Installation
M&C SP180-H probes are designed for stationary use. The professional choice of the right sample probe for the
application and the professional installation garantees a long service life and minimum maintenance work. The
ideal position for mounting the sample probe is horizontally with a tilt angle of 10% towards the process. Please
follow these installation steps and see details in Figure 2.
•
Screw the sample probe tube directly into the ¾" inner thread
of the probe flange and tighten it.
•
To mount the SP180-H at the sampling flange, please loosen the metal clamp
around the heat insulating
protection cover and remove the cover
.
•
If the probe connection is not the same size as the standard flange connection DN65 PN6, an adapter flange
(optional) should be mounted to the probe (see Figure 2). Place the gasket
between the adapter
flange and the probe flange.
•
If the heated probe tube type SP30/35 is used then the probe is to be screwed to its flange (with welded
threaded bolt). First insert the flange seal
between the two flanges.
•
Attach the flange seal
to the probe connection.
•
Insert sample probe with mounted probe tube into the connection piece and fasten the probe using the
supplied bolts and nuts.
•
After mounting of the probe at the sampling flange put the heat insulating protection cover
over the
probe flange again and secure it with the metal clamp
.
Note
For the preferred mounting position of the probe, the sample gas
outlet is pointing downwards. This is just the preferred mounting
position, it is not necessary for correct operation.
Figure 2
Mounting of the SP180-H