background image

 

User Manual

  

58 

 

40005E (Enhanced Bargraph Indicator) 

 

REF NO: m42B/om/101  

 

           

 

Issue No: 02 

 

 
 

Table 41 Holding Registers Parameters 

– Part 1 

Sr. 
No 

Parameter Description 

Modbus 

Address 

Parameter 

Type 

Min 

Value 

(Single 

Channel) 

Max 

Value 

(Single 

Channel) 

Access 

Remarks 

Ch 

– 1 – AL1 – SET 

40001 

Integer 

R/W 

Refer 

Table 44

 

Ch 

– 2 – AL1 – SET 

40002 

Integer 

R/W 

Ch 

– 1 – AL2 – SET 

40025 

Integer 

R/W 

Ch 

– 2 – AL2 – SET 

40026 

Integer 

R/W 

Ch 

– 1 – AL3 – SET 

42501 

Integer 

R/W 

Ch 

– 2 – AL3 – SET 

42502 

Integer 

R/W 

Ch 

– 1 – AL4 – SET 

42525 

Integer 

R/W 

Ch 

– 2 – AL4 – SET 

42526 

Integer 

R/W 

Ch 

– 1 - 2– ASP Hysteresis 

40049 

Integer 

250 

R/W 

10 

Ch 

– 1 – C – SET 

40061 

Integer 

R/W 

Refer 

Table 44

 

11 

Ch 

– 2 – C – SET 

40062 

Integer 

R/W 

12 

Ch 

– 1 - 2– CSP Hysteresis 

40085 

Integer 

250 

R/W 

13 

Ch

–1 - 2 – Input Type 

40097 

Integer

 

R/W 

Refer 

Table 44

 

14 

       Ch 

–1 – USER - Z    

40109 

Integer 

R/W 

15 

       Ch 

–2 – USER - Z    

40110 

Integer 

R/W 

16 

       Ch 

–1 – USER – S    

40133 

Integer 

R/W 

17 

       Ch 

–2 – USER – S    

40134 

Integer 

R/W 

18 

Ch 

– 1-2 – DP 

40157 

Integer

 

R/W 

19 

Ch 

– 1-2 – ASP1 MAP 

40169 

Integer

 

R/W 

Refer 

Table 15

 

20 

Ch 

– 1-2 – ASP2 MAP 

40181 

Integer

 

R/W 

21 

Ch 

– 1-2 – ASP3 MAP 

42549 

Integer

 

R/W 

22 

Ch 

– 1-2 – ASP4 MAP 

42561 

Integer

 

R/W 

23 

Ch 

– 1-2 – CSP MAP 

40193 

Integer

 

R/W 

 
 

 

 

Table 42 Holding Registers Parameters 

– Part 2 

 

Sr. 
No 

Parameter Description 

Modbus 

Address 

Parameter 

Type 

Min 

Value 

(Single 

Channel) 

Max 

Value 

(Single 

Channel) 

Access 

Remarks 

24 

Square root for Channel  2 

– 1            

(MSB- LSB) 

42002 - H 

Bit 

R/W 

Refe

Table 45

 

25 

CJC 

42004 - H 

Integer 

R/W 

Refe

Table 48

 

26 

ACJC 

42004 - L 

Integer 

R/W 

27 

FCJC 

42005 

Integer 

-10.0 

65.0 

R/W 

28 

CCJC 

42006 - H 

Integer 

24 

R/W 

29 

Open Sensor for Channel  2 - 1         

(MSB- LSB) 

42007 - H 

Bit 

R/W 

Refe

Table 46

 

Table 49

 

30 

Percentage Range for Over Reading 

Display 

42180 

Integer 

10 

R/W 

31 

Percentage Range for Under Reading 

Display 

42181 

Integer 

10 

R/W 

32 

Number Of Channel 

42009 

Integer 

24 

R/W 

33 

Alarm Acknowledge 

42011 

Integer 

34 

DO Logic for Relay 5 

– 1                     

(MSB- LSB) 

42012 - H 

Bit 

R/W 

Refe

Table 51

 

35 

Relay 1 - 2 - DO Delay 

42014 

Integer 

99 

R/W 

36 

ASP1 Function for Channel 2 - 1     

(MSB- LSB) 

42030 - H 

Bit 

R/W 

Refe

Table 46

 

&

Table 52

 

37 

ASP1 type for Channel 2 - 1           

(MSB- LSB) 

42032 - H 

Bit 

R/W 

Refe

Table 46

 

&

Table 53

 

38 

ASP2 Function for Channel 2 - 1     

(MSB- LSB) 

42034 - H 

Bit 

R/W 

Refe

Table 46

 

&

Table 52

 

Содержание 40005E

Страница 1: ...INDICATOR Single Channel Dual Channel Masibus Automation Instrumentation Pvt Ltd B 30 GIDC Electronics Estate Sector 25 Gandhinagar 382044 Gujarat India 91 79 23287275 79 91 79 23287281 82 Email supp...

Страница 2: ...ecification 14 3 5 Network Connectivity Optional 14 3 6 Data logging Optional 14 3 7 Display Specification 15 3 8 Power Supply Specification 15 3 10 Signal Isolations And Insulation Specification 15 3...

Страница 3: ...42 Table 19 Sub Parameters of Communication Configuration Mode 42 Table 20 Sub Parameters of Display Configuration Mode 43 Table 21Sub Parameters of Logging Configuration Mode 43 Table 22 Calibration...

Страница 4: ...actory Reset Parameters 64 Table 62 WatchDog Output 64 Table 63 PV Status during Burn Out Condition 66 Table 64 Retransmission Output during Open Over Under Condition 67 LIST OF FIGURES Figure 1 Panel...

Страница 5: ...sible harm to users and damage to the property and to use the product safely Understand the following description signs and symbols read the text and observe Descriptions DESCRIPTION OF SIGNS This ind...

Страница 6: ...to change without notice as a result of continual improvements to the instrument s performance and functions Every effort has been made to ensure accuracy in the preparation of this manual Should any...

Страница 7: ...NG CODE Model No of I p channel Input Type Ch 1 Display Ch 2 Display Aux Power Supply Mounting Communication Retransmi ssion o p Relay o p PV Bar PV Bar 40005E X X X X X X XX XX XX X X S Single 1 E R...

Страница 8: ...R Red G Green Power Supply U1 85 265 VAC U2 18 36 VDC Mounting Method P0 Panel Mounting W1 Wall Mounting Communication 1X 1 RS485 Serial Port 2X 2 RS485 Serial Port 1E 1 RS485 Serial Port and 1 Ether...

Страница 9: ...current carrying wire in the product s mains supply 3 Resistance between the protective conductor terminal and earth must be no greater than 0 1 ohms Use thicker gauge wire if the resistance is too hi...

Страница 10: ...10 40005E Enhanced Bargraph Indicator REF NO m42B om 101 Issue No 02 2 2 Mounting of 40005E Mounting Method Panel Mounting External Dimensions and Panel Cut Out Dimensions Unit mm Figure 1 Panel Cutou...

Страница 11: ...anel LEDs Check the Relay Output DO RL The corresponding LED illuminates when input signal Analog Input is above or below Alarm Set point Check if the relay voltage and current is within the specified...

Страница 12: ...F NO m42B om 101 Issue No 02 through compliance tests Installation state Check for looseness of the cable connector and damage of the cable No looseness or damage Ambient environment Check if the temp...

Страница 13: ...een three wires should be equal LINEARIZATION FOR LINEAR INPUT 10 POINT Linearization applicable only for linear input Input Range for RTD PT 100 CU53 and NI120 TC and Linear input Table 4 Input types...

Страница 14: ...ided for modification of all control and functional parameters CONFIGURATION SOFTWARE All Configurable parameters can be set through PC Based software MEMORY Non volatile restored after power loss 3 4...

Страница 15: ...20 Mohms at 500V DC 1 Primary terminals indicate power terminals and relay output terminals 2 Secondary terminals indicate analog input signals Digital Contact output terminals communication terminal...

Страница 16: ...ue No 02 3 12 Environmental Specification AMBIENT TEMPERATURE 10 to 55 C HUMIDITY 30 to 95 RH Non Condensing TEMPERATURE COEFFICIENT For All Analog input circuits 100ppm WEIGHT 1 25 KG INSTRUMENT WARM...

Страница 17: ...Panel Diagram Figure 3 Front Panel Diagram for Dual Channel Name of Part Indication on Front Panel Channel Data Display Channel Data Bar Display 101 LED segment for indicating Channel Data in percent...

Страница 18: ...ode Sub mode It is also used to escape from edit mode without saving the respective parameter INCREMENT KEY It is used to increment value in run mode and in other modes DECREMENT KEY It is used for de...

