background image

7.

 

The Insulation Resistance Tester Structure 

 

 

1.

 

Insulation Resistance Test/ Lock button 

2.

 

Relative & Time-down button 

3.

 

Data Hold button 

4.

 

Internet Battery Check & Time-up button 

5.

 

MAX/MIN & Insulation resistance test emergency stop button 

6.

 

LCD display 

7.

 

DC 13.5V Power adapter Input Terminal 

8.

 

EARTH: High resistance measurement input 

9.

 

V/Ὠ input terminal & GUARD: Grounding protection input terminal 

10.

 

COM & High voltage line shielding input terminal 

11.

 

LINE:500V to 5000V High voltage output terminal 

12.

 

PI/DAR, Peak Voltage measurement & Time Clear button 

13.

 

Back-light Levels button 

14.

 

Function switch 

15.

 

Test leads 

 

 

Содержание MIR-505

Страница 1: ...HIGH VOLTAGE INSULATION TESTER Product Code MIR 505 USER MANUAL WWW MARMONIX CO HIGH VOLTAGE INSULATION TESTER...

Страница 2: ...rement Continuity Check D Measuring Insulation Resistance 9 Auto discharge function 10 Principal of Insulation Resistance Measurement A Continuous Insulation Resistance Measurement B Timer Insulation...

Страница 3: ...ory II Equipment of Overvoltage Category II is energy consuming equipment to be supplied from the fixed installation Note Examples include household office and laboratory appliances Overvoltage Catego...

Страница 4: ...Resistance tester with ant parts or cover removed When carrying out insulation measurement do not come in contact with the circuit test Warning Do not use the Insulation resistance tester if it is da...

Страница 5: ...When the instrument will not be in use for a long period place it in storage after removing the batteries Symbols Risk of electric shock Equipment protected by double or reinforced insulation DC measu...

Страница 6: ...or at nighttime work Live circuit warning symbols plus audible warning Live circuit detection prevents insulation test if voltage 30V is detected for added user protection With auto power off functio...

Страница 7: ...load 1 to 1 2mA at 1M load 1 to 1 2mA at 2 5M load 1 to 1 2mA at 5M load Recommended use of Power Adaptor Short Circuit Current Approx 1mA Accuracy 2 5 reading 15digits at 0 005 to 600 0M 3 reading 15...

Страница 8: ...ts 601 to 6 00k 0 001k 1 5 reading 15digits Continuity Buzzer 50 Short Circuit Current 200mA Open Circuit Test Voltage 4 5V DC AC Voltage Measurement Range Resolution Accuracy DC Voltage 0 5 to 600 0V...

Страница 9: ...egory II Pollution Degree 2 Altitude up to 2000 meters Indoor Use only Power source 8 x 1 5V R14 batteries Operating conditions 0 to 40 C 32 to 104 F 80 RH Storage conditions 10 to 60 C 14 to 140 F 80...

Страница 10: ...ce test emergency stop button 6 LCD display 7 DC 13 5V Power adapter Input Terminal 8 EARTH High resistance measurement input 9 V input terminal GUARD Grounding protection input terminal 10 COM High v...

Страница 11: ...display reading Measurement Inputs Terminal Description Insulation resistance measurement LINE 500V 1000V 2500V 5000V range Insulation resistance measurement high voltage line output LINE COM two plug...

Страница 12: ...utomatically at any set time Time down Set timer time range down This is a function to conduct a test automatically at any set time PI Polarization index measurement press 2sec Clear timer function DA...

Страница 13: ...o not touch it A Measuring Voltage DANGER To avoid electric shock please do not test if voltage to ground AC DC is over 600V even if the line voltage is below 600V Under voltage measurement please not...

Страница 14: ...Low Resistance Measurement continuity Check Warning Do not run this test unless ACV DCV 0 Do not use this mode to check diodes 1 Set the Rotary switch to the Position 2 Connect the red test lead to t...

Страница 15: ...ESS TO TEST button is pressed To check the insulation of electric equipment s or electric circuits measure their insulation resistance with the instrument Be sure to check the voltage which can be app...

Страница 16: ...ouch the circuit under test immediately after testing Capacitance stored in the circuit may cause electric shock Leave test leads connected to the circuit and never touch the circuit until the dischar...

Страница 17: ...in to 1 min 3 Pressing and hold turning the PRESS TO SET button clockwise to perform a timer measurement of insulation resistance Then the button is locked and timer measurement can be performed 4 Aft...

Страница 18: ...seconds NOTES The following for reference only According to the standard IEEE 43 2000 with respect to the motor and related equipment for trouble shooting it will be for the insulation resistance tes...

Страница 19: ...o repair or replace parts 12 DAR Standard DAR value 1 6 or more 1 25 1 6 1 25 or less Tested material status Excellent PASS Fail 13 Use of Guard Terminal When measuring a cable leakage current flowing...

Страница 20: ...y the present reading but will continue to update and store the max and min readings 5 To exit MAX MIN PEAK mode press and hold MAX MIN for 2 seconds or presses PEAK MAX MIN Button Only for AC DC volt...

Страница 21: ...on Press the key to turn ON OFF and select backlight levels The backlight will automatically turn off in 0 seconds 15 Auto Power Off To reserve battery power the tester enters into the Sleep Mode and...

Страница 22: ...ejoin the case bottom and battery compartment and reinstall the screw 17 The Use of Power Adaptor Make sure the meter is power off and insert the power adaptor to the input terminal It is highly recom...

Страница 23: ...Service Periodically wipe the case with a damp cloth and mild detergent Do not use abrasive or solvents To clean the terminals with cotton bar with detergent as dirt or moisture in the terminals can...

Отзывы: