background image

3

2. INTRODUCTION

2.1 Inspection

Examine the shipping carton for any sign of damage. Inspect the 
unit and any accessories for damage. If there is any damage then 
consult your distributor immediately.

2.2 Description

The Martindale SPC70 has the following functions and features:

‹

  Complies with BS EN 60942, Class 2

‹

  94dB and 114dB output at 1kHz

‹

  Fits 1 inch and ½ inch (using adaptor) diameter microphones

‹

  Low battery indication

2.3 Accessories (included)

The SPC70 comes with the following accessories:

‹

  1 x 9V PP3 battery  

‹

  ½ inch microphone adaptor

‹

 Carry case

‹

 Instructions

2.4 Battery Installation

Refer to Section 4.1 (Battery Replacement) for the battery 
installation instructions for the SPC70.

4

3. OPERATION

3.1 1 SPC70 Elements

3.2 Low Battery Indication 

If the 

Power ON/Low Battery

 LED illuminates RED, the battery 

needs replacing as output sound pressure level accuracy can no 
longer be guaranteed (See section 4.1 Battery Replacement).

3.3 Testing/Adjusting a Sound Level Meter

Insert the cavity of the SPC70 over the microphone of the sound 

1

Power and low battery LED

2

Power and output level switch

3

Transducer assembly with 1 inch 
cavity for microphone insertion 

4

½ inch microphone adaptor

5

Battery cover

5

level meter being tested / adjusted. 

If the sound level meter has a ½ inch microphone, the ½ inch 
microphone adaptor should first be inserted into the cavity of the 
SPC70.

Set the SPC70 switch to the 94 dB or 114 dB position as required, 
and set the controls of the sound level meter being checked / 
adjusted in accordance with the manufacturer’s instructions and 
power it on.

If required make any adjustment to the sound level meter in 
accordance with the manufacturer’s instructions.

Switch off the SPC70 after use to conserve the battery.

3.4 Testing Considerations

Make sure the sound level meter microphone is properly seated into 
the SPC70 microphone cavity. If the ½ inch microphone adaptor is 
used make sure it is properly seated.

High ambient sources of noise or vibration may cause incorrect 
readings. In such circumstances it is advisable to make any tests 
with the SPC70 set to the 114 dB output sound pressure level.

Always use the certified values for the SPC70 when testing or 
adjusting a sound level meter.

6

4. MAINTENANCE

4.1 Battery Replacement

The battery compartment is underneath the unit.

Undo the screw securing the battery cover and lift off the battery 
cover.

Remove the old battery and replace with a new 9V battery (IEC 
6LR61, NEDA 1604A), observing correct polarity.

Replace the battery cover and secure with the screw.

4.4 Calibration 

To maintain the integrity of measurements made using your 
instrument, Martindale Electric recommends that it is returned 
at least once a year to an approved Calibration Laboratory for 
recalibration and certification.

Martindale Electric is pleased to offer you this service. Please 
contact our Service Department for details.
Email: [email protected]
Tel: 01923 650660

4.3 Cleaning  

 

If contamination is found, clean with a damp soft cloth and if 
necessary a mild detergent or alcohol. Do not use abrasives, 
abrasive solvents, or detergents which can cause damage to the 
unit. If a mild detergent is used, the unit should subsequently 
be thoroughly cleaned with a water dampened soft cloth. After 
cleaning, dry and allow to remain in a dry environment for 2 hours 
before use.

Отзывы: