Chapter 7: Specifications
Display Ranges and Resolution
Parameter Range
Resolution
rSO
2
0 to 99%
1%
Δ
SpO2
0 to 99%
1%
Δ
base
-100 to 890%
1%
AUC
0 to 9999
1 minute-%
Accuracy (ARMS)
Regional Hemoglobin Oxygen Saturation of Blood (rSO
2
)
rSO
2
(trending)
(from 45% to 85% SavO2)
Adult [1], Pediatric [2] 3%
rSO
2
(absolute)
(from 45% to 85% SavO2)
Adult [1]
4%
Pediatric [2]
5%
* ARMS accuracy is a statistical calculation of the difference between device measurements
and reference measurements. Approximately two-thirds of the device measurements fell
b/- ARMS of the reference measurements in a controlled study.
Note:
ARMS is indicative of use with the large adhesive O3
®
Sensor >40kg only for adults and
the pediatric O3
®
Sensor ≥5 kg to <40 kg only for pediatric patients.
[1] Absolute and trending rSO
2
(ARMS) accuracy were determined by testing on healthy adult
volunteers with light to dark pigmentation in the range of 45% to 85% SavO
2
against 30%
arterial and 70% jugular venous blood oxygen saturations, measured with a laboratory
CO-Oximeter.
[2] Absolute rSO
2
accuracy (ARMS) was determined by testing on pediatric patients ≥5 kg, <40
kg with varying skin pigmentation in the range of 45% to 85% SavO
2
against 30% arterial and
70% jugular venous blood oxygen saturations, measured with a laboratory CO-Oximeter.