18
Any fault in the appliance should be repaired as
necessary before retesting.
3.7 Earth Continuity/Insulation Failure
Possible Reasons for Failure other than a faulty
appliance or Lead
u
Appliance or lead being tested not correctly
connected to the HPAT500.
u
Test probe is not making good contact with the
appliance, or is damaged.
Rectify the problem and start the test again.
3.8 Earth Continuity displays ‘NOTE’
If result for Earth Continuity is displayed as
‘NOTE x.xx PASS’ the instrument has granted a PASS
for the Appliance under test, where Earth Continuity is
higher than the lower pass level, but within the higher
pass level. The advisory “NOTE.... alerts the end user,
who should decide whether a pass is appropriate.