Zero Settings Advanced
This advanced section allows the use of a specially configured external module to supply the system zero value
for the handheld to use.
Example:
The same T24-PR1 is used with a truck that picks up different trailers and is required to display the sum of 4
strain gauges connected to each trailer (Using T24-SAs).
Because each trailer will have a different system zero requirement we would add a further device to each
trailer set to transmit the system zero value. It is the Data Tag that is entered here.
Note: On all trailers the acquisition module sets share the same Data Tags.
Items you can change:
Data Tag
Enter the
Data Tag
of the message to use for the external system zero.
ID
Contains the ID of the device used to supply the external system zero. This is only necessary to
provide a visible record of the remote device and is shown to keep compatibility with the
Mode
and Communications
page.
You do not need to enter anything here although it will be filled in automatically if you perform a
pair to retrieve data.
P
Click this to give 5 seconds to perform pairing to automatically provide the Data Tag and ID from a
specific device. Usually pairing is activated by removing and replacing the power supply.
X
Click this to reset the Data Tag and ID to zero (disabling the external system zero function).