Страница 19: ...12 Pre Feb Cable RS 485 Communication 4 Wire Size 26 16AWG Screw Size M2 0 Steel Ni Plated Ethernet Communication 1 RJ 45 Connector Figure 6 Connection Terminals 1 Do not touch the terminals of unit w...

Страница 20: ...use a line filter on the secondary side Do not place the primary and secondary power cables close to each other 3 High voltage transients may occur when switching inductive loads such as some contact...

Страница 21: ...User Manual 21 40005E Enhanced Bargraph Indicator REF NO m42B om 101 Issue No 02 5 2 Cable Details 1 Digital Output Relay Cable Details Figure 7 Relay Cable Connection...

Страница 22: ...e DIGITAL OUTPUT RELAY DO RL Sr No Connector Pin No Pin Detail Ferrule No 1 2 COMMON 1 2 2 3 NO 1 3 3 4 NC 1 4 4 5 COMMON 2 5 5 6 NO 2 6 6 7 NC 2 7 7 8 COMMON 3 8 8 9 NO 3 9 9 10 NC 3 10 10 11 COMMON...

Страница 23: ...User Manual 23 40005E Enhanced Bargraph Indicator REF NO m42B om 101 Issue No 02 2 Analog Input Output Cable Details Figure 8 Analog Input Output Cable Connection...

Страница 24: ...101 Issue No 02 Table 7 Pin Details of Analog Input Output Cable ANALOG INPUT OUTPUT AI 1 Sr No Connector Pin No Pin Detail Ferrule No 1 1 AO 1 1 2 2 AO 1 2 3 3 AO 2 3 4 4 AO 2 4 5 5 TPS 1 5 6 6 1A 6...

Страница 25: ...User Manual 25 40005E Enhanced Bargraph Indicator REF NO m42B om 101 Issue No 02 3 Back Plate Wiring Details Figure 9 Back Plate Wiring Details...

Страница 26: ...ferences This is the preferred option in areas where there is a potential for high electrical noise or if cabling lacks the cleanliness of conduit or wire trays Good installation practice for RS485 sy...

Страница 27: ...Input cable and Communication cable to 40005E as shown in figure 7 figure 8 figure 9 respectively 2 Connect Power supply Either 85 265 VAC or 18 36 VDC as per the ordering code to 40005E The power on...

Страница 28: ...User Manual 28 40005E Enhanced Bargraph Indicator REF NO m42B om 101 Issue No 02 7 MENU LAYOUT 7 1 Parameter Flow Diagram Figure 12 Program Mode Flow diagram...

Страница 29: ...User Manual 29 40005E Enhanced Bargraph Indicator REF NO m42B om 101 Issue No 02 Figure 13 Configuration Mode Flow diagram...

Страница 30: ...User Manual 30 40005E Enhanced Bargraph Indicator REF NO m42B om 101 Issue No 02 Figure 14 Configuration Mode Flow diagram...

Страница 31: ...REF NO m42B om 101 Issue No 02 Figure 15 Calibration Mode and Security Mode Flow diagram Note For All modes Ch 2 and Ch All will be displayed for Dual Channel Bargraph only For Single Channel Bargraph...

Страница 32: ...enu press Menu Enter key To change the submenu parameter press increment or decrement key To come out from any menu press Esc key 1 Program Mode Pressing MENU key CHANNEL window shows Prog Prog messag...

Страница 33: ...put is required CSP C S Hy Hystresis Control Set point Hysteresis for Channel 1 and 2 Where total Number of Channels 2 0001 for all 2 channels Control Output is required C S Hy 2 Configuration Mode By...

Страница 34: ...ion Configuration Different parameters for Communication Configuration Comm Disp Display Configuration Different parameters for Display Configuration diSP Log Logging Configuration Different parameter...

Страница 35: ...C 000 0 CJC op se Open Sensor UP Down UP Down 0 Down 1 UP 0 for all 2 channels oP SE Ov pe Percentage Range for Over Value Display 0 to 10 0 Ov PE un pe Percentage Range for Under Value Display 0 to 1...

Страница 36: ...Point Linearization This instrument has feature of Ten point linearization for each channel User can define up to ten different segments of the full scale input in percentage with Full scale engineeri...

Страница 37: ...play varies between 1 0 2000 for input of 0 10 2 2000 4000 for input of 10 50 3 4000 4500 for input of 50 60 4 4500 7500 for input of 60 80 5 7500 9999 for input of 80 100 Similarly the configuration...

Страница 38: ...all 2 channels A1 Fn a1 mp Alarm 1 Mapping See DO Digital Output Configuration Refer Table 15and note 1 Digital Output is required A1 mP a1 tp Alarm 1 Type lo Hi 0 LOW 1 HIGH 0 for all 2 channels A1 t...

Страница 39: ...CSP Watch dog Mappin g DO number for W D MP 1 Relay 3 3 Relay 3 3 None 0 Relay 2 2 Relay 1 1 Relay 5 5 2 None 0 Relay 4 4 Relay 4 4 Relay 4 4 None 0 Note 1 1 None means no DO Digital Output is assigne...

Страница 40: ...6 Based On Watchdog Output Disable Enable selection Error messages will be shown i e if Watchdog Output is disabled then no error message will be shown and other operation goes smoothly When Watchdog...

Страница 41: ...tputs Chapter 8 Table 17 Sub Parameters of AO Analog Output Configuration Mode Sub parameters of AO Analog output Configuration Mode Optional Channel Window Name Setting name and description Default v...

Страница 42: ...de Channel Window Name Setting name and description Default value Shows only if Symbol sr no Unit ID 1 to 247 1 Sr no C1 B r Com Port 1 Baud rate 9600 19 2k 57 6k 9600 19 2k 0 9600 1 19 2 K 2 57 6 K 1...

Страница 43: ...parameters of Display Configuration Mode Channel Window Name Setting name and description Default value Shows only if Symbol t out Timeout for display back to Run Mode 10 to 100 Seconds 60 t oUt Fltr...

Страница 44: ...our RTC Hour 0 to 23 0 Hour Min RTC Minute 0 to 59 0 Min Sec RTC Second 0 to 59 0 Sec Date RTC Date 1 to 31 0 date Mon RTC Month 1 to 12 0 Mon year RTC Year 1 to 255 0 yEAr Note 5 1 Periodic and Event...

Страница 45: ...Name Setting name and description Default value Shows only if Symbol PWD Password Asking for Password 0 Password is other than zero PWD 0 Calibration Mode Password 0 to 9999 0 Password is other than...

Страница 46: ...menu F rSt Factory Reset Parameter By Pressing MENU key CHANNEL window shows pwd Pwd message Press MENU Key to enter into edit mode of password and then press INCREMENT key or DECREMENT key to modify...

Страница 47: ...y default value Calibration and parameters will set to default value Refer Note 6 F RST 2 ALL Note 6 1 After resetting Calibration Parameters All values to factory reset values the CHANNEL window will...

Страница 48: ...CK 3 HIGH ALARM RELAY OFF ON ON OFF OFF OFF ASP1 TRIP RELAY OFF ON OFF OFF ON OFF LOW ALARM RELAY OFF ON OFF ON OFF OFF ASP1 TRIP RELAY OFF ON OFF OFF ON OFF ALARM AL2 MOMEMTARY ALARM when in abnormal...

Страница 49: ...LARM RELAY OFF ON OFF ON OFF OFF OFF ASP2 TRIP RELAY OFF ON OFF OFF ON ON OFF ALARM AL3 MAINTAINED ALARM when in abnormal condition ack not pressed CONDITION NORMAL ABNORMAL UP O S DOWN O S ACK 2 NORM...

Страница 50: ...e goes below set point by a certain amount before then only output will turn off or on again On Off Hysteresis prevents the output from chattering or making fast continual switches if the cycling abov...

Страница 51: ...ble 28 Device Fault can be monitored by a failsafe relay output which is mapped for watchdog output When WDT is disable device will continue to work with fault The Fault LED will be on in this conditi...

Страница 52: ...ape key to come out to normal operation PV input sensor calibration Note that all signal cards and all input type group must be calibrated before proceeding further There is no need to calibrate all i...

Страница 53: ...By pressing MENU key CHANNEL windows shows rtr 1 rtr 1 It indicates which Retransmission output needs to be calibrated By pressing MENU key CHANNEL window shows the value being outputted on Retransmi...

Страница 54: ...n RTU mode is Coding System 8 bit binary hexadecimal 0 9 A F Two hexadecimal characters contained in each 8 bit field of the message Bits per Byte 1 start bit 8 data bits least significant bit sent fi...

Страница 55: ...ges when a master device sends a query to a slave Device it expects a normal response One of four possible events can occur from The master s query If the slave device receives the query without a com...

Страница 56: ...ed in the query data field is not an allowable value for the Slave 10 3 Modbus Addresses Table 35 Process Value and Status Read Parameters Sr No Parameter Description Modbus Address Parameter Type Acc...

Страница 57: ...tatus Register and its bits arrangement Parameter Alarm 1 Status Alarm 1 Status Modbus Address 30028 30029 Bit No 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 Descriptio...

Страница 58: ...40181 Integer 0 5 R W 21 Ch 1 2 ASP3 MAP 42549 Integer 0 5 R W 22 Ch 1 2 ASP4 MAP 42561 Integer 0 5 R W 23 Ch 1 2 CSP MAP 40193 Integer 0 5 R W Table 42 Holding Registers Parameters Part 2 Sr No Para...

Страница 59: ...M 2 42062 Integer 0 1 R W Refer Table 57 54 Parity COM2 42063 Integer 0 2 R W Refer Table 58 55 Ethernet IP address 1 2 42064 Integer 0 255 R W 56 Ethernet IP address 3 4 42065 Integer 0 255 R W 57 Su...

Страница 60: ...l 1 42581 Integer 0 100 R W 97 Linearization Percent Step 2 LP02 Channel 1 42582 Integer 0 100 R W 98 Linearization Percent Step 3 LP03 Channel 1 42583 Integer 0 100 R W 99 Linearization Percent Step...

Страница 61: ...5 Linearization Percent Step 10 LP10 Channel 2 42610 Integer 0 100 R W 126 Linearization Range Step 1 LR01 Channel 2 42611 Integer 1999 9999 R W 127 Linearization Range Step 2 LR02 Channel 2 42612 Int...

Страница 62: ...1 to 5 V 17 0 x 11 1 5V 1999 9999 0 to 10 V 18 0 x 12 0 10V 1999 9999 Table 45 Bit accessible Holding Register Parameter Square Root Parameter Square root Square root Modbus Address 42002 42003 Bit No...

Страница 63: ...e Modbus Index Parameter value 0 NO 0 ACJC 0 DOWN 0 OFF 1 YES 1 FCJC 1 UP 1 ON 2 CCJC Table 51 DO Digital Output Logic Selection Table 52 Set Point Function Selection Table 53 Set Point Type Selection...

Страница 64: ...le 62 Logging Enable Selection Log Mode Selection Factory Reset Parameters WatchDog Output Modbus Index Parameter value Modbus Index Parameter value Modbus Index Parameter value Modbus Index Parameter...

Страница 65: ...NO m42B om 101 Issue No 02 11 TROUBLESHOOTING If the operating display does not appear after turning on the ENHANCED BARGRAPH INDICATOR s power follow the measures in the procedure below If a problem...

Страница 66: ...100 LEDs will blink 101 st LED will be OFF 0 to 10V PV LOW No LED will blink OR ON if PV value is greater or equal to 1 of total range Note If set PV_low PV_high for input type is less then maximum v...

Страница 67: ...ds the last value of retransmisison o p PV DOWN REV 20 8 Holds the last value of retransmisison o p Holds the last value of retransmisison o p PV UP REV 3 2 Holds the last value of retransmisison o p...

Страница 68: ...L DATA HOW TO FETCH HISTORICAL DATA 1 For Periodic Logging Periodic Data logging Record Frame Detail Parameter Detail Bytes Log Frame detection 101 102 2 Record Number 4 Time Stamp 6 No of Bytes for P...

Страница 69: ...hannels for periodic data logging so that Total Bytes in 1 record 12 2 8 28 Record Per page 256 28 9 Odd Number So in this case for fetching data follow below procedure 1 Write Hold Data 1 2 Write Fet...

Страница 70: ...Log 42401 42356 Integer R 2 Hold Event Data 42460 Integer R W 3 Fetch Event Data 42459 Integer R W 4 Total Periodic Records 42457 42458 Integer R 42457 Higher byte 42458 Lower Byte 5 Log Roll Over cou...

Страница 71: ...ers o Linearization note added in calibration procedure o Modbus parameters for Linearization added 2 Version No 02 o Back Plate Wiring Details are added Masibus Automation Instrumentation Pvt Ltd Cus...

Отзывы: