Martel BetaGauge PI Скачать руководство пользователя страница 19

Error Codes

Error

Description

101

A non-numeric entry was received where it should be a
numeric entry

102

Too many significant digits entered

103

Invalid units or parameter value received

105

Entry is above the upper limit of the allowable range

106

Entry is below the lower limit of the allowable range

108

A required command parameter was missing

109

An invalid pressure unit was received

117

An unknown command was received

120

The serial input buffer overflowed

121

Too many entries in the command line

122

Pressure module not connected

17

Содержание BetaGauge PI

Страница 1: ...taGauge PI PI PRO Reference Class High Accuracy Digital Pressure Test Gauge User Reference Manual This document includes information for both the CSA and CSA ATEX certified versions of the BetaGauge P...

Страница 2: ......

Страница 3: ...martelcalibrators com e mail sales martelcorp com Tel 603 434 1433 800 821 0023 Fax 603 434 1653 Martel Electronics 3 Corporate Park Drive Derry NH 03038 1 2 Standard Equipment Check to see that your...

Страница 4: ...reas An Ex hazardous area as used in this manual refers to an area made hazardous by the potential presence of flammable or explosive vapors These areas are also referred to as hazardous locations see...

Страница 5: ...n objects or water capable of impairing safety Caution To avoid possible damage to calibrator or to equipment under test If the message changes to OL the range limit is exceeded and the pressure sourc...

Страница 6: ...can scroll forward and backwards through the 18 standard engineering units plus a one custom unit scale When the desired unit is displayed press ENTER or CONFIG Pressure will now be displayed in the c...

Страница 7: ...e gauge 8 Function Lock Access to each of the settable parameters above can be turned off once set to prevent unauthorized changes to configuration This is accomplished through a password protected su...

Страница 8: ...iplied by the PSI measured the result will be displayed For example 40 PSI is the equivalent of 1000 lbs of product in a tank You want to display the product weight using a 100 PSI gauge By setting a...

Страница 9: ...lt powered version the terminals for power input are located on the rear of the gauge To apply power simply connect 24 volts to the rear terminal block taking care to observe proper polarity BetaGauge...

Страница 10: ...Input Ranges See page 13 for a table of available ranges in PSI plus equivalent ranges and resolution for all engineering units Accuracy Positive Pressure 0 04 of reading 0 01 of full scale Vacuum 0...

Страница 11: ...II Mechanical Dimensions 4 5 diameter x 2 2 depth x 5 height 11 4cm x 5 6cm x 12 7cm PIR PRO depth is 1 5 or 3 7cm Pressure Connection NPT Male Housing Stainless steel meets NEMA 4 IP65 PIR PRO cast Z...

Страница 12: ...hen measurement deviates by more than 75 of the specified accuracy or 0 04 Test Equipment Verification and calibration of the BetaGauge PIR requires pressure and or vacuum standards able to produce an...

Страница 13: ...ssure Standard to output 437 50 psi 87 5 After the output has stabilized press the ENTER key to continue As the unit takes readings the screen will show When the readings are complete the screen shoul...

Страница 14: ...the unit takes readings the screen will show When the readings are complete the screen should look as shown below Use the Pressure Standard to output 125 00 psi 25 After the output has stabilized pre...

Страница 15: ...Note Only some ranges use vacuum calibration If your gauge is not one than this step will be automatically skipped and calibration will be complete Use the Pressure Standard to output 12 00 psi After...

Страница 16: ...34 1056 5 2113 0 1056 5 2113 0 7043 4 21130 35217 70434 mmH2O 4 C 703 089 281 24 703 09 10546 21093 10546 21093 70309 mmH2O 20 C 704 336 281 73 704 34 10565 21130 10565 21130 70434 mH2O 4 C 0 703089 0...

Страница 17: ...ion level of the Calibrator TARE Tares the offset pressure of the reading on the calibra tor TARE Returns the current tare value PRES_UNIT Returns the pressure unit for the upper display PRES_UNIT Set...

Страница 18: ...uare inch Bar Pressure in bars mBar Pressure in millibars Kg cm2 Pressure in kilograms per centimeter squared InH2O4C Pressure in inches of water at 4 C InH2O20C Pressure in inches of water at 20 C In...

Страница 19: ...value received 105 Entry is above the upper limit of the allowable range 106 Entry is below the lower limit of the allowable range 108 A required command parameter was missing 109 An invalid pressure...

Страница 20: ...horization form can be obtained from our website www martelcorp com under customer service You can also call 1 800 821 0023 to have a form faxed Martel will not be responsible for damage as a result o...

Страница 21: ...19...

Страница 22: ...20...

Страница 23: ...21...

Страница 24: ...P N 0200018 Rev D 7 10 www martelcalibrators com e mail sales martelcorp com Tel 603 434 1433 800 821 0023 Fax 603 434 1653 Martel Electronics 3 Corporate Park Drive Derry NH 03038...