Maquet Datascope CS300 Скачать руководство пользователя страница 4

 | 

2

 | 

CS300

 

Abbreviated Operator’s Guide

 |

IAB Sensor Low Level Pressure Output – to Bedside Monitor

This output connector is used to provide an electrical signal from the IAB Sensor 
module to the physiologic pressure transducer input on the patient’s bedside monitor 
when a Sensor IAB is connected and calibrated. Its low level output emulates that of a 
typical Arterial Pressure Transducer, allowing direct connection to the patient inputs of 
the bedside monitor when a Sensor IAB is connected and calibrated.

IAB Sensor Output – Vent

This button facilitates the zeroing process required for proper calibration of the 
physiologic pressure channel of an attached bedside monitor. When pressed, the IAB 
Sensor Output signal will simulate that of an atmospheric vented transducer. Venting 
will be simulated for 15 seconds to give users adequate time to zero the pressure 
channel on the bedside monitor.
The pressure signal will then be restored automatically to permit monitoring  
through the bedside and centralized monitoring systems.

OPERATING MODES

AUTO Operation Mode

All aspects of pump Operation are automated

n

  Most appropriate trigger source is automatically selected.

n

  Inflation and deflation timing is automatically set.

n

  Lead I, II, III, or External ECG source is automatically selected.

n

   System searches for the next best trigger source in the event of a loss of 

trigger source.

n

   Timing adjusts automatically if trigger source changes, heart rate changes, 

rhythm changes.

n

   With sustained unpredictable rhythms, R-wave deflation is automatically selected. 

The message 

“Auto R-Wave Deflate”

 will be displayed. The deflation indicator 

will automatically be repositioned to the far right, “R-Trac” will be displayed in the 
IAB deflation indicator. The system will return to predictive timing once the rhythm 
becomes predictable.

SEMI AUTO Operation Mode

n

   Operator selects the most appropriate trigger source.

n

    Operator establishes initial timing and thereafter, software algorithms automatically 

track changes in patient heart rate or rhythm and adjusts timing accordingly.

n

   Changing the trigger source will cause the pump to go to Standby. Pumping will 

resume when the 

START

 key is pressed.

n

   Lead I, II, III, AVR, AVL, AVF, V or External ECG source can be selected.

VAULT COPY

Содержание Datascope CS300

Страница 1: ...DATASCOPE IS NOW MAQUET CARDIOVASCULAR THE CS300 ABBREVIATED OPERATOR S GUIDE VAULT COPY...

Страница 2: ...CS300 Bezel and Keypad The CS300 provides comprehensive HELP SCREENS with easy to follow step by step instructions on set up as well as alarm and advisory conditions VAULT COPY...

Страница 3: ...nsor IAB Calibration On initial insertion pressing the START key automatically purges and fills the IAB and calibrates the Sensor IAB if connected While assisting press and hold the Zero Pressure key...

Страница 4: ...ATING MODES AUTO Operation Mode All aspects of pump Operation are automated n Most appropriate trigger source is automatically selected n Inflation and deflation timing is automatically set n Lead I I...

Страница 5: ...d IAB assist is desired For example whenever circulatory bypass or a laminar flow left ventricular assist device is in use Trigger Source n Trigger is the signal the CS300 uses to identify the beginni...

Страница 6: ...he A V interval is between 80 224 msec and the pacing rate is less than 125 beats per minute Pacer A n This trigger mode is only available in the SEMI AUTO and Manual Operation modes n The R wave of t...

Страница 7: ...eration mode n Use the IAB Deflation arrows to adjust the end of the highlighted segment prior to ventricular ejection After initiating assist the IAB INFLATION only operational in the Semi Auto and M...

Страница 8: ...is sensed In SEMI AUTO Operation mode if the ectopic beat is of small amplitude reliable triggering can be maximized if an ECG Lead is selected which minimizes the amplitude difference between the nor...

Страница 9: ...energy coupled from the ESU to the ECG input of the CS300 generally resulting in stable ECG triggering When the CS300 is in AUTO Operation mode it will automatically select Pressure Trigger when ESU i...

Страница 10: ...or ALARM messages are displayed in the ALARM MESSAGES section of the monitor display HIGH PRIORITY ALARMS High Priority Alarms indicate situations that require immediate Operator response Pumping is s...

Страница 11: ...o improve ECG signal by ensuring electrode contact and optimal placement n If signal problem persists switch Operation mode to SEMI AUTO Verify trigger source n Use IAB Inflation and IAB Deflation con...

Страница 12: ...rigger is required and manual threshold is selected reduce the pressure trigger threshold by using the PRESSURE THRESHOLD sub menu from the PUMP OPTIONS menu n If appropriate select a different trigge...

Страница 13: ...M The IAB fill mode has been changed from Manual to AUTO without an autofill Operator Intervention n Inspect the IAB and catheter extender tubing for any evidence of a leak including the safety disk c...

Страница 14: ...Check integrity of all cable lead connections n Attempt to improve Arterial Pressure by checking integrity of all cable connections If conventional transducer in use verify the transducer was not lef...

Страница 15: ...no fill had been performed Verify Proper Timing SA M Operator has changed the trigger source Manual Fill IAB SA M Notifies the operator when to manually fill the IAB catheter with shuttle gas Display...

Страница 16: ...ed the Semi Auto Operation mode Maintenance Required Code ___ System maintenance may be required Low Helium Helium supply is below a preset reserve as determined by tank pressure Prolonged Time in Sta...

Страница 17: ...elect an alternative item use the UP and DOWN arrow keys 24hr Emergency Helpline 1 800 777 4222 Sweep Speed Balloon Waveform Brightness ECG Inflation Markers Alarm Volume Standby Advisory Tone Beep Vo...

Страница 18: ...ous printout hold the PRINT STRIP key for two 2 seconds The associated LED will blink as printing continues To stop the continuous printout press the PRINT STRIP key again Doppler Information The dopp...

Страница 19: ...p the IAB Catheter Extender Input when the message PLUG DISK OUTLET appears on the display If the message UNPLUG DISK OUTLET appears remove any connections that may exist to the IAB Catheter Extender...

Страница 20: ...The condensate removal module will not operate to conserve remaining power Battery Charging To charge the internal battery 1 Leave the system power cord plugged in and set the MAINS On Off to ON 2 Ch...

Страница 21: ...t and that a noticeable pressure drop has not occurred NOTE Prior to using the system in any other position except vertical consult with Technical Service NOTE For information on outfitting an aircraf...

Страница 22: ...off the cart To remove the console by detaching the battery pack With the console on the cart lift the battery release levers to the unlock position fig 2 Lift the console straight up and off the cart...

Страница 23: ...e by pressing the button on the rear of the monitor fig 4 Attach the monitor to the top of the pump console Make sure the monitor is securely attached before moving the system Extend the console handl...

Страница 24: ...o technical development Caution U S Federal Law restricts this device to sale by or on the order of a physician Refer to Instructions for Use for current indications warnings contraindications and pre...

Отзывы: