SAFETY INFORMATION: Always read before proceeding.

WARNING

These instructions contain information and warnings that are necessary for safe
operation and maintenance of the MM35. It is recommended that you read the
instructions carefully and ensure the contents are understood. Failure to
understand the instructions and comply with warnings and instructions herein
can result in serious injury, damage or even death.

In order to avoid the danger of electrical shock, it is important that proper safety
measures are taken when working with voltages exceeding 30V AC rms, 42V AC
peak or 60V DC. Never exceed the maximum allowable input level for each
function and range. Refer to the specifications for maximum inputs. Never
touch exposed wiring, connections or live circuits. 

The MM35 must only be used in conditions and for the purpose which it has
been constructed. Attention should be paid to safety instructions, technical
specifications and use of the MM34 in dry surroundings.

Always inspect your meter, test leads and accessories for any sign of damage
before use. If any abnormal conditions exist (e.g: broken test leads, cracked
case, display not reading, etc.), do not attempt to use it. Do not expose it to
direct sunlight, excessive temperature or moisture.

Keep these instructions for future reference. Updated instructions and product
information are available at: www.martindale-electric.co.uk/instruct.html

SYMBOLS:

Equipment complies with relevant EU Directives

AC (Alternating Current)

Ground

Direct Current

Equipment protected by Double Insulation (Class II)

Caution - refer to accompanying documents

Caution - risk of electric shock

4. MAINTENANCE

4.1 Cleaning
Maintenance consists of periodic cleaning and battery replacement. 

The exterior of the instrument can be cleaned with a dry clean cloth to
remove any oil, grease or grime. Never use liquid solvents or detergents.

Repairs or servicing not covered in this manual should only be performed
by qualified personnel.

4.2 Battery Replacement
WARNING - To avoid electrical shock, disconnect the test leads and
any input signals before replacing the battery. Replace only with the
same type of battery.

This meter is powered by NEDA type 1604 or equivalent 9V battery. When
the meter displays the “     ” symbol, the battery must be replaced to
maintain proper operation. To replace the battery, remove the three screws
from the back of the meter and open the bottom case, remove the battery
from battery compartment and replace it with a new one.

4.3 Fuse Replacement 
If no current measurements are possible, check for a blown overload
protection fuse. There are two fuses: F1 for the µA/mA socket and F2 for
the 10A socket. For access to fuses, remove the three screws from the
back of the meter and open the bottom case. Replace F1 only with the
original type 0.5A/250V, fast acting fuse (part no: FUSE500250). 

Replace F2 only with the original type 10A/600V, fast acting ceramic fuse
(part no: FUSE10600).

4.4 Calibration
The recommended calibration interval is 12 months.

2

11

Содержание MM35

Страница 1: ...erve the working life of the multimeter It is strongly advised that the unit is not kept in a tool box where other tools may damage it 4 7 Warranty Faults in manufacture and materials are fully guaranteed for 2 years from date of invoice and will be rectified by us free of charge provided the unit has not been tampered with and is returned to us with its housing unopened Damage due to dropping abu...

Страница 2: ...t html SYMBOLS Equipment complies with relevant EU Directives AC Alternating Current Ground Direct Current Equipment protected by Double Insulation Class II Caution refer to accompanying documents Caution risk of electric shock 4 MAINTENANCE 4 1 Cleaning Maintenance consists of periodic cleaning and battery replacement The exterior of the instrument can be cleaned with a dry clean cloth to remove ...

Страница 3: ...t and black lead to COM socket 4 Connect the test leads to the measurement points and take a reading 3 9 1 Diode test 1 Select the rotary switch position and toggle the button until the symbol appears 2 Remove power from the circuit under test 3 Connect the red test lead to the VΩ socket and black lead to COM socket 4 Touch the probes across the diode and take a reading reverse the probes and agai...

Страница 4: ...urement 1 Select one of the three current functions using the rotary switch and use button A to choose either DC or AC The appropriate symbol appears at the left hand side of the display 2 When measuring less than 320mA connect the red test lead to the µA mA socket and the black lead to the COM socket 3 For current above 320mA the red lead should be connected to the 10A socket 1 INTRODUCTION 1 1 D...

Страница 5: ... or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement circuit Measurement instruments will also respond to unwanted signals that may be present Users should exercise care and take appropriate precautions to avoid misleading results in the presence of electromagnetic interference 3 OPERATION 3 1 Precautions Warning Before using the instrument read the Safet...

Страница 6: ...SISTANCE Ranges 320Ω 3 2kΩ 32kΩ 320kΩ 3 2MΩ 30MΩ Resolution 100mΩ Accuracy below 320Ω 2 rdg 3 digits 320Ω to 320kΩ 1 5 rdg 3 digits above 320kΩ 2 5 rdg 3 digits above 3 2MΩ 5 rdg 5 digits Test current 320Ω range 700µA 3 2kΩ range 130µA 32kΩ range 13µA 320kΩ range 1 3µA 3 2MΩ range 0 13µA Overvoltage protection 500V DC or AC rms DIODE TEST Test current 500µA at 0 6V forward voltage Resolution 1mV A...

Отзывы: