10

9

Memory Input

Your Automatic Digital Blood Pressure Monitor has a 

memory  feature.   When  a  valid  measurement  is  made,  

the  reading  data  (systolic,  diastolic  and  pulse  rate)  will 

be  stored  in  the  memory  automatically  as  soon  as  the 

unit  is  turned  off,  automatically  shuts  off,  or  when 

another  measurement  is  initiated.

The  unit  has  the  capacity  to  store  7  readings,  and  

also computes an average of up to the 7 memory 

readings.

  When more than 7 readings are taken and   

recorded  to  memory,  the  oldest  ones  will  be  deleted.

Memory Recall

By pressing the 

MEMORY

 button,  you will see "A" in the 

lower left hand corner of the display.  It means you are 

looking at the average of all the measurement results.  

Pressing the 

MEMORY

 button again, the newest set of 

measurement  results  is  indicated  by  "1".   Press  it  

again,  the  next  set  will  be  displayed.

Memory  Erase

When the batteries are taken out of the monitor, all re-  

cords   will  be  erased.

Note : 

When  taking   a measurement, the 

MEMORY

 button  

is  inactive.

Memory Number

Error indicator

Indication:  Cuff pressure reaches 330 

mmHg.

Cause:       Twisted bladder inside of the 

cuff.

Correction: Make sure that the bladder 

lays flat within the cuff.

Indication:  Display "Err", cuff pressure 

releases and "0" pressure 
display.

Cause:       Excessive arm movement or 

talking during measurement.

Correction: Turn off and on the unit and 

measure again while keeping 
still.

Indication:  Could not obtain pulse rate.
Cause:       Cuff was loosely applied, 

extremely weak pulse or 
arrhythmia.

Correction: Reapply cuff and measure 

again.

Batteries are exhausted.

MS-701K: Replace with four new "AA" batteries.
MS-702:   Replace with four new "AAA" batteries.

Possible sources of errors

Error Indicators

Make sure that the batteries are installed 

in the proper direction. (Polarity matches 
to the indications in the battery case.)

Check that the cuff is positioned properly.
Go over the measurement procedure 

again.

Check if the wrist is positioned properly.
Rest your arm on a table so that the cuff 

is level with your heart during 
measurement.

Remain seated during the entire 

measurement period.

Refrain from hand and body movements 

during measurement.

Your blood pressure can fluctuate 

considerably through out the day.   All of 
the following factors can influence your 
blood pressure:

¡@ Emotional state
¡@ Daily activities or exercise
¡@ Smoking
¡@ Drinking alcoholic beverages
¡@ Eating
¡@ Taking certain medications

Nothing 
appears in 
display

No 
measurement 
occurs

Blood 
pressure 
readings are 
too high or 
low

Blood 
pressure 
readings are 
variable

Troubleshooting

Recommended Action

Problem

Содержание MS-701K

Страница 1: ...R R MS 702 MS 701K Semi Auto Digital Blood Pressure Monitor R R...

Страница 2: ...ethod of Measurement 5 Parts Identification 6 Description of Display Marks 6 Preparation before Taking a Measurement 7 Taking a Measurement 8 Memory Function For MS 701K Only 9 Err Indicators Troubles...

Страница 3: ...y of the device has been verified and conforms to the provisions of the EC council directive 93 42 EEC of 14 June 1993 on medical devices as well as the EMC directive 89 336 EEC EN 1060 1 Non invasive...

Страница 4: ...ated mmHg For example an individual s blood pressure may be mea sured as 120 mmHg systolic and 80 mmHg diastolic This would be spoken as 120 over 80 and written as 120 80 Remember that blood pressure...

Страница 5: ...inuously throughout the day Try to take your blood pres sure at the same time each day 5 Do not be concerned with the results of one meas urement Many measurements recorded over a long period of time...

Страница 6: ...relax for 5 to 10 minutes before taking another measurement 7 Many factors can affect your blood pressure such as exercising eating talking moving nervousness environment and temperature changes Emoti...

Страница 7: ...ing Low Battery Inflating Deflating Monitor Accessories MS 702 2 3 4 LCD Display Air Tube Socket ON OFF Button MEMORY Button MS 701K 1 2 3 1 2 3 4 Exhaust Value Push Button Rubber Bulb Cuff Hose Arm C...

Страница 8: ...ur wrist 4 Any gap between the wrist and cuff will affect the result of measurement 5 The fabric cuff can be folded down for ease of viewing the display Battery Installation Replacement 1 Slide the ba...

Страница 9: ...symbol being displayed that means you need to inflate the cuff to a higher pressure than you did the first time If you do not see this symbol it means it is taking your measurement Please remain calm...

Страница 10: ...to the 7 memory readings When more than 7 readings are taken and recorded to memory the oldest ones will be deleted Memory Recall By pressing the MEMORY button you will see A in the lower left hand co...

Страница 11: ...tors Make sure that the batteries are installed in the proper direction Polarity matches to the indications in the battery case Check that the cuff is positioned properly Go over the measurement proce...

Страница 12: ...11...

Страница 13: ...atic deflation Arm size ranges 22 33 cm 9 13 inches Batteries MS 701K 1 5V LR6 AA x 4 MS 702 1 5V LR3 AAA x 4 Automatic power off Approx 3 minutes after measurement Reference method for clinical trial...

Страница 14: ...13 Date Time Pulse Body Condition 1 1 1 1 1 1 60 80 100 120 140 160 180 200 220 240 mmHg 128 134 123 84 90 76 70 73 69 7 00 13 30 20 00...

Страница 15: ...14...

Страница 16: ...R R MBM MS1U9 ED3...

Отзывы: