Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5-4
5.8.2
Insert the non-dominant hand’s ring finger into the finger
sensor.
5.8.3
Under %SpMET, a healthy, non-smoking individual would
have a typical reading between 0-2% on the display. (These
numbers are for reference only and should not be used for
diagnostic purposes)
5.9 PVI (Pleth Variability Index) Testing (If Applicable)
Note: This is performed as a functional test only.
5.9.1
Connect a Red Masimo SET Rainbow sensor to the Radical-
7 Handheld.
5.9.2
Insert the non-dominant hand’s ring finger into the finger
sensor.
5.9.3
PVI will fluctuate depending on motion, perfusion, physiology
and other factors, so there is no ‘normal’ value. If a number
is displayed, then the PVI feature is functioning properly.
5.10 Nurse Call Test
Note: Refer to Section 8 in this manual for connector pin outs noted in
Sections 5.10
5.10.1 Disconnect the patient cable or the Masimo SET Tester from
the Radical-7 and turn the instrument on. Ensure that there
are no audible alarms and that the audible alarms are not
silenced.
5.10.2 Connect the common lead of a digital multimeter to pin 12
(Nurse Call - Common) of the Analog Output/Nurse Call
Connector (see Service Manual, Troubleshooting, Section 8,
Analog Output/Nurse Call Test) on the Radical-7. Connect
the positive lead of the multimeter to pin 6 (Nurse Call -
Normally Open) of the Analog Output/Nurse Call Connector
and measure that the resistance is greater than 1 M
Ω
(open
circuit).
5.10.3 Trigger an alarm on the monitor (e.g. by disconnecting a
sensor after it was measuring data) and verify the resistance
is < 20
Ω
.