Page 60
Servicing Instructions 227 470 35 B - 97.12
Control lights (LED)
For mains voltage applied, battery condition and Start/Stop function.
Lead selection
Manual selection of individual lead groups or automatic stepping keys for the lead groups using
the "Auto" mode.
*
Lead programs:
EINTHOVEN, GOLDBERGER, WILSON and CABRERA lead sequence
Automatic modes
To support and simplify operation by:
*
Automatic blocking function
*
Automatic control, of lead selection, paper feed, calibration
*
Formated output
*
Automatic zero position
*
Anti-drift system to compensate fluctuations in polarization voltage
Detection of pacemaker pulses
*
Pulse duration between 0.1 and 2.5 ms
*
Marking irrespective of sign
*
Pulse amplitude between ± 5 mV and ± 700 mV
Heart rate indication
Determining the heart rate from all ECG signals.
*
Display range 30 ... 300 P/min
*
Display cycle with each heart beat, max. 2 s
Signal inputs
Insulated patient connection type CF in compliance with IEC; all pickup electrode connections
and N-connectors high voltage resistant (only in connection with original Marquette Hellige
patient lead), interference compensation via neutral electrode (N), electrode monitoring for
breaks.
*
Electrode connections for R, L, F, N, C1 ... C6
*
Input impedance for differential signals between any two electrode connections at 10 Hz
> 10 MOhm
*
Input impedance for common-mode signals against N up to 60 Hz > 50 MOhm
Содержание MicroSmart
Страница 1: ...Servicing Instructions MicroSmart Version 2 xx 227 470 35B SA e ...
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 78: ......
Страница 79: ......
Страница 88: ......
Страница 89: ......
Страница 90: ......
Страница 91: ......