MightySat Rx
Appendix
www.masimo.com
26
Masimo
Error or Error Message
Possible Causes
Recommended Solutions
Device keeps turning off
while on the finger
•
Incorrect finger
placement
•
Environmental
influences
•
Device damaged
•
Reposition finger (See
Basic
Setup and Use
on page 17 in
this manual)
•
Relocate device
•
Replace with new batteries
•
Contact Masimo Technical
Services
Measurement does not
display on the smart device
using optional Bluetooth
LE
•
Bluetooth LE not
connected
•
Compatible app not
installed on smart
device
•
Device damaged
•
Smart device
damaged
•
Confirm Bluetooth LE is on
for the MightySat Rx and the
smart device
•
Confirm a compatible app is
installed on the smart device
•
Close and re-launch the
compatible app on the smart
device
•
Check that MightySat Rx is
connected to the correct
smart device
•
Contact Masimo Technical
Services
Specifications
Display Ranges
Parameter
Display Ranges
SpO
2
(Oxygen Saturation) 0% to 100%
PR (Pulse Rate)
25 bpm to 240 bpm
Pi (Perfusion Index)
0.02% to 20%
PVi (Pleth Variability Index) 0% to 100%
RRp (Respiration Rate)
4 rpm to 70 rpm
The emitted wavelengths range from 600 nm to 1000 nm and the peak optical power is less
than 15 mW. Information about the wavelength range can be especially useful to clinicians.
Содержание MightySat Rx
Страница 1: ...Home Care Manual MightySat Rx Fingertip Pulse Oximeter ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 16: ......
Страница 22: ......
Страница 44: ......
Страница 51: ......
Страница 52: ...www masimo com 300162 LAB 10168A 0119 ...