O3® Regional Oximeter
Chapter 4: Operation
www.masimo.com
27
Masimo
Site Selection Menu
Note:
The O3
®
system is intended for the monitoring of rSO
2
in the cerebral region under the
sensors. Other Body Sensor Site selections are for reference only and are not intended for
patient monitoring.
When the connector end on the O3
®
Sensor is first inserted into the O3
®
Module, "1" is
defaulted to the left forehead and "2" is defaulted to the right forehead. The sensor site
selection can be modified by accessing the Site Selection menu.
There are two (2) ways to access this menu:
1.
To access this menu tap the action menu on the top right corner of the O3
®
Module window. Then press on the
Site Selection
button.
Or
2.
Press the "Gear" icon on the lower right corner of the main window of Root
®
to
access the Main Menu, then press the "O3" icon to access the O3
®
Module Menu
and select the
Site Selection
Menu.
When the Site Selection menu is accessed, the sensor site screen displays on Root
®
, as
illustrated below:
Select the
Head
or
Body
buttons at the upper left corner of the screen to switch between Head
and Body Patient View site selection menus.
Note:
Sensor icons appear white when not selected and blue (with the corresponding sensor
number) when selected.
Содержание O3 Regional Oximeter