21
ERROR MESSAGES
ERROR
MESSAGE
REASON
ACTION
Low battery:
This warning shows that the
voltage of batteries is too low
to use.
Please
replace
with
new
batteries or plug the AC
adaptor for operation.
Overload:
The total load exceeds the
maximum capacity of the
scale.
Please reduce the loading and
try again.
Counting error (too high):
Indicates that the signal from
the load cells is too high.
This error is normally caused
by a serious fault on the scales
i.e. a faulty load cell or wiring.
Please
contact
the
local
Service Representative.
Counting error (too low):
Indicates that the signal from
the load cells is too low.
This error is normally caused
by a serious fault on the scales
such as a faulty load cell or
wiring. Please contact the
local Service Representative.
Zero count over calibration:
Zero
range
+10%
while
powered on.
Please re-calibrate the
instrument.
*1
Zero count under calibration:
Zero
range
-10%
while
powered on.
Please re-calibrate the
instrument.
*1
EEPROM error:
Indicates that there is a fault
with the scales software.
This error is normally caused
by a serious fault on the scales
such as a faulty load cell or
wiring. Please contact the
local Service Representative.
ADC Error:
Component defect.
Please
contact
the
local
Service Representative
*1
– These errors are usually less serious and can be rectified by re-
calibration; however they can indicate an underlying problem with the
scales. Zero Range is set to 10% of Scale Capacity E.g. Capacity
200.0kg = Zero Range +/-20.0kg