Input / Output Configuration
The module is factory calibrated to provide between 0% and 100% output value when the positive input varies
between the negative and positive excitation.
Here you can calibrate the acquisition module and set a system zero if required.
This simple page allows semi-automated calibration where you can apply known inputs to calibrate.
This calibration includes linearization and is automatically applied.
See later for
By Cert
and
Advanced
page where you can adjust individual gains and offsets.
Calibration Process
Decide on how many points you will calibrate over.
Decide what inputs will be applied (in ascending order) at each point.
Enter the actual input (in the required units) that you want the module to read at each point.
Now proceed to apply each input in turn (allowing a settle time) and click the
Acquire
button at that
point. You can now apply the next input and click
Acquire
until all the points are completed.
The bottom of the page shows the
Input Value
and the
Calibrated Value
. Once the second point has been
acquired this
Calibrated Value
should display the actual calibrated value.