Wartungs- und Reparaturanleitung
Alphaclassic / Betaclassic
Maintenance and Repair Instruction
1118
1118.0XGB_ 06
Page: 18
5. Repairwork, Replacement of hydraulic and mechanical components
Replacement of the column casings and the bellows
Bellows
Lock the O.R. table on the floor.
Remove all screws (4) of the seat plate (3).
Remove the seat plate (3).
Remove 6 screws (7).
Secure the column casings to prevent it from dropping
Remove all the rivets (1) of the bellows (2) .
Lower the casings and remove the bellows.
Column casing
Bend the 2 clips (10) of the uppermost segment of the casings (5) to enable removal of the bracket
(11).
Remove the uppermost segment of the casings.
Disassemble the centre part of the casings in the same order.
Remove the 3 screws (8) of the lower casing part.
Disassemble the lower part of the casings as described above.
Removal of the base casing:
Remove 4 fixation screws (12).
The base casing (9) can now be removed from the base in an upward direction.
Plase note:
To replace the base casing the table top must be separated from the column – see 5.2
(column guiding).
5.1