3.Table top components
3.1 Head rest 1130.64X0
To fix the head rest (
1
), insert the two pins into the locat-
ing holes until they fully engage. Then tighten the two
fastening screws (
17
).
Check to see if the head rest is attached securely
by pulling it.
3.2 Adjusting the head rest
The head rest can be inclined upwards by pulling the end
of the plate up. The head rest engages automatically step
by step.
To incline the head rest downwards, pull the operating
lever (
20
) towards the padding. The head rest is auto-
matically located in the next ratchet after you release the
operating lever (
20
).
Î
Make sure you comply with the Operating instruc-
tions for the head rest.
3.3 Leg plates 1118.50X0
z
To attach the leg plate
1118.50X0
, insert the pin into
the locating hole up to the stop. Pull the locking lever
(
6
) towards the foot end up to the stop. Insert the pin
fully into the locating hole, with the locking lever still
pulled. Release the locking lever so that it returns to
its original position.
z
The leg plate can be folded down and into the horizon-
tal position by pressing down lever (
5
) at the foot end
of the operating table.
z
The leg plates can be abducted by slackening the grip
screw (
7
). To secure the leg plate in a detent posi-
tion, re-tighten the grip screw (
7
).
z
To remove the leg plate, pull the locking lever (
6
) to-
wards the foot end of the operating table and pull the
leg plate out from the locating holes.
17