Instruction Manual Mi 306 EC/TDS Meter

10

10

10

10

10

3

Make sure the meter is calibrated before taking measurements.

4

If measurements are taken successively in different samples, to have accurate readings it is
recommended to rinse the probe thoroughly with deionized water before immersion in the
samples.

5

To maximize battery life, the meter is automatically switched off after 5 minutes of non-use. To
reactivate the instrument press the ON/OFF key. This feature can be disabled by entering the
setup mode and selecting the "AoF" item (see SETUP section for details).

6

TDS reading is obtained multiplying the EC reading by the TDS factor, which has a default
value of 0.50. It is possible to change the TDS factor in the 0.40 to 0.80 range by entering
the setup mode and selecting the "tdS" item (see SETUP section). Always set the reference
temperature to 25°C when measuring TDS.

7

When the use of an alternate function (MR, SETUP, CFM, FIXED, TC and HOLD) is
requested, press and hold the SHIFT key first and then the second key.

AUTORANGING

The EC and TDS scales are autoranging. The meter automatically sets the scale with
the highest possible resolution.

By pressing SHIFT+FIXED, the autoranging feature is disabled and the
current range is frozen on the LCD. "FIXED" symbol blinks on the LCD.
To restore the autoranging option press SHIFT+FIXED again.

Note: Autoranging is automatically restored if the RANGE key is pressed, if the setup or

calibration modes are entered and if the meter is turned off and back on again.

AUTO ENDPOINT MODE

The Auto Endpoint feature allows the user to freeze the display, when a stable reading is
reached (EC/TDS/NaCl and temperature range).

To enter the Auto Endpoint mode, press SHIFT+HOLD keys.

Содержание MI 306

Страница 1: ...C TDS Meter EC TDS Meter EC TDS Meter EC TDS Meter EC TDS Meter Mi 306 A A A A Automatic L utomatic L utomatic L utomatic L utomatic Logging EC TDS Meter ogging EC TDS Meter ogging EC TDS Meter ogging...

Страница 2: ...LOGGING FUNCTION 14 DATA TRANSFER TO PC 17 BATTERY REPLACEMENT 18 CONDUCTIVITY VS TEMPERATURE CHART 18 ACCESSORIES 19 DISPLAY DISPLAY DISPLAY DISPLAY DISPLAY A PRIMARY DISPLAY B AUTO ENDPOINT MODE IND...

Страница 3: ...or with SHIFT to enter setup mode 5 ATC TC key to select temperature compensation mode or with SHIFT to view the temperature coefficient value 6 MEM MR key to store or with SHIFT recall measurements 7...

Страница 4: ...ly ATC or manually MTC compensated for temperature The temperature coefficient value is user selectable It is possible to disable the temperature compensation and measure the actual conductivity NOTC...

Страница 5: ...0 1 S cm from 30 0 to 299 9 S cm 1 S cm from 300 to 2999 S cm 0 01 mS cm from 3 00 to 29 99 mS cm 0 1 mS cm over 30 0 mS cm TDS 0 01 mg L from 0 00 to 14 99 mg L 0 1 mg L from 15 0 to 149 9 mg L 1 mg...

Страница 6: ...aCl Calibration 1 point with HI 7037 buffer optional Temperature 2 points at 0 and 50 C Calibration plus 1 C adjustment Temperature Automatic or Manual from 0 to 60 C Compensation can be disabled to m...

Страница 7: ...upper LCD The lower LCD displays the blinking code of the current setup item Select the desired setup item using the or key Press SHIFT CFM to confirm Note Note Note Note Note If SHIFT SETUP are press...

Страница 8: ...to 2098 1999 dAT Date dd mm 01 01 hou Hour hh mm 00 00 uEr Firmware release Chr Battery level test To change the value use the or key Press SHIFT CFM to confirm Note Note Note Note Note Press SHIFT S...

Страница 9: ...sired range EC TDS NaCl is selected on the LCD Allow for the reading to stabilize The upper LCD displays the measure in the selected range while the temperature is displayed on the lower LCD Notes Not...

Страница 10: ...setup mode and selecting the tdS item see SETUP section Always set the reference temperature to 25 C when measuring TDS 7 When the use of an alternate function MR SETUP CFM FIXED TC and HOLD is reques...

Страница 11: ...the range options TEMPERATURE COMPENSATION Three options of compensating temperature are available Automatic Atc The probe has a built in temperature sensor the value of the tempera ture is used to a...

Страница 12: ...S cm 12 88 mS cm 80 0 mS cm 111 8 mS cm To enter EC calibration select the EC range and press the CAL key Note TDS reading is automatically derived from the EC reading and no specific calibration for...

Страница 13: ...the MA 9066 MA 9066 MA 9066 MA 9066 MA 9066 calibration solution sea water solution as a 100 NaCl standard solution To enter NaCl calibration select the NaCl from RANGE key and press CAL Rinse the pro...

Страница 14: ...and the MEM key is pressed the sample will not be stored and the LCD will display FULL In this case it is necessary to delete some data from memory to proceed To view logged data To view logged data...

Страница 15: ...o skip to another sample using the up and down arrow keys For example if the reading of a sample is displayed pressing the up arrow key will cause the meter to display the reading of the next sample A...

Страница 16: ...Press SHIFT HOLD to escape without data deletion When viewing through the logged data the NULL message will be displayed when selecting a deleted sample To delete all data in memory To delete all dat...

Страница 17: ...cable start the terminal program and set the communication options as follows 8 N 1 no flow control To send a command to the meter the scheme is command CR The computer sends the command expressed as...

Страница 18: ...es with increasing temperature It is affected by the type and number of ions in the solution and by the viscosity of the solution itself Both parameters are temperature dependent The depen dency of co...

Страница 19: ...t defects in materials and manufacturing for a period of 3 years from the date of purchase Electrodes and Probes are warranted for 6 months If during this period the repair or replacement of parts is...

Страница 20: ...ls kik t sor 11 6726 Szeged Hungary Tel 36 62 428 050 Fax 36 62 428 051 e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA Tel 1 252 443 3630 Fa...

Отзывы: