Manson Engineering Industrial HCS-3600 USB Скачать руководство пользователя страница 6

9.

      

   REMOTE CONTROL MODE

 

 

There are two methods for remote control of current and voltage adjustment.
Both methods require current remote control part to be set up in order for remote control mode to be functional,
otherwise unit will be in CC mode all the time.

9.1 Method 

A  

Using two external variable DC voltage sources

Remote Socket Pin Assignment for external variable voltage source

PIN

FUNCTIONS

REMARKS

1

Internal DC +5V

Less then 50mA

2

Voltage Adjust

0 - 5V

3

Current Adjust

0 - 5V

4

Ground

5

Output OFF

Short to Ground

6

N.A.

7

N.A.

8

N.A.

Check the output voltage range of the power supply by varying the external voltage source.
Short circuit the main output with 

8AWG

 wire to check the display for CC setting varying the external voltage source.

9.2 Method 

B  

Using two 0 – 5K Ohm variable resistors

Remark: variable resistors 5Kohm.

Remote Socket Pin Assignment for variable resistor

PIN

FUNCTIONS

REMARKS

1

Internal DC +5V

Resistor end

2

Voltage Adjust

Variable part of resistor

3

Current Adjust

Variable part of resistor

4

Ground

Another resistor end

5

Output OFF

Short to Ground

6

N.A.

7

N.A.

8

N.A.

Check the output voltage range of the power supply by adjusting the 5Kohm variable resistor.
Short circuit the main output with 

8AWG

 wire to check the display for CC setting by adjusting the variable resistor.

Содержание HCS-3600 USB

Страница 1: ...o not operate or touch this power supply with wet hands Do not open the casing of the power supply when it is connected to ac mains Refer all servicing to qualified service personnel only Before replacing the AC fuse at AC socket find out and clear up the cause first Replace the AC fuse with the same type and rating as the original fuse The max output voltage of Model HCS 3604 is over 60VDC avoid ...

Страница 2: ...3600 The total rated current Aux Main is 60A HCS 3602 The total rated current Aux Main is 30A HCS 3604 The total rated current Aux Main is 15A 7 Output Terminal Rated 60A for HCS 3600 Rated 30A for HCS 3602 Rated 15A for HCS 3604 8 Remote Sensing Terminal for model HCS 3600 only Warning Shorting the remote sensing terminal or Connect sensing terminal in reverse polarity will damage the power suppl...

Страница 3: ... 15V HCS 3602 25V HCS 3604 55V Maximum 6 3 Set Mode First enter into the Set Mode by pushing Switch 9 to Set Mode slot The power supply is then ready to be preset 6 3 1 To define the preset output P1 P2 P3 a Select the Recall Switch 10 to the position P1 P2 or P3 which you want to set b Adjust the front panel voltage control knob to set your desired voltage value c Adjust the front panel current c...

Страница 4: ...d make sure it complies with your AC mains voltage Connect the power supply to the AC Mains using the provided power cord Make sure the Mode Switch 9 is at Normal Position Below table to show the self test sequence Self test display and Sequence Test contents To show software version Segment check C V Indicator check C C Indicator check Rear control indicator check Return to C V Start to check Ove...

Страница 5: ...rity input of the equipment 7 6 Switch on the power supply first and the panel meter green CV Indicator should light up again 7 7 Switch on the equipment and the panel meter green CV Indicator should still remain in green 7 8 You can now operate the equipment When an operation is finished switch off the equipment first and then switch off the power supply 8 REMOTE SENSING ONLY HCS 3600 HAS THIS FU...

Страница 6: ...put voltage range of the power supply by varying the external voltage source Short circuit the main output with 8AWG wire to check the display for CC setting varying the external voltage source 9 2 Method B Using two 0 5K Ohm variable resistors Remark variable resistors 5Kohm Remote Socket Pin Assignment for variable resistor PIN FUNCTIONS REMARKS 1 Internal DC 5V Resistor end 2 Voltage Adjust Var...

Страница 7: ...e modes Normal Preset Remote and Set mode A By default Pin 5 is open and output is on B Shorting Pin 5 to Pin 4 ground and output is off C When output is off the LED will show OP OFF Remark using the 8pin remote plug provided and connect with 22AWG wires Pin numbers are marked on the black portion ...

Страница 8: ...unit and remove all loading Check your load and output setting Allow the unit to cool down for at least 30 minutes Check if any of the ventilation is blocked check enough clearance around power supply Listen carefully for the short wind noise from the cooling fan when you turn on the unit again If you cannot hear this routine self test wind noise on switch on the fan is fault and do not use the po...

Страница 9: ...4 5A Efficiency 85 86 88 Switching Frequency 65 85Khz 75 95Khz 65 85Khz Tracking Over Voltage Protections O P 1 5V set voltage 2V O P 5 15V set voltage 3V O P 1 5V set voltage 2V O P 5 20V set voltage 3V O P 20 30V set voltage 4V O P 1 5V set voltage 2V O P 5 20V set voltage 3V O P 20 60V set voltage 4V Transient Response Time 50 100 Load 1 5ms Power Factor Control Power factor correction 0 95 at ...

Отзывы: