1
|
|
Abbreviated Operator’s Guide
|
CS300
™
|
QUICK REfERENCE – INITIAL SET-UP
1. Establish power and verify the
MAINS ON/Off
and the
IABP ON /Off
switches are
on.
2. Open the Helium Tank.
3. Establish ECG and Pressure.
4. If using a Sensor IAB, ensure that the IAB sensor cable (orange) has been
connected.
*
Otherwise, zero the transducer
**
:
n
Open the transducer to air
n
Press the
zERO PRESSURE
key for 2 seconds
n
Close the transducer
5. Confirm that the operation mode is
AUTO
.
6. Attach the IAB catheter and the appropriate extender to the safety disk.
7. To initiate pumping, press the
START
key. In response the IABP will Autofill and then
begin pumping. If desired, IAB DEFLATION timing can be fine-tuned using the IAB
DEFLATION controls.
8. Verify the setting of the AUG. ALARM:
n
Verify that the AUG. ALARM setting is approximately 10 mmHg less than
the patient’s diastolic augmented pressure.
n
If needed, adjust by pressing the AUG ALARM key and using the NAVIGATION
ARROW keys to change the value displayed on the screen.
9. Initial setup is now complete.
*
for Sensor IAB Calibration:
–
On initial insertion, pressing the START key automatically purges and fills the IAB and
calibrates the Sensor IAB if connected.
–
While assisting, press and hold the Zero Pressure key for two (2) seconds to
perform an operator-initiated calibration of the Sensor IAB. The displayed blood
pressure parameters from the IAB Sensor are only valid if the IAB Sensor is properly
calibrated.
To prevent calibration of the IAB Sensor at inappropriate times, the Zero Pressure key is
disabled when:
n
The IABP is in
STANDBy
.
n
The IAB Sensor is disconnected.
n
The
Pressure Source
is set to External.
n
The IAB Fill Mode is set to Manual Fill.
**
for conventional IAB/Transducer:
– Zero the transducer as noted above
To prevent zeroing the transducer at inappropriate times, this key is disabled when:
n
Pulsatility is detected on the AP waveform.
n
The transducer is disconnected.
n
The
Pressure Source
is set to
External
.
NOTE:
A connected and functional Sensor IAB takes priority over an arterial pressure
transducer, as the Direct pressure source. The Sensor IAB must be disconnected to
activate the arterial pressure transducer input. The pump automatically searches for a
viable pressure source when multiple sources are available.