SECTION 5 ASSEMBLY
DOOR ADJUSTMENT
The Door Adjustment is Located in the
Fulcrum Casting at the base of the door
opening. This adjustment employs the use
of a screw and locknut In order to adjust
the Sterilmatic Door to a tighter closed
position (to prevent steam from leaking by
the door gasket as pressure builds up), it
is necessary to loosen the locknut and
back off the screw at least one-quarter of
a turn and re- tighten the locknut.
THE DOOR GASKET
Keep the gasket clean. With normal
closing and locking of the door assembly,
a steam-tight seal should be made
between the door gasket and the door opening. This seal cannot be maintained if particles of
foreign matter are allowed to accumulate upon either of the contacting surfaces.
If there is leakage by the door gasket before a steam build-up within the cooking chamber
causes it to stop, regard the door assembly as improperly adjusted. A re-adjustment must
then be made of the door screw.
To change the door gasket, remove the entire door assembly as a unit. Discard the old
gasket, replace it with a new one (no cement is required), and reinstall the door assembly.
Make an operational check for leakage and adjust the door, if necessary.
DOOR LIFT SPRING
Market Forge supplies door lift springs in sets only. This policy has been found to be in the
best interest of the customer. Through continuous use, some of the original qualities of the
springs are lost and it becomes advantageous to make replacements to both the left and
right door lift springs in the event that one becomes damaged or broken.
Replacement door lift springs are marked with tabs at the factory prior to shipment to identify
a right from a left spring. These springs must be installed with the right door lift spring on the
right of the door and the left door lift spring on the left of the door as viewed from the front of
the sterilizer.
Содержание STM-E
Страница 2: ...35 Garvey Street Everett Massachusetts 02149 ...
Страница 6: ......
Страница 7: ...5 SECTION 1 INSTALLATION INSTRUCTIONS ...
Страница 10: ...8 SECTION 1 INSTALLATION INSTRUCTIONS ...
Страница 11: ...9 SECTION 1 INSTALLATION INSTRUCTIONS ...
Страница 12: ...10 SECTION 1 INSTALLATION INSTRUCTIONS TYPICAL CIRCUIT CONNECTION FOR STM E AND STM EL ...
Страница 13: ...11 SECTION 1 INSTALLATION INSTRUCTIONS TYPICAL CIRCUIT CONNECTION FOR STM EX AND STM ELX EXPORT MODELS ...
Страница 19: ......
Страница 21: ......
Страница 22: ......
Страница 28: ...26 SECTION 6 ILLUSTRATED PARTS ...
Страница 29: ......
Страница 30: ...28 SECTION 6 ILLUSTRATED PARTS ...
Страница 31: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......