User Manual
P a g e
|
83
of
100
Procedure for calibration of RTD Source
To enter into the RTD Source Calibration, Select the
RTD
option in
Calibration page and SOURCE RTD option in CAL. RTD page.
For RTD source calibration first calibrate Offset for all 4 groups.
Procedure for calibration of Offset
To enter into the Offset Calibration, Select the
Offset Calibration
Radiobuttonbox in
CAL. RTD SOURCE
page.
Example: - Calibrating Offset
Offset measured value for zero and span is always 0.000mV
To calibrate
ZERO,
Select
ZERO Measured
Edit Box, when this Edit Box is selected
UC-12 will source value that is seen in
ZERO Expected
Value (here
UC-12 will source 0.000mV).For Example the external measure unit
is measuring 0.023 mV offset. Then enter 0.023 value in
ZERO
Measured
Edit Box & Press
MENU/ENT
key to calibrate the ZERO.
To calibrate
SPAN,