background image

18

4.5 Cleaning 

  To reduce the risk of surface leakage, this instrument must 

be kept in a clean condition.

 

 Prior to cleaning, ensure that the instrument is disconnected 
from any voltage source.

If contamination is found, clean with a damp soft cloth and if 
necessary a mild detergent or alcohol. Do not use abrasives, 
abrasive solvents, or detergents which can cause damage to the 
unit. If a mild detergent is used, the unit should subsequently 
be thoroughly cleaned with a water dampened soft cloth. After 
cleaning, dry and allow to remain in a dry environment for 2 hours 
before use.

4.6 Repair & Service

There are no user serviceable parts in this unit other than those 
that may be described in section 4. Return to Martindale Electric 
if faulty. Our service department will quote promptly to repair any 
fault that occurs outside the guarantee period. 

Before the unit is returned, please ensure that you have checked 
the: unit, battery, leads and poor connections.

4.7 Storage Conditions

The instrument should be kept in warm dry conditions away from 
direct sources of heat or sunlight, and in such a manner as to 
preserve the working life of the unit. It is strongly advised that the 
unit is not kept in a tool box where other tools may damage it.

19

5. WARRANTY AND LIMITATION OF LIABILITY

This Martindale product is warranted to be free from defects in material 

and workmanship under normal use and service. The warranty period 

is 2 years and begins on the date of receipt by the end user. This 

warranty extends only to the original buyer or end-user customer, 

and does not apply to fuses, disposable batteries, test leads or to any 

product which, in Martindale’s reasonable opinion, has been misused, 

altered, neglected, contaminated, or damaged by accident or abnormal 

conditions of operation, handling or storage. 

Martindale authorised resellers shall extend this warranty on new and 

unused products to end-user customers only but have no authority to 

extend a greater or different warranty on behalf of Martindale.

Martindale’s warranty obligation is limited, at Martindale’s option, to 

refund of the purchase price, free of charge repair, or replacement of 

a defective product which is returned to Martindale within the warranty 

period.

This warranty is the buyer’s sole and exclusive remedy and is in lieu 

of all other warranties, expressed or implied, including but not limited 

to any implied warranty of merchantability or fitness for a particular 

purpose. Martindale shall not be liable for any special, indirect, 

incidental or consequential damages or losses, including loss of data, 

arising from any cause or theory.

Since some jurisdictions do not allow limitation of the term of an implied 

warranty, or exclusion or limitation of incidental or consequential 

damages, the limitations and exclusions of this warranty may not 

apply to every buyer. If any part of any provision of this warranty 

is held invalid or unenforceable by a court or other decision-maker 

of competent jurisdiction, such holding will not affect the validity or 

enforceability of any other provision or other part of that provision.

Nothing in this statement reduces your statutory rights.

20

ELECTRICAL

All specified accuracies are at 23°C ± 5°C, <70% R.H. for 1 year.

Temperature coefficient: 

0.1 x (specified accuracy) per °C. (0°C to 18°C, 28°C to 50°C).

All accuracies below are expressed as ± (percentage of r digits)

DC Voltage

Range

Resolution

Input 

impedance

Accuracy

600mV

0.01mV

10M

:

0.08% + 5

6V

0.0001V

11M

:

60V

0.001V

10M

:

600V

0.01V

1000V

0.1V

Overload protection: 1000V DC or 750V AC rms

Specification
MM94 TRMS Digital Multimeter

Содержание MM94

Страница 1: ...Measurements 15 3 24 Continuity Testing 15 3 25 Diode Testing 16 4 Maintenance 16 4 1 Battery Replacement 16 4 2 Fuse Replacement 17 4 3 Test Lead Replacement 17 4 4 Calibration 17 4 5 Cleaning 18 4 6 Repair Service 18 4 7 Storage Conditions 18 5 Warranty 19 Measurement Categories 20 Specifications 1 1 SAFETY INFORMATION Always read before proceeding REMEMBER SAFETY IS NO ACCIDENT These instructio...

Страница 2: ...th CAT III and CAT II installations up to 1000V to earth and within the operating temperature and humidity range specified The CAT III 1000V marking does not mean the clamp meter can be used to 1000V AC rms but that it will be safe to the user if inadvertently connected across a voltage up to 1000V AC rms to earth If the removable probe tip caps are not fitted to the probes of the test leads their...

Страница 3: ... of LCD Symbols 7 HFR1 High frequency reject 1 kHz selected HFR 2 High frequency reject 100 kHz selected RANGE Manual ranging is selected MAX Maximum indication is displayed MIN Minimum indication is displayed HOLD Display hold is activated P Indicates peak measurement P Indicates peak measurement REL Relative function selected Indicates 30V at input terminals AC Indicates AC measurement DC Indica...

Страница 4: ...ions connect the test leads to the circuit being measured and read the measured ac voltage from the display 11 3 15 DC Voltage Measurements Connect the black test lead to the COM terminal and the red test lead to the terminal Set the rotary switch to the or position as required Taking all necessary safety precautions connect the test leads to the circuit being measured and read the measured dc vol...

Страница 5: ... if d600 15 3 25 Diode Testing If the diode to be tested is in circuit be sure the circuit power is switched off Connect the black test lead to the COM terminal and the red test lead to the terminal Set the rotary switch to the position Press the SHIFT button twice The LCD will display the symbol Taking all necessary safety precautions connect the test leads to the diode being tested If the diode ...

Страница 6: ...ered neglected contaminated or damaged by accident or abnormal conditions of operation handling or storage Martindale authorised resellers shall extend this warranty on new and unused products to end user customers only but have no authority to extend a greater or different warranty on behalf of Martindale Martindale s warranty obligation is limited at Martindale s option to refund of the purchase...

Страница 7: ...range Crest factor 3 at full scale 6 at half scale Peak hold accuracy on AC current ranges 45Hz to 500 Hz 3 5 rdg 500 dts Peak hold response time 1ms Input protection 0 5 A 1000 V fast blow ceramic fuse on 600μA to 400mA ranges 20A 600V fast blow ceramic fuse on 20 A range Note 1 Apply currents 10 A for 30 seconds maximum then allow a 10 minute cooling period Specification MM94 TRMS Digital Multim...

Страница 8: ...0 R H Altitude up to 2000m Pollution degree 2 indoor use SAFETY Conforms to BS EN 61010 1 BS EN 61010 2 033 CAT IV 600V CAT III 1000V Class II double insulation EMC Conforms to BS EN 61326 1 Specification MM94 TRMS Digital Multimeter SPECIFICATION FOR TL16 TEST LEADS Maximum voltage 1000V AC DC Maximum current 10A continuous Connector 4mm banana plug with fixed shroud Environmental Temperature Ope...

Отзывы: