Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
3-4
3.5.4
To remove the fuses from the fuse holder, use the edge of
the screwdriver blade to pry against the bottom of the metal
portion of the fuse.
3.5.5
Place fuses as needed (1 Amp, Metric, fast acting, 5x20mm,
250V) in the fuse holder, observing the proper orientation.
3.5.6
Slide the fuse holder back into the power entry module and
press firmly to make sure it is firmly inserted.
3.5.7
The unit is ready to be reconnected to AC power.
Note: If the fuses blow after replacement, troubleshoot the unit.
WARNING: FIRE HAZARD: TO PROTECT AGAINST FIRE HAZARD,
REPLACE ONLY WITH FUSES OF THE SAME TYPE,
CURRENT RATING, AND VOLTAGE RATING.