad 5 Features These features are common to the Rad 5 family Clinically proven Masimo SET technology performance Applicable fo...

Страница 10: ...and motion conditions and for patients who are well or poorly perfused in hospitals hospital type facilities mobile and home environments Rad 5v The Masimo SET Rad 5v Pulse Oximeter is indicated for t...

Страница 11: ...osition that might cause it to fall on the patient Do not lift the Rad 5 5v by the patient cable WARNING Explosion hazard Do not use the Rad 5 5v in the presence of flammable anesthetics or other flam...

Страница 12: ...is to be operated by qualified personnel only This manual accessory directions for use all precautionary information and specifications should be read before use WARNING The Rad 5 is intended for con...

Страница 13: ...lied coloring and texture such as nail polish acrylic nails glitter etc Moisture birthmarks skin discoloration nail aberration deformed fingers or foreign objects in the light path Elevated levels of...

Страница 14: ...sor Directions for Use for the specified duration of patient monitoring time Note Physiological conditions that result in loss of pulsatile signal may result in no SpO2 readings Note A functional test...

Страница 15: ...and if not installed and used in accordance with the instructions may cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in a pa...

Страница 16: ......

Страница 17: ...because they cannot distinguish between the arterial and venous blood movement sometimes referred to as noise Masimo SET pulse oximetry utilizes parallel engines and adaptive filtering Adaptive filter...

Страница 18: ...actual or physiological and the speed timing and behavior of each The stability of the readings over time is affected by the averaging mode being used The longer the averaging time the more stable the...

Страница 19: ...n also be used to identify the occurrence of a patient s pulse With motion the plethysmographic waveform is often distorted and may be obscured by noise artifact Shown as a vertical line the SpO2 SIQ...

Страница 20: ......

Страница 21: ...The Rad 5 5v includes the following Rad 5 5v Device For a list of compatible sensors and cables visit http www masimo com Front View Rad 5 Rad 5v No Control Indicator Description 1 Patient Cable Conne...

Страница 22: ...onitoring 8 Power On Off Used to turn the unit on and off 9 Battery Level Indicator Indicates the status of batteries See Battery Level Indicator on page 26 10 Visual Alarm Indicator Illuminates when...

Страница 23: ...olume Cycles through three volume levels and mute At the loudest level pressing the Pulse Tone Volume button will return the volume to mute 19 Display Brightness Provides control of the front panel in...

Страница 24: ...ndicator Description 1 Serial Number Label Located inside battery compartment and on back of device 2 Agency Approvals Label Agency Approvals Label 3 Battery Cover Covers the batteries 4 Battery Cover...

Страница 25: ...talled 4 Press and hold the Power On Off Button for 2 seconds to turn the oximeter off See Front View on page 19 5 Press the Power On Off Button to turn the device on again Verify all LEDs light up an...

Страница 26: ...ving the sensor from the patient SEn OFF message appears on the display The alarm tone sounds The Visual Alarm Indicator flashes 10 Disconnect the sensor from the patient cable or device Confirm that...

Страница 27: ...On Off Button 2 The Rad 5 5v will power down and turn off LED Indicators For LED indicator descriptions and locations see Front View on page 19 To modify the brightness of the LEDs see Menu Level 4 L...

Страница 28: ...00 to 75 3 LEDs 75 to 50 2 LEDs 50 to 25 1 LED 25 to 10 1 Flashing LED with Audible Alarm 10 to 0 Normal Patient Monitoring During normal operation the Rad 5 5v Display shows oxygen saturation as SpO2...

Страница 29: ...the lowest setting the pulse tone is muted A low pitch tone indicates the highest or lowest setting has been reached In the setup menu system the Up and Down Arrow keys select among the options for ea...

Страница 30: ...tlines the default values that the Rad 5 5v revert to after a power cycle Option Default Setting Brightness display Set to pre power down setting Pulse tone volume Set to pre power down setting The fo...

Страница 31: ...Averaging Sensitivity FastSat and SmartTone on page 34 They are as follows Normal Sensitivity Nor This is the recommended mode for patients that are experiencing some compromise in blood flow or perf...

Страница 32: ...will accompany the display unless the oximeter has been set to Alarm Suspend Mode To modify the alarm limits see Menu Level 2 Averaging Sensitivity FastSat and SmartTone on page 34 Setting Range SpO2...

Страница 33: ...Rad 5 patient cable connector see Front View on page 19 and connect the DB 9 end to a COM port on the PC 3 Turn the Rad 5 on 4 Start the TrendCom Utility 5 Select the appropriate COM port number if ne...

Страница 34: ...meter Specification Date MM DD YY Time HH MM SS SpO2 001 to 100 or meaning parameter not available Pulse rate 001 to 240 or meaning parameter not available Perfusion Index 00 00 to 20 00 Exception Mes...

Страница 35: ...d 5 settings and configuration options are accessed through the menu system The Mode Enter key is used to enter the menu system and to move through the different menu levels Within each level of the s...

Страница 36: ...t buttons are pressed Use the Up Arrow key to turn on SmartTone Sensitivity 2 Hi Maximum Nor Normal APO APOD FastSat 1 On Off SmartTone 3 On Off 1 FastSat is automatically enabled in 2 and 4 second av...

Страница 37: ...own Arrow Keys to adjust parameter to desired setting Set month Set Day Set Hour Set Minute A valid date must be entered If an invalid date is entered i e February 31 the trend will not turn on and tn...

Страница 38: ...and Down keys located on the front panel of the oximeter The following sub sections describe each menu item in more detail The oximeter has options that allow user configuration to suit specific needs...

Страница 39: ...andard Mode Use Up or Down Arrow Keys to adjust parameter to desired setting Note The parameter is set selected when the Mode Enter and Next buttons are pressed Sleep Mode SLP LED Display Brightness 4...

Страница 40: ......

Страница 41: ...ufficient for manual corrective action Prompt Death or irreversible injury Having the potential for the event to develop within a period of time usually sufficient for manual corrective action Low Del...

Страница 42: ...ded except for the Sleep Mode Power On Alarms are active and Alarm Suspended Indicator is off Push Once Alarm is suspended for 120 seconds and Alarm Suspended Indicator flashes Push Twice Audible alar...

Страница 43: ...bars Pulse Search Wait for found pulse This search should occur whenever a sensor is first applied to a patient Pulse bar turns red bottom two LEDs only Low Perfusion Rule out occlusion of blood flow...

Страница 44: ...roperly applied and cover the sensor site with opaque material if required Rad 5v Messages The following messages display on Rad 5v Display Message Type Solution NO CBL No cable connected Connect cabl...

Страница 45: ...lace sensor as soon as possible NO ADH No adhesive sensor connected Connect adhesive sensor INC ADH Incompatible adhesive sensor Replace adhesive sensor RPL ADH Defective or unrecognized adhesive sens...

Страница 46: ......

Страница 47: ...ife at a faster rate than normal Continuous Speaker Tone Internal Failure Device requires service Press the Alarm Silence button If alarm continues to sound power down the device and remove batteries...

Страница 48: ......

Страница 49: ...ion Index 0 02 to 20 Accuracy ARMS Oxygen Saturation SpO2 No Motion 1 SpO2 from 70 to 100 Adults Pediatrics Infants 2 Neonates 3 Motion SpO2 from 70 to 100 Adults 2 Pediatrics 2 Infants 2 2 Neonates 3...

Страница 50: ...when compared to the reference device under a controlled study Resolution Parameter Resolution SpO2 1 PR 1 bpm Electrical Battery Type 4 AA Alkaline 5 Capacity 30 hours 6 Environmental Environmental...

Страница 51: ...oz Modes Modes Rad 5 Averaging Mode 2 4 8 10 12 or 16 seconds 7 Sensitivity Normal Maximum and APOD Rad 5v Averaging Mode 8 seconds Sensitivity Normal Alarms Alarm Details High Priority 799 Hz tone 5...

Страница 52: ...bar perfusion index bar battery status Rad 5 Only MAX FastSat Indicator Type LED Compliance EMC Compliance IEC 60601 1 2 Class B Safety Standards Compliance ANSI AAMI ES 60601 1 2005 A1 2012 CAN CSA C...

Страница 53: ...Guidance and Manufacturer s Declarations Electromagnetic Emissions The ME Equipment is intended for use in the electromagnetic environment specified below The customer or the user of the ME Equipment...

Страница 54: ...should be at least 30 Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 1 kV for input output lines Mains power quality should be that of a typical commercial or hospital envir...

Страница 55: ...601 Test Level Compliance Level Recommended Separation Distance Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 Vrms 150 kHz to 80 MHz Radiated RF IEC 61000 4 3 3 V m 150 kHz to 80MHz 3 V m 80 M...

Страница 56: ...operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the ME Equipment b Over the frequency range 150 kHz to 80 MHz field strengths shou...

Страница 57: ...symbols are found on the Rad 5 5v or packaging and are defined below Symbols Definition Symbols Definition Follow Instructions for Use Consult instructions for use Mark of Conformity to European Medic...

Страница 58: ...ive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70 100 SpO2 against a laboratory CO Oximeter and ECG monitor 3 Masimo SET technology has been valid...

Страница 59: ...5 sodium hypochlorite solution Up to 55 alcohol 0 5 quaternary ammonium chloride solution such as Ecolab Asepti Wipe II Germicidal Wipes or PDI Super Sani Cloth Germicidal Wipes Accelerated Hydrogen...

Страница 60: ...ll available LEDs are illuminated and a brief beep tone sounds 2 The oximeter begins normal operation Key Press Button Test With the monitor turned on with the exception of the Power Button press each...

Страница 61: ...Verify that an audible alarm occurs and the SpO2 measurement and the Alarm indicator are both flashing 8 Press the Alarm Silence button once and verify that the alarm is silenced and the Alarm Silenc...

Страница 62: ...e invoice or other applicable documentation Purchase order number to cover repair if the Rad 5 5v is not under warranty or for tracking purposes if it is Ship to and bill to information Contact person...

Страница 63: ...ication reproduction release performance display or disclosure of the Software including the related documentation by the U S Government or any of its agencies shall be governed solely by the terms of...

Страница 64: ...nditions or warranties regarding the Product No other warranty applies express or implied including without limitation any implied warranty of merchantability fitness for a particular purpose satisfac...

Страница 65: ...hooting 47 Chapter 7 Specifications 49 Chapter 8 Service and Maintenance 44 61 Citations 58 Cleaning 61 Cleaning and Service Warnings and Cautions 12 Compliance 52 Compliance Warnings and Cautions 13...

Страница 66: ...ng 25 28 O Operation 33 P Performance Verification 62 Performance Warnings and Cautions 10 Perfusion Index 27 Physical Characteristics 51 Power Rad 5 5v On and Off 27 Power On Self Test 62 Product Des...

Страница 67: ...9 Special Menu 39 Standard and Sleep Mode 39 Successful Monitoring for SpO2 and PR 17 25 Symbols 57 T Testing Rad 5 5v with Masimo SET Tester 63 Trend Data Format 34 Trend Settings 33 37 TrendCom Util...

Страница 68: ......

Страница 69: ......

Страница 70: ...www masimo com 33606 LAB 5724H 0920...

Отзывы: