2. TECHNICAL SPECIFICATION

Display:  3.5 digit liquid crystal (LCD), maximum reading 3200 with floating
decimal point. Updates at 2 times per second. A responsive, 34 segment
analogue bargraph is shown below the digital read out.
Polarity:  Automatic, implied for positive values, "-" to show negative
readings.
Overrange:  O.L is displayed.
Zero:  Automatic.
Low battery indication:  A "      " symbol appears when the voltage drops
below that required for correct operation.
Auto Power Off:  After approx. 10 minutes.
Operating:  0 to 50°C @ < 70% relative humidity (RH).
Storage:  -20 to 60°C, 0 to 80% RH (battery must be removed).
Temperature coefficient:  Add (0.1 x % accuracy) per °C from 0 to 18°C,
and from 28 to 50°C.
Maximum altitude:  2,000m.
Battery:  1 x 9 volt, type PP3, MN1604, NEDA1604, JIS006P, IEC 6F22.
Battery life:  500 hours, typical (Zinc - carbon).
Dimensions:  143 x 68 x 47mm (5.6 x 2.7 x 1.9 inches).
Weight:  206g (7.3oz) with battery fitted.
Included: Set of test leads, 9V battery (fitted) & user manual.

DC VOLTS
Ranges:  
320mV, 3.2V, 32V, 320V, 600V.
Resolution:  100µV
Accuracy: below 320mV, ± 1.2% rdg + 1 digit

3.2V range, ± 0.8% rdg + 1 digit
above 3.2V, ± 1.2% rdg + 1 digit 

Input impedance:  1000M

Ω 

up to 320mV

11M

Ω

on 3.2V range

10M

Ω

on all other ranges

Overload protection:  600V DC or AC rms.

However, electrical noise or intense electromagnetic fields in the vicinity of
the equipment may disturb the measurement circuit. Measurement
instruments will also respond to unwanted signals that may be present.
Users should exercise care and take appropriate precautions to avoid
misleading results in the presence of electromagnetic interference.

3. OPERATION

3.1. Precautions
Warning: Before using the instrument, read the Safety Information
Section, and examine it for damage or contamination (excessive dirt,
grease, etc.). Check the test leads for any evidence of cracked or
frayed insulation. If any abnormal conditions exist, do not use the
meter.

Warning: When using probes, keep your fingers behind the guards.

Warning: Measuring voltages which exceed the limits of the meter
may damage it and expose the operator to a shock hazard. Always
work within the meter voltage limits indicated on it's front panel.

Warning: Never ground yourself when taking electrical
measurements. Do not touch exposed metal pipes, outlets, fixtures
etc. which may be at ground potential. Keep your body insulated by
wearing dry clothing, rubber shoes or stand on a rubber mat, or other
suitable material.

3.2 Range selection
When the function (for example, AC Volts) has been selected and a signal
is present, the MM35 will autorange until the most suitable range has been
selected. The reading will be displayed and the meter will continue to
change range according to variations in the input.

5

8

Содержание MM35

Страница 1: ...erve the working life of the multimeter It is strongly advised that the unit is not kept in a tool box where other tools may damage it 4 7 Warranty Faults in manufacture and materials are fully guaranteed for 2 years from date of invoice and will be rectified by us free of charge provided the unit has not been tampered with and is returned to us with its housing unopened Damage due to dropping abu...

Страница 2: ...t html SYMBOLS Equipment complies with relevant EU Directives AC Alternating Current Ground Direct Current Equipment protected by Double Insulation Class II Caution refer to accompanying documents Caution risk of electric shock 4 MAINTENANCE 4 1 Cleaning Maintenance consists of periodic cleaning and battery replacement The exterior of the instrument can be cleaned with a dry clean cloth to remove ...

Страница 3: ...t and black lead to COM socket 4 Connect the test leads to the measurement points and take a reading 3 9 1 Diode test 1 Select the rotary switch position and toggle the button until the symbol appears 2 Remove power from the circuit under test 3 Connect the red test lead to the VΩ socket and black lead to COM socket 4 Touch the probes across the diode and take a reading reverse the probes and agai...

Страница 4: ...urement 1 Select one of the three current functions using the rotary switch and use button A to choose either DC or AC The appropriate symbol appears at the left hand side of the display 2 When measuring less than 320mA connect the red test lead to the µA mA socket and the black lead to the COM socket 3 For current above 320mA the red lead should be connected to the 10A socket 1 INTRODUCTION 1 1 D...

Страница 5: ... or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement circuit Measurement instruments will also respond to unwanted signals that may be present Users should exercise care and take appropriate precautions to avoid misleading results in the presence of electromagnetic interference 3 OPERATION 3 1 Precautions Warning Before using the instrument read the Safet...

Страница 6: ...SISTANCE Ranges 320Ω 3 2kΩ 32kΩ 320kΩ 3 2MΩ 30MΩ Resolution 100mΩ Accuracy below 320Ω 2 rdg 3 digits 320Ω to 320kΩ 1 5 rdg 3 digits above 320kΩ 2 5 rdg 3 digits above 3 2MΩ 5 rdg 5 digits Test current 320Ω range 700µA 3 2kΩ range 130µA 32kΩ range 13µA 320kΩ range 1 3µA 3 2MΩ range 0 13µA Overvoltage protection 500V DC or AC rms DIODE TEST Test current 500µA at 0 6V forward voltage Resolution 1mV A...

Отзывы: