4
• Output cables having a modular plug ( similar to
a telephone connector) must never be connec-
ted to a telephone outlet.
• Products with a welded plastic housing are
not repairable. Please contact your supplier for
replacement part.
• This product contains hazardous voltages and
there are no user replaceable parts inside the
product. Never attempt to remove the cover.
• WARNING: No modification of this equipment is
allowed. Any repair/service should be carried out
by qualified personnel who may get assistance
by contacting the manufacturer or the manu-
facturer’s agent.
• Products specified to have automatic polarity
protection must be switched off if a battery is
connected with reverse polarity. The protection
will be automatically reset when the polarity has
been corrected.
• In chargers specified to have a replaceable fuse
as polarity protection the fuse must be replaced
if the battery has been connected with reverse
polarity. When replacing the fuse; a fuse of the
same type and rating must be used.
• If the product is specified to comply with the
standard for Medical Electrical Equipment (stan-
dards based on IEC60601-1) it complies with
some of the requirements for medical electrical
equipment and may be used in medical applica-
tions and hospital environments.
• The product must not be used in the vicinity of
flammable anesthetic gases or in other environ-
ments with flammable or explosive atmosphere.
• If the product is specified to comply with the
standard for Medical Electrical Equipment
for Home Healthcare Environment (standard
IEC60601-1-11) it may be used in medical
applications used in a home healthcare environ-
ment.
NOTE: Products relying on safety earth for
protection (Class I) may not be used in home
healthcare environment unless they are perma-
nently wired to the building installation:
Installation must only be carried out by qualified
service personnel, following the below
instructions:
- The protective earth conductor must be min.
0.75 mm
2
.
- Connect the protective earth conductor to the
external protective earthing system.
- Verify that the protective earth terminal used is
connected to the external protective earthing
system.
- Verify the integrity of the external protective
earthing system.
• This product converts the mains voltage to
a safety extra low voltage.
Some products may be treated as Applied
Part (Type BF) according to standard EN/IEC
60601-1 and may come in physical contact with
a patient.
• This product must be operated in an environ-
ment within temperature range +5 to +40°C,
humidity 15 - 93 % RH and atmospheric pres-
sure 70 - 106 kPa (700 - 1060 hPa).
• Expected service life of this product and acces-
sories delivered with this product is three (3)
years, if operated as indicated above. However;
the guarantee times indicated in document
”TERMS OF SALES AND DELIVERY FOR MAS-
COT AS” apply (available at www.mascot.com).
• Environmental parameters during transport and
storage between uses: temperature range -25 to
+85 °C, humidity 15 - 93% RH NC and atmosp-
heric pressure 70 - 106 kPa (700 - 1060 hPa).
• If stored for longer periods of time the
environmental parameters should be within
the temperature range +5 to +35°C, humidity
range 10 - 75% RH NC and atmospheric pres-
sure 70 - 106 kPa (700 - 1060 hPa) to maintain
the products expected service life.
• Expected shelf life of this product is one (1) year,
if stored as indicated above.
• This product complies with the requirements to
electromagnetic compatibility for medical ele-