SECTION 1 INSTALLATION INSTRUCTIONS

INSTALLATION FOR OPTIONAL RECORDING THERMOMETER

The Optional Recording Thermometer may be surface or panel mounted as best accommodates the layout of

the sterilization area.

1. Remove the top cover. Remove the pipe plug located in the Y-fitting that also accommodates the

safety relief valve and which is connected directly to the sterilizing chamber.

2. Insert the thermometer bulb of the temperature recorder into the pipe, from which the plug was

removed.

3. Tighten the fittings. Bend that portion of the Thermometer bulb coil that extends inside the shell

toward the rear of the cylinder and at an angle of 45° so that it does not interfere with mounting

the overhead baffle plate (Item #20 Master Illustration).

4. For mounting the recorder on a wall or on the unit itself, the 3/16" diameter tapped holes in the

angles which are attached to the recorder must be drilled out. This allows for the use of wood or

sheet metal screws on walls or panels, and nuts and bolts on the unit itself.*

5A. Electrical connections to the recorder are made by running a line from the recorder to any 120

volt supply. Connection can also be made at the supply connection for the unit by utilizing the

proper wires to gain 120 volts. Connect either L 1, L2, or L3 for power lead (black wire and the

neutral white wire) line for the other side of the 120 volt line.

5B. Electrical connections to the recorder are made by running a line from the recorder to any 220

volt supply. Connection can also be made at the supply connection for the unit by utilizing the

proper wires to gain 220 volts. Connect either L 1, l2, L3 for power lead (black wire and the

neutral white wire) line for the other side of the 220 volt line.

6. The United Electric Recording Thermometer is a 24 hour clock and should be run continuously.

Charts should be changed every 24 hours.

"For further mounting instructions. see the Manufacturers Instruction and Maintenance Instructions, Form #17-0099."

TO REMOVE THE CHART

1. Unscrew and remove the chart knob and lift the pen by depressing the pen lifter.

2. Grasp edge of the chart, lift off the chart hub and slide downward and out without permitting the

paper to touch the pen.

TO INSTALL NEW CHART

1. Slide edge of the chart under the pen and over the hub, insert the upper edge under the

nameplate chart guide and seat the chart hole on the hub.

2. Set correct chart time to the chart plate time index line and replace the knob.

PEN ADJUSTMENT

- If the pen point is not aligned properly with the temperature to be recorded (

as

checked by an ordinary thermomete

r), it may be corrected in the following manner:

1. Locate the small fillister head screw on the right side of the extreme upper end of the Pen

Assembly.

2. Turn the screw clockwise for pen movement toward the hub (or counter-clockwise for pen

movement away from the hub) until the temperature value of the ink line has been adjusted to its

proper position.

PART NO.

10-7933

REPLACEMENT 24 HOUR THERMOMETER CHART FAHRENHEIT

PART NO.

08-6244

REPLACEMENT 24 HOUR THERMOMETER CHART CELSIUS

Содержание STM-E

Страница 1: ...uaranteed Alfa Medical 1 800 762 1586 265 Post Ave 516 280 7822 Westbury NY 11590 516 280 7832 fax www sterilizers com eMail sterilizers com The attached manual is for your records Go to the below web site to look for parts http bit ly Market Forge Sterilmatic Parts ...

Страница 2: ...35 Garvey Street Everett Massachusetts 02149 ...

Страница 3: ...ND 17 STERILMATIC DOOR ASSEMBLY 17 DOOR ADJUSTMENT 18 THE DOOR GASKET 18 DOOR LIFT SPRING 18 TO REMOVE THE DOOR ASSEMBLY 19 DOOR ASSEMBLY PARTS LIST 19 THE FULCRUM AND DRAIN ASSEMBLY 20 FULCRUM AND DRAIN PARTS LIST 20 ROLLER ASSEMBLY 21 DOMESTIC THE PRESSURE ACTUATED TEMPERATURE CONTROL STM E AND STM EL 21 DOMESTIC HOW IT WORKS 21 DOMESTIC ADJUSTING THE RANGE OR OPERATION STM E 21 EXPORT THE PRESS...

Страница 4: ... left side of unit Whether the supply current is 208 volt or 240 volt single or three phase all control circuits are 120 volts In order to accomplish this connection a current carrying grounded neutral must be provided Thus a three phase system must be 4 wires and a single phase system must be 3 wires If a current carrying grounded neutral is not available from the power source a separate 120 volt...

Страница 5: ...oration at the rear of the upper tray support channel OPERATION CHECK To check for proper operation of unit 1 Close drain valve by turning handle clockwise WARNING DO NOT OPEN DRAIN VALVE WHILE UNIT IS OPERATING PREMATURE OPENING MAY RESULT IN SCALDING OF OPERATOR 2 Fill chamber with 4 to 6 quarts 3 7 to 5 6 liters of ordinary tap water DO NOT USE DISTILLED OR DIONIZED WATER 3 Close chamber door 4...

Страница 6: ......

Страница 7: ...5 SECTION 1 INSTALLATION INSTRUCTIONS ...

Страница 8: ... key way slot Slide the rack down into its correct position See Fig 1 Fig 1 4 After installing one pan support rack correctly you can install the baffle Position the baffle so that the 45o bend is facing up and towards the front of the unit Slide the mounting tab onto the flat bend on the pan stop bracket The baffle should now stay in place by itself but in a tilted state See Fig 2 Fig 2 5 Positio...

Страница 9: ...om the power source a separate 120 volt circuit must be run Most electrical codes require and we recommend that a separate disconnect switch be located within sight of the sterilizer When a separate 120 volt control circuit must be run this must also be part of the disconnect box assembly THE ELECTRIC SUPPLY CONNECTIONS FOR STM EX AND STM ELX EXPORT Connect to proper electrical supply as indicated...

Страница 10: ...8 SECTION 1 INSTALLATION INSTRUCTIONS ...

Страница 11: ...9 SECTION 1 INSTALLATION INSTRUCTIONS ...

Страница 12: ...10 SECTION 1 INSTALLATION INSTRUCTIONS TYPICAL CIRCUIT CONNECTION FOR STM E AND STM EL ...

Страница 13: ...11 SECTION 1 INSTALLATION INSTRUCTIONS TYPICAL CIRCUIT CONNECTION FOR STM EX AND STM ELX EXPORT MODELS ...

Страница 14: ... by running a line from the recorder to any 220 volt supply Connection can also be made at the supply connection for the unit by utilizing the proper wires to gain 220 volts Connect either L 1 l2 L3 for power lead black wire and the neutral white wire line for the other side of the 220 volt line 6 The United Electric Recording Thermometer is a 24 hour clock and should be run continuously Charts sh...

Страница 15: ...s follows These conditions have been highlighted in our documentation Grains of hardness in the water supply should be as follows 25 to 2 a pH imbalance in the water supply can greatly affect the life to the aluminum cylinder The pH range that would be recommended is between 7 0 8 5 The lack of a positive electrical ground can cause an electrolytic reaction that will accelerate pitting Another con...

Страница 16: ...In no case should the timer be set to less than 15 minutes Sterilization will not be accomplished in less than 15 minutes exposure time 6 TURN TIMER Located at upper right front of sterilizer Select desired length of sterilizing period This turns power supply on and starts the cycle after pressure temperature combination has been reached Amber pilot light indicates that the timer is running 7 When...

Страница 17: ...nt of air and complete contact of steam to surfaces Drying is also facilitated PETRI DISHES PIPETTES DESICCATORS etc Should be inverted UTENSILS TREATMENT TRAYS Placed on edges to facilitate drying INSTRUMENT SETS Place instruments set in trays having mesh or perforated bottoms Place trays flat on shelves COMBINING FABRICS HARD GOODS Place hard goods on lowest shelves PLASTIC UTENSILS DO NOT stack...

Страница 18: ...muslin or paper 45 Flasked solutions 1500 2000 ml SECTION 4 DAILY CLEANING DAILY CLEANING PROCEDURE AT THE END OF THE DAY a Remove bottom splash baffle NOTE IMPORTANT STERILIZING CHAMBER MUST BE CLEANED AND DRAINED DAILY USING THE FOLLOWING PROCEDURE WASH WETTED PORTION OF THE CYLINDER THOROUGHLY BY ADDING A MILD DETERGENT TO WATER IN CYLINDER b If a soft cloth or brush is used with the detergent ...

Страница 19: ......

Страница 20: ...auses it to stop regard the door assembly as improperly adjusted A re adjustment must then be made of the door screw To change the door gasket remove the entire door assembly as a unit Discard the old gasket replace it with a new one no cement is required and reinstall the door assembly Make an operational check for leakage and adjust the door if necessary DOOR LIFT SPRING Market Forge supplies do...

Страница 21: ......

Страница 22: ......

Страница 23: ...ure at 13 PSI controlling the two outer banks of heating elements Switch 2 is set to cut out on rising pressure at 15 5 PSI and cut in on falling pressure at approximately 15 PSI controlling center heater element only Thus a balancing effect of steam pressure build up and heater element current is constantly maintained during the sterilizing cycle When the Timer signifies the end of the cycle the ...

Страница 24: ...ssure at approximately 1 05 kg cm 2 controlling center heater element only Thus a balancing effect of steam pressure build up and heater element current is constantly maintained during the sterilizing cycle When the Timer signifies the end of the cycle the electric current to the contactors will automatically be broken CHECKING THE 110 C SETTING STM ELX The Unit should be completely evacuated then...

Страница 25: ...eld The elements cannot be removed and in the unlikely event that one or all fail the complete cylinder must be replaced THE LOW WATER CUT OFF MANUAL RESET Fastened to a special mounting brace behind the front panel the Low Water Cut Off acts to shut off the complete unit should the water run dry The Low Water Cut Off is factory set to shut the unit off when the cylinder temperature rises between ...

Страница 26: ... steam which is within the sterilmatic sterilizing chamber To replace this unit it is necessary to disconnect the copper tubing and to remove the two nuts holding the gauge framework in place THE SAFETY VALVE The Safety Valve is factory set to automatically open and exhaust excess steam from within the sterilizing cylinder thereby assuring that operating pressures remain within safe limits The lev...

Страница 27: ...le of sterilization Also a slot is milled at an angle through the seat to allow a constant bleed off of a slight amount of steam during the cycle to completely eliminate any air pockets in the cylinder see A in figure Failure of the trap to operate properly will result in an uneven distribution of live steam within the compartment HOW IT WORKS With the introduction of steam into the sterilizing co...

Страница 28: ...26 SECTION 6 ILLUSTRATED PARTS ...

Страница 29: ......

Страница 30: ...28 SECTION 6 ILLUSTRATED PARTS ...

Страница 31: ......

Страница 32: ......

Страница 33: ......

Страница 34: ......

Страница 35: ......

Страница 36: ...fective This warranty is effective for 90 days in regard to labor This warranty does not apply to damage resulting from use of hard or corrosive water from failure to drain and dry cylinder daily or from inadequate cleaning procedures nor does it cover any part or assembly which has been subjected to accident alteration or is from a machine on which that serial number has been removed or altered N...

Отзывы: