2. TECHNICAL SPECIFICATION

Display:  

3.5 digit liquid crystal (LCD), maximum reading 1999 with floating

decimal point. Updates at 2.5 times per second.

Polarity:  

Automatic, implied for positive values, "-" to show negative

readings.

Overrange:  

1. or -1. is displayed.

Zero:  

Automatic.

Low battery indication:  

A "      " symbol appears when the voltage drops

below that required for correct operation.

Operating:  

0 to 50°C @ < 70% relative humidity (RH).

Storage:  

-20 to 60°C, 0 to 80% RH (battery must be removed).

Temperature coefficient:  

Add (0.1 x % accuracy) per °C from 0 to 18°C,

and from 28 to 50°C.

Maximum altitude:  

2,000m.

Battery:  

1 x 9 volt, type PP3, MN1604, NEDA1604, JIS006P, IEC 6F22.

Battery life:  

300 hours, typical (Zinc - carbon).

Dimensions:  

143 x 68 x 47mm (5.6 x 2.7 x 1.9 inches).

Weight:  

206g (7.3oz) with battery fitted.

Included: 

Set of test leads, 9V battery (fitted) & user manual.

3. OPERATION

3.1. Precautions
Warning: Before using the instrument, read the Safety Information
Section, and examine it for damage or contamination (excessive dirt,
grease, etc.). Check the test leads for any evidence of cracked or
frayed insulation. If any abnormal conditions exist, do not use the
meter.

Warning: When using probes, keep your fingers behind the guards.

Warning: Measuring voltages which exceed the limits of the meter
may damage it and expose the operator to a shock hazard. Always
work within the meter voltage limits indicated on it's front panel.

Warning: Never ground yourself when taking electrical
measurements. Do not touch exposed metal pipes, outlets, fixtures
etc. which may be at ground potential. Keep your body insulated by
wearing dry clothing, rubber shoes or stand on a rubber mat, or other
suitable material.

3.2 Voltage measurement

1. Connect the red test lead to the "V

Ω

" socket, and the black lead to the

"COM" socket.
2. Select the correct function & range using the rotary switch, ensuring that
AC, or DC, ranges are chosen as appropriate. If the magnitude of the
signal is uncertain, then begin with the highest setting, and reduce it until a
satisfactory indication is obtained.
3. Connect the other end of both test leads to the unit under test.
4. In the DC function a "-" is displayed for negative values, positive
indication is implied.

5

8

Содержание MM34

Страница 1: ...ase rotation G Proving units G Socket Testers G Thermometers Probes G Test Leads G Voltage Indicators and G Specialist Metrohm testers 4 5kV G Specialist Drummond testers Martindale Electric Company Limited Metrohm House Penfold Trading Estate Watford WD24 4YY UK Tel 44 0 1923 441717 Fax 44 0 1923 446900 E mail sales martindale electric co uk Website www martindale electric co uk Martindale Electr...

Страница 2: ...re available at www martindale electric co uk instruct html SYMBOLS Equipment complies with relevant EU Directives AC Alternating Current Ground Direct Current Equipment protected by Double Insulation Class II Caution refer to accompanying documents Caution risk of electric shock 4 4 Calibration The recommended calibration interval is 12 months Martindale Electric will carry out routine calibratio...

Страница 3: ...placement WARNING To avoid electrical shock disconnect the test leads and any input signals before replacing the battery Replace only with the same type of battery This meter is powered by NEDA type 1604 or equivalent 9V battery When the meter displays the symbol the battery must be replaced to maintain proper operation To replace the battery remove the three screws from the back of the meter and ...

Страница 4: ... both directions indicates open circuit a reading lower than 0 5 in either direction indicates a shorted diode 3 6 Battery test 1 Connect the red test lead to the µA mA socket and black lead to COM 2 Select either the 1 5V or the 9V Batt switch position 3 Connect the test probes across the battery a satisfactory reading for a 1 5V battery is 1 25V and for a 9V battery is 8 1V Further information r...

Страница 5: ...irt grease etc Check the test leads for any evidence of cracked or frayed insulation If any abnormal conditions exist do not use the meter Warning When using probes keep your fingers behind the guards Warning Measuring voltages which exceed the limits of the meter may damage it and expose the operator to a shock hazard Always work within the meter voltage limits indicated on it s front panel Warni...

Страница 6: ...ccuracy 3 rdg 1 digit Open circuit voltage 3V DC typical Overload protection 500V DC or AC rms BATTERY TEST Ranges 1 5V 9V Resolution 1mV 10mV Accuracy 3 5 rdg 2 digits Load current 150mA for 1 5V range typical 6mA for 9V range typical NOTE This instrument complies with class ll overvoltage CAT lll 600V of the IEC1010 1 EN61010 1 standard and pollution degree 2 in accordance with IEC 664 indoor us...

Отзывы: