Instruction Manual Mi 306 EC/TDS Meter
8
8
8
8
8
Item
Item
Item
Item
Item
Description
Description
Description
Description
Description
Default
Default
Default
Default
Default
Valid value
Valid value
Valid value
Valid value
Valid value
tc
Temp. compensation coeff.
0.00 to 6.00 %/°C
1.90
tcE
Temp. compensation mode
Atc, Mtc, notc
Atc
rEF
Reference Temperature
20 or 25°C
25°C
tdS
TDS factor
0.40 to 0.80
0.50
CEL
Cell constant (K)
0.500 to 1.700
1.000
Aof
Auto-Off enable
On, OFF
On
YEA
Year
1999 to 2098
1999
dAT
Date
dd:mm
01:01
hou
Hour
hh:mm
00:00
uEr
Firmware release
Chr
Battery level test
To change the value use the
▲
or
▼
key.
Press SHIFT+CFM to confirm.
Note:
Note:
Note:
Note:
Note: Press SHIFT+SETUP before confirmation to escape without changing the previ-
ously set value.
The following table lists the setup items, their valid range of values and the factory
settings (default):
Notes:
Notes:
Notes:
Notes:
Notes:
1
Once enabled, the Auto Off time is fixed at 5 minutes.
2
When the battery level test is selected (Chr), LCD will display the remaining percent-
age of battery charge. 100% means fully charged battery and 0% corresponds to the
minimum battery level that allows the meter to operate.