IX. Specification

 

 s  

Models

NEP – 6185

NEP - 6303

Protection

Over Voltage Protection
(Main Output)

Vout +1V when Vout 

5V

Vout +1V when Vout 

5V

Vout +1.5V when 
5.1V 

Vout 

10V with <0.2sec.

Vout +1.5V when 
5.1V 

Vout 

10V with <0.2sec.

18~22% at tracking
10.1V <Vout 

19V

18~22% at tracking
10.1V <Vout 

33V

Short Circuit or Over Load

Yes

Over Temperature

100°C

100°C

Meter Accuracy

Voltage Meter Accuracy (V

2V) ±1% + 4 Digits

Voltage Meter Accuracy (V>2V) ±1% + 3 Digits

Current Meter Accuracy (I

1A)

±1% + 4 Digits

Voltage Meter Accuracy (I>1A)

±1% + 3 Digits

LCD Indications

Voltage and Current, C.C. And C.V., Aux.1 and 2 Status,
Output ON/OFF, 3 User Defined V & I Presets,
Running the Cycled Program, Fault Indication, OVP-Over Voltage,
OCP-Over Current or Load, OTP-Over Temperature

Presets Cycle Programming

Max Preset time: 999 seconds ;
Max Preset Cycles 999 Times or infinite cycles “0”

Output Terminals

Standard binding pole screw type or safety jack type

Approvals

CE  EMC: EN 55011, EN 61000  LVD: EN 61010

Fuse

2A x 1 (Slow)

Cooling System

Variable speed thermally control fan (0 to full speed)

Dimensions (WxHxD)

151 x 132 x 280 mm

Weight

5.4 Kgs

*All values are based on the Standard ambient Temperature 25°C and Pressure 0.1Mpa.

* SPECIFICATIONS ARE SUBJECT TO CHANGE WITHOUT PRIOR NOTICE *

Содержание NEP Series

Страница 1: ...NEP SERIES User Manual REV 0 2011 08 7673 6185 0000...

Страница 2: ...nels A FRONT PANEL 1 LCD display Indicates output voltage current presets CC CV status of main auxiliary outputs output On OFF running the Cycled Program and faults 2 Output Voltage Control knob Adjus...

Страница 3: ...to the unit 5 Do not replace the fuse before ceasing the problems use same fuse rating 6 Never touch the power supply when your hand is wet 7 Never remove the metal cover without unplugging of AC main...

Страница 4: ...e Knob then quickly press the Current Knob 1 Press the Voltage Knob and then press the Current Knob Icon 1 is flashing to show ready for setting 2 Adjust Voltage Knob for voltage value and Current Kno...

Страница 5: ...ew settings and Icon 3 will start flashing to indicate Preset 3 is ready for setting 5 Adjust Voltage and Current as before and press Voltage and Current Knobs to confirm Preset 3 new settings 6 All t...

Страница 6: ...cordingly 3 Repeat the above to get to Preset P2 then P3 4 Quick press of Voltage and Current Knobs at P3 will exit from the Preset Mode to normal operation mode N at which has no flashing Presets Ico...

Страница 7: ...oves to the next lot of Timed Presets When the power supply is in cycle operation the cycle icon flashes with the time preset in operation Setting the Timed Presets 1 Push Button 4 to Output OFF posit...

Страница 8: ...d P3 7 To set the number of cycles After t3 of P3 has been set and confirmed the P icon will then appears and ready for setting of number of cycles Adjust the voltage knob to set the number of cycles...

Страница 9: ...ycle Icon is flashing This serves as a warning of wrong setting of 3 preset timers Remarks A If the number of cycle is set at zero then the unit will run infinite cycles B To exit from the Timed Prese...

Страница 10: ...ent user presets 3 Programming of timer and auto cycling of preset in sequence 4 Rotary encoder tuning of V I 5 Output On Off switch with status indication 6 Large LCD display for V I Presets CC CV st...

Страница 11: ...on Ripple Noise r m s Current 10mA Load Regulation 10 90 Rated Voltage 30mA Line Regulation 5 Variation 30mA Setting Accuracy 0 3 0 02 V Current Resolution 0 01A Input Voltage Input Voltage 230VAC 50H...

Страница 12: ...LCD Indications Voltage and Current C C And C V Aux 1 and 2 Status Output ON OFF 3 User Defined V I Presets Running the Cycled Program Fault Indication OVP Over Voltage OCP Over Current or Load OTP Ov...

Отзывы: