Guidance and manufacturer

’s declaration – electromagnetic immunity. 

 

The M-550 is intended for use in the electromagnetic environment specified below. The customer or user of 
the M-550 should ensure that it is used in such an environment. 

 

Immunity Test 

IEC 60601 test level 

Compliance level 

Electromagnetic environment-
guidance 

 
 
 
 
 
Conducted RF 
IEC61000-4-6 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Radiated RF 
IEC 61000-4-3 

 
 
 
 
 
3 Vrms 
150KHx to 80MHz 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
3 V/m 80MHz to 2,5GHz 

 
 
 
 
 
3 Vrms 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
3 V/m 

Portable and mobile RF 
communications equipment should be 
used no closer to any part of the M-
550 including cables, than the 
recommended separation distance 
calculated from the equation 
applicable to the frequency of the 
transmitter. 
 

Recommended separation 
distance: 

d = 1,2 √

P

 

d = 1,2 √

P

 80MHz to 800MHz 

d = 2,3 √

P

 800MHz to 2,5GHz 

 
Where 

is the maximum output 

power rating of the transmitter in 
watts (w) according to the transmitter 
manufacturer and 

d

 is the 

recommended separation distance in 
metres (m). 
Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey, should 
be less than the compliance level in 
each frequency range. 
 
Interference may occur in the vicinity 
of equipment marked with the 
following symbol: 

 

NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, 
objects and people. 

A) 

  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured 
field strength in the location in which the M-550 is used exceeds the application RF compliance level above, the M-550 should be 
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating the M-550. 

B) 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

 
 
 
 
 
 
 
 
 
 
 
 
 

Page 12 

Содержание M-550BT

Страница 1: ...USER MANUAL M 550BT Please take time to read these instructions before starting to use the scale Version 1 1 01 18...

Страница 2: ......

Страница 3: ...c Symbols 6 Power Supply and Low Battery 7 Operation 7 Setting the Bluetooth Function 8 Error Messages 9 Troubleshooting 10 EMC Guidance and Manufacturer s Declaration 11 Recommended Separation Distan...

Страница 4: ...Product Specification Model M 550BT Accuracy Class Class III Capacity Division 200kg x 200g Column No Units of Measure Kg Function Keys ON OFF ZERO TARE HOLD BMI Stabilization Time 1 2 Seconds Operat...

Страница 5: ...f swallowed promptly seek medical assistance If you have any problems contact Marsden your local dealer your service partner Cleaning We recommend using alcohol based wipes or similar when cleaning th...

Страница 6: ...entification of manufacturer of medical product including address Type B Applied Part Dispose of old appliances separately from your household waste This product must be disposed of at a communal coll...

Страница 7: ...1 Press this key to switch on the scale The display will show 0 0kg 2 Pressing the ON OFF key on the opposite side of the display will reverse the scale s display therefore the display can be viewed...

Страница 8: ...or three seconds The scale will enter BMI mode 2 The display will show the last inputted height 3 Press the HOLD BMI button to decrease the height and ZERO TARE to increase the height 4 When the corre...

Страница 9: ...the scale and try again If the scale continues to show the error message it indicates a fault with the electronics or wiring High Low Zero Count 1 The scale is above its zero range Please remove any w...

Страница 10: ...h the following section carefully Self checking tips Some functional defects can be identified and maintained by users as listed below 1 Power Failure Check if the mains power adaptor has been correct...

Страница 11: ...r 6kV contact 8kV air Floors should be wood cement or ceramic tile If floors are covered with synthetic material the relative humidity should be at least 30 Electrical fast transient burst IEC 61000 4...

Страница 12: ...as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol N...

Страница 13: ...aration distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 P 80MHz to 800MHz d 1 2 P 800MHz to 2 5GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 10...

Страница 14: ...Panel and LCD Display LCD Display 1 Stable Symbol 2 Minor weight value 3 Zero Manufacturer s Declaration of Conformity Page 14...

Страница 15: ...dengroup co uk www marsden weighing co uk Manufacturing and Distribution Unit 7 Centurion Business Park Coggin Mill Way Rotherham S60 1FB Head Office Unit 1 Genesis Business Park Sheffield Road Rother...

Отзывы: