11.

 

Check ADC and Set To Factory Setting 

Operation 

Display 

Descriptions 

Press

 

000000

 

Go to Calibrations and Configurations menu and enter 
password 002003 

Press

 

Or

T

 

002003

 

To enter password 002003: 

 

Press

to move blinking digit to the right 

 

Press

T

to increment blinking digit 

 

Press

0

to decrement blinking digit 

Press

  * 

 

CAL SP

  In Calibrations menu 

Press

  * 

 

-SEt- 

 

In Configurations menu 

Press

  * 

 

-A-d- 

 

In Check ADC and Set To Factory menu 

Press

 

365235

 

ADC count 

Press

  * 

 

    0.00

 

Current weight 

 

-A-d- 

 

Operate step 1 to step 5 again 

Press

  * 

 

FACt   

  Set to factory setting 

Press

 

    0.00

 

Indicator is set to factory setting: 
d=0.001    FS=3.000    FLt=2    AUtP=10    Adr=00 
b=2400    L=00749    H=03723    gAIn=1    rAtE=1 

12.

 

Setting set point 

Operation 

Display 

Descriptions 

Press

 

000000

 

Go to Set Point menu and enter password 110906 

Press

 

Or

T

 

110906

 

To enter password 110906: 

 

Press

to move blinking digit to the right 

 

Press

T

to increment blinking digit 

 

Press

0

to decrement blinking digit 

Press

  * 

 

A1       

 

Set point #1 for relay #1 

Press

 

000150

 

Press

to access set point #1 

Press

 

Or

T

 

000200

 

Press a combinations of

,

T

and

0

to 

set point A1 to 000200 

Press

  * 

 

A2       

 

Set point #2 for relay #2,   
Follow step 4 and 5 to change set point 

Press

  * 

 

A3       

 

Set point #3 for relay #3 
Follow step 4 and 5 to change set point 

Press

  * 

 

A4       

 

Set point #4 for relay #4 
Follow step 4 and 5 to change set point 

Press

  * 

 

A5       

 

Set point #5 for relay #5 
Follow step 4 and 5 to change set point 

Press

  * 

 

A6       

 

Set point #6 for relay #6 
Follow step 4 and 5 to change set point 

Press

  * 

 

OUt-d 

 

Set state of relay output 

Press

 

010001

 

The six bits in OUt-d from left to right represent the 
open/close state of the relays from #6 to #1 
Example 1

OUt-d=010001 ( relay #5 bit is 1), A5=005000 

Содержание TD 2014 Series

Страница 1: ...User Manual TD 2014 Series Weighing Controller Applications Electronic platform scale weighing scale etc ...

Страница 2: ... Size 95 48 110 mm Weight 0 5 kg Signal output 4 20 mA analog output Options a Opto isolated RS232 or RS485 b 6 relay outputs with 2 opto isolated inputs that can be used to control the output of relay 1 and 2 2 Model Number TD 2014 0 No interface 1 Optical isolated RS232C 2 Optical isolated RS485 AC AC 110 220V 50 Hz available for XK315A2 6A series only DC DC 9V 1A adapter A Basic model with 4 20...

Страница 3: ...ntinuous ASCII regular data output if weight is 123 45 the output is 123 45 3 Adr 01 98 data output using ModBus protocol 8 Operation 1 Turn on the indicator it displays the version number and checks the LED from 0 to 9 2 Zero When the weight is stable between 2 FS press 0 zero the display and 0 is on When the zero reference is 2 FS it displays HHHHHH When the zero reference is 2 FS it displays LL...

Страница 4: ... to select filter Press AUtP00 Set zero trace range and zero at start range AUtP 00 no zero set at start zero trace range is 0 4d AUtP 10 zero at start range is 20 FS zero trace range is 0 4d AUtP 20 zero at start range is 20 FS zero trace range is 0 8d AUtP 30 zero at start range is 20 FS zero trace range is 1 2d AUtP 40 zero at start range is 20 FS zero trace range is 1 6d AUtP 50 zero at start ...

Страница 5: ...a combinations of T and 0 to change the magnification Press rAtE 1 Set AD sample rate 0 10 times second 1 40 times second 2 250 times second Press Or T rAtE 2 Press a combinations of T and 0 to change the sample rate Press 0 00 Save all parameters 10 Calibrations Operation Display Descriptions Press 000000 Go to Calibrations and Configurations menu and enter password 002003 Press Or T 002003 To en...

Страница 6: ... 000000 Go to Set Point menu and enter password 110906 Press Or T 110906 To enter password 110906 Press to move blinking digit to the right Press T to increment blinking digit Press 0 to decrement blinking digit Press A1 Set point 1 for relay 1 Press 000150 Press to access set point 1 Press Or T 000200 Press a combinations of T and 0 to set point A1 to 000200 Press A2 Set point 2 for relay 2 Follo...

Страница 7: ...g Press En o5 Set the output of 5 relay Follow step 14 to 16 to change relay output setting Press En o6 Set the output of 6 relay Follow step 14 to 16 to change relay output setting Press 0 00 Save set points 13 Software version VER2 01 packing scale VER2 02 3 speed rating scale VER2 03 check weighing VER2 04 2 materials 2 speed batching scale VER2 05 3 materials 2 speed batching scale VER2 06 4 m...

Страница 8: ...7 maintenance and calibration facilities 16 Relay output ...

Отзывы